Fatigue Resistance of Nickel-titanium Instruments Exposed to High-concentration Hypochlorite

被引:37
作者
Huang, Xiangya [1 ,2 ]
Shen, Ya [2 ,3 ]
Wei, Xi [1 ]
Haapasalo, Markus [2 ]
机构
[1] Sun Yat Sen Univ, Guangdong Prov Key Lab Stomatol, Guanghua Sch Stomatol, Dept Conservat Dent & Endodont, Guangzhou, Guangdong, Peoples R China
[2] Univ British Columbia, Fac Dent, Dept Oral Biol & Med Sci, Div Endodont, 2199 Wesbrook Mall, Vancouver, BC V6T 1Z3, Canada
[3] Univ British Columbia, Dept Mat Engn, Vancouver, BC, Canada
基金
加拿大创新基金会;
关键词
Ceramic canal; corrosion; fatigue resistance; nickel-titanium instrument; sodium hypochlorite; temperature; PHASE-TRANSFORMATION BEHAVIOR; LOW-CYCLE FATIGUE; SODIUM-HYPOCHLORITE; ROTARY INSTRUMENTS; CLINICAL-USE; ENDODONTIC INSTRUMENTS; PROFILE VORTEX; PROTAPER GOLD; TEMPERATURE; FILES;
D O I
10.1016/j.joen.2017.06.033
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: The purpose of this study was to introduce a new fatigue test model that simulates the clinical situation for evaluating the corrosion effect of 5.25% sodium hypochlorite (NaOCl) on nickel-titanium (NiTi) files and to evaluate the effect of 3 different temperatures (22 degrees C, 37 degrees C, and 60 degrees C) on the cyclic fatigue of these files. Methods: Three NiTi files (size 25/.04), K3 (SybronEndo, Orange, CA), K3XF (SybronEndo), and Vortex (Dentsply Tulsa Dental Specialties, Tulsa, OK), were subjected to cyclic fatigue tests inside a novel artificial ceramic canal with a curvature of 60 degrees and a 5-mm radius. A 19-mm-long file segment from the tip was introduced into the canal and immersed in water or 5.25% NaOCl at 3 different temperatures, and the number of revolutions to fracture (N-f) was recorded. The fracture surface of all fragments was examined by a scanning electron microscope. Data were analyzed using univariate analysis of variance with the significance level at 0.05. Results: The N-f of Vortex files was the highest followed by K3XF and K3 (P < .05) at all conditions. The Nf of all files was highest at 22 degrees C and lowest at 60 degrees C (P < .05). However, no difference in N-f was detected in Vortex files between 22 degrees C and 37 degrees C. The N-f of all files in 5.25% NaOCl was shorter than that in water although there was no statistically significant difference. No pitting or crevice corrosion was observed on the fracture surface. Conclusions: NaOCl, 5.25%, does not significantly affect the fatigue behavior of NiTi files. The fatigue resistance should be tested under specific temperature conditions. The austenite finish temperature of a file is important in determining the fracture risk at body temperature.
引用
收藏
页码:1847 / 1851
页数:5
相关论文
共 32 条
[21]   A Review of Cyclic Fatigue Testing of Nickel-Titanium Rotary Instruments [J].
Plotino, Gianluca ;
Grande, Nicola M. ;
Cordaro, Massimo ;
Testarelli, Luca ;
Gambarini, Gianluca .
JOURNAL OF ENDODONTICS, 2009, 35 (11) :1469-1476
[22]   Measurement of the trajectory of different NiTi rotary instruments in an artificial canal specifically designed for cyclic fatigue tests [J].
Plotino, Gianluca ;
Grande, Nicola M. ;
Cordaro, Massimo ;
Testarelli, Luca ;
Gambarini, Gianluca .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2009, 108 (03) :E152-E156
[23]   Electropolishing Enhances the Resistance of Nickel-Titanium Rotary Files to Corrosion Fatigue-Failure in Hypochlorite [J].
Praisarnti, Chonrada ;
Chang, Jeffrey W. W. ;
Cheung, Gary S. P. .
JOURNAL OF ENDODONTICS, 2010, 36 (08) :1354-1357
[24]   Cyclic fatigue testing of nickel-titanium endodontic instruments [J].
Pruett, JP ;
Clement, DJ ;
Carnes, DL .
JOURNAL OF ENDODONTICS, 1997, 23 (02) :77-85
[25]   Fatigue and nanomechanical properties of K3XF nickel-titanium instruments [J].
Shen, Y. ;
Zhou, H. ;
Campbell, L. ;
Wang, Z. ;
Wang, R. ;
Du, T. ;
Haapasalo, M. .
INTERNATIONAL ENDODONTIC JOURNAL, 2014, 47 (12) :1160-1167
[26]   Comparison of defects in ProFile and ProTaper systems after clinical use [J].
Shen, Y ;
Cheung, GSP ;
Bian, ZA ;
Peng, B .
JOURNAL OF ENDODONTICS, 2006, 32 (01) :61-65
[27]   ProFile Vortex and Vortex Blue Nickel-Titanium Rotary Instruments after Clinical Use [J].
Shen, Ya ;
Zhou, Huimin ;
Coil, Jeffrey M. ;
Aljazaeri, Bassim ;
Buttar, Rene ;
Wang, Zhejun ;
Zheng, Yu-feng ;
Haapasalo, Markus .
JOURNAL OF ENDODONTICS, 2015, 41 (06) :937-942
[28]   Phase Transformation Behavior and Mechanical Properties of Thermomechanically Treated K3XF Nickel-Titanium Instruments [J].
Shen, Ya ;
Zhou, Hui-min ;
Wang, Zhejun ;
Campbell, Les ;
Zheng, Yu-feng ;
Haapasalo, Markus .
JOURNAL OF ENDODONTICS, 2013, 39 (07) :919-923
[29]   Effect of Environment on Fatigue Failure of Controlled Memory Wire Nickel-Titanium Rotary Instruments [J].
Shen, Ya ;
Qian, Wei ;
Abtin, Houman ;
Gao, Yuan ;
Haapasalo, Markus .
JOURNAL OF ENDODONTICS, 2012, 38 (03) :376-380
[30]   Fatigue Testing of Controlled Memory Wire Nickel-Titanium Rotary Instruments [J].
Shen, Ya ;
Qian, Wei ;
Abtin, Houman ;
Gao, Yuan ;
Haapasalo, Markus .
JOURNAL OF ENDODONTICS, 2011, 37 (07) :997-1001