Calibration of CryojetHT and Cobra Plus Cryosystems used in X-ray diffraction studies

被引:21
作者
Dudka, A. P. [1 ]
Verin, I. A. [1 ]
Smirnova, E. S. [1 ]
机构
[1] Russian Acad Sci, Shubnikov Inst Crystallog, Leninskii Pr 59, Moscow 119333, Russia
基金
俄罗斯基础研究基金会;
关键词
DATA-COLLECTION; CRYOCRYSTALLOGRAPHY;
D O I
10.1134/S1063774516040052
中图分类号
O7 [晶体学];
学科分类号
0702 ; 070205 ; 0703 ; 080501 ;
摘要
CryoJetHT (Oxford Instruments) and Cobra Plus (Oxford Cryosystems) cryosystems, which are used for sample cooling in X-ray diffraction experiments, have been calibrated. It is shown that the real temperature in the vicinity of the sample differs significantly (the deviation is as high as 8-10 K at low temperatures) from the temperature recorded by authorized sensors of these systems. The calibration results are confirmed by measurements of the unit-cell parameters of GdFe3(BO3)(4) single crystal in the temperature range of its phase transition. It is shown that, to determine the real temperature of a sample, one must perform an independent calibration of cryosystems rather than rely on their ratings.
引用
收藏
页码:692 / 696
页数:5
相关论文
共 10 条
[1]  
Coppens P., 1997, X-ray charge densities and chemical bonding
[2]   Raman scattering from phonons and magnons in RFe3(BO3)4 [J].
Fausti, Daniele ;
Nugroho, Agung A. ;
van Loosdrecht, Paul H. M. ;
Klimin, Sergei A. ;
Popova, Marina N. ;
Bezmaternykh, Leonard N. .
PHYSICAL REVIEW B, 2006, 74 (02)
[3]   Thermal equilibration of samples for neutron scattering [J].
Herman, Tobias K. ;
Parks, Sarah C. ;
Scherschligt, Julia .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 2013, 46 :279-285
[4]   CRYOCRYSTALLOGRAPHY OF BIOLOGICAL MACROMOLECULES - A GENERALLY APPLICABLE METHOD [J].
HOPE, H .
ACTA CRYSTALLOGRAPHICA SECTION B-STRUCTURAL SCIENCE, 1988, 44 :22-26
[5]   Low-temperature synchrotron radiation study of a twinned disordered crystal of bis(4,4′-bromophenyl)-61,61-diyl methano fullerene C60 [J].
Iversen, BB ;
Darovsky, A ;
Bolotovsky, R ;
Coppens, P .
ACTA CRYSTALLOGRAPHICA SECTION B-STRUCTURAL SCIENCE, 1998, 54 :174-179
[7]  
Levitin RZ, 2004, JETP LETT+, V79, P423, DOI 10.1134/1.1776236
[8]   High-resolution synchrotron data collection for charge-density work at 100 and 20 K [J].
Luger, P ;
Messerschmidt, M ;
Scheins, S ;
Wagner, A .
ACTA CRYSTALLOGRAPHICA A-FOUNDATION AND ADVANCES, 2004, 60 :390-396
[9]   Macromolecular cryocrystallography: Cooling, mounting, storage and transportation of crystals [J].
Parkin, S ;
Hope, H .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1998, 31 :945-953
[10]   A modified Pinkerton-type helium gas-flow system for high-accuracy data collection at the X3 SUNY synchrotron beamline at NSLS [J].
Ribaud, L ;
Wu, G ;
Zhang, YG ;
Coppens, P .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 2001, 34 :76-79