The Effects of Hearing Aid Digital Noise Reduction and Directionality on Acceptable Noise Level

被引:12
作者
Ahmadi, Roghayeh [1 ]
Jalilvand, Hamid [1 ]
Mahdavi, Mohammad Ebrahim [1 ]
Ahmadi, Fatemeh [2 ]
Baghban, Ali Reza Akbarzade [1 ]
机构
[1] Shahid Beheshti Univ Med Sci, Dept Audiol, Sch Rehabil, Damavand St, Tehran, Iran
[2] Allameh Tabatabai Univ, Sch Econ, Tehran, Iran
关键词
Directional Hearing Aid; Hearing Aids; Acceptable Noise Level; Hearing Loss; Digital Noise Reduction; WORKING-MEMORY CAPACITY; SPEECH RECOGNITION; MICROPHONE;
D O I
10.21053/ceo.2018.00052
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Objectives. Two main digital signal processing technologies inside the modern hearing aid to provide the best conditions for hearing aid users are directionality (DIR) and digital noise reduction (DNR) algorithms. There are various possible settings for these algorithms. The present study evaluates the effects of various DIR and DNR conditions (both separately and in combination) on listening comfort among hearing aid users. Methods. In 18 participants who received hearing aid fitting services from the Rehabilitation School of Shahid Beheshti University of Medical Sciences regularly, we applied acceptable noise level (ANL) as our subjective measure of listening comfort. We evaluated both of these under six different hearing aid conditions: omnidirectional-baseline, omnidirectional-broadband DNR, omnidirectional-multichannel DNR, directional, directional-broadband DNR, and directional-multichannel DNR. Results. The ANL, results ranged from -3 dB to 14 dB in all conditions. The results show, among all conditions, both the omnidirectional-baseline condition and the omnidirectional-broadband DNR condition are the worst conditions for listening in noise. The DIR always reduces the amount of noise that patients received during testing. The DNR algorithm does not improve listening in noise significantly when compared with the DIR algorithms. Although both DNR and DIR algorithms yielded a lower ANL, the DIR algorithm was more effective than the DNR. Conclusion. The DIR and DNR technologies provide listening comfort in the presence of noise. Thus, user benefit depends on how the digital signal processing settings inside the hearing aid are adjusted.
引用
收藏
页码:267 / 274
页数:8
相关论文
共 24 条
[1]   The Acceptable Noise Level Benefit From Directionality for Listeners With Severe Hearing Loss [J].
Aghsoleimani, Mina ;
Jalilvand, Hamid ;
Mahdavi, Mohammad Ebrahim ;
Nazeri, Ahmad Reza ;
Karnali, Mohammad .
CLINICAL AND EXPERIMENTAL OTORHINOLARYNGOLOGY, 2018, 11 (03) :166-173
[2]  
Ahmadi A, 2015, J Rehab Med, V4, P109
[3]   Evaluation of the noise reduction system in a commercial digital hearing aid [J].
Alcántara, JI ;
Moore, BCJ ;
Kühnel, V ;
Launer, S .
INTERNATIONAL JOURNAL OF AUDIOLOGY, 2003, 42 (01) :34-42
[4]  
[Anonymous], 2012, HEARING AIDS
[5]   Validation of a Mini-Mental State Examination (MMSE) for the Persian Population: A Pilot Study [J].
Ansari, Noureddin Nakhostin ;
Naghdi, Soofia ;
Hasson, Scott ;
Valizadeh, Leala ;
Jalaie, Shohreh .
APPLIED NEUROPSYCHOLOGY, 2010, 17 (03) :190-195
[6]  
Boymans M, 2000, AUDIOLOGY, V39, P260
[7]   Acoustical and Perceptual Comparison of Noise Reduction and Compression in Hearing Aids [J].
Brons, Inge ;
Houben, Rolph ;
Dreschler, Wouter A. .
JOURNAL OF SPEECH LANGUAGE AND HEARING RESEARCH, 2015, 58 (04) :1363-1376
[8]   The Effects of Hearing Aid Directional Microphone and Noise Reduction Processing on Listening Effort in Older Adults with Hearing Loss [J].
Desjardins, Jamie L. .
JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2016, 27 (01) :29-41
[9]  
Freyaldenhoven Melinda C, 2005, J Am Acad Audiol, V16, P228, DOI 10.3766/jaaa.16.4.4
[10]  
Kochkin S., 2002, Hearing Rev, V9, P18