A SIMPLE AND NOVEL INTEGRATED OPTO-ELECTRONIC SYSTEM FOR BLOOD VOLUME PULSE SENSING AND HEART RATE MONITORING

被引:3
作者
Jayasree, Vadakke Kadangote [1 ]
Shaija, Palackappillil Jacob [1 ]
Nampoori, Vadakkedathu Parameswaran Narayanan [1 ]
Girijavallabhan, Chakkalakkal Pavothil [2 ]
Radhakrishnan, Padmanabhan [1 ]
机构
[1] Cochin Univ Sci & Technol, Int Sch Photon, Kochi 682022, India
[2] Cochin Univ Sci & Technol, Ctr Excellence Lasers & Optoelect Sci, Kochi 682022, India
关键词
D O I
10.1080/15599610701779325
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A high sensitivity, low power, low cost sensor has been developed for sensing the blood volume pulse using transmission mode photoplethysmography (PPG) from the finger tip. It uses standard light emitting diode (LED) and phototransistor as emitter and detector, respectively. The sensor probe head is made of a material called delrin which is an acetal polymer having high dimensional stability and biocompatibility. A peripheral interface controller (PIC) microcontroller is used for the implementation of the measurement protocol. The wavelet denoising algorithm is used to suppress the noise component in the PPG signal. For heart rate (HR) estimate comparison, PPG signals are evaluated by comparing their beat-to-beat estimates with the corresponding R-R intervals from an electrocardiogram (ECG). The main advantages of the proposed approach are the reduction in cost, dimensions and power consumption. The probe can be well tolerated by the subject and is self-contained and portable.
引用
收藏
页码:392 / 403
页数:12
相关论文
共 21 条
[11]   Standard pulse oximeters can be used to monitor respiratory rate [J].
Leonard, P ;
Beattie, TF ;
Addison, PS ;
Watson, JN .
EMERGENCY MEDICINE JOURNAL, 2003, 20 (06) :524-525
[12]  
MENDELSON Y, 1992, CLIN CHEM, V38, P1601
[13]   Determination of age-related increases in large artery stiffness by digital pulse contour analysis [J].
Millasseau, SC ;
Kelly, RP ;
Ritter, JM ;
Chowienczyk, PJ .
CLINICAL SCIENCE, 2002, 103 (04) :371-377
[14]   Noninvasive assessment of the digital volume pulse - Comparison with the peripheral pressure pulse [J].
Millasseau, SC ;
Guigui, FG ;
Kelly, RP ;
Prasad, K ;
Cockcroft, JR ;
Ritter, JM ;
Chowienczyk, PJ .
HYPERTENSION, 2000, 36 (06) :952-956
[15]  
NIJBOER J A, 1981, Clinical Physics and Physiological Measurement, V2, P205, DOI 10.1088/0143-0815/2/3/004
[16]   Artifact-resistant power-efficient design of finger-ring plethysmographic sensors [J].
Rhee, S ;
Yang, BH ;
Asada, HH .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2001, 48 (07) :795-805
[17]   Comment on the modified Beer-Lambert law for scattering media [J].
Sassaroli, A ;
Fantini, S .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (14) :N255-N257
[18]   A LED-LED-based photoplethysmography sensor [J].
Stojanovic, R. ;
Karadaglic, D. .
PHYSIOLOGICAL MEASUREMENT, 2007, 28 (06) :N19-N27
[19]   The effect of contacting force on photoplethysmographic signals [J].
Teng, XF ;
Zhang, YT .
PHYSIOLOGICAL MEASUREMENT, 2004, 25 (05) :1323-1335
[20]  
TROFS T, 2007, SENSORS TRANSDUCERS, V80, P1230