A comparison of polyacrylamide gels and radiochromic film for source measurements in intravascular brachytherapy

被引:14
作者
Amin, MN
Horsfield, MA
Bonnett, DE [1 ]
Dunn, MJ
Poulton, M
Harding, PF
机构
[1] Univ Leicester, Div Med Phys, Leicester LE1 5WW, Leics, England
[2] Maidstone & Tunbridge Wells NHS Trust, Kent Oncol Ctr, Dept Med Phys, Kent ME16 9QQ, England
[3] Univ Hosp Leicester, Dept Med Phys, Leicester LE1 5WW, Leics, England
关键词
D O I
10.1259/bjr/25639755
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
For intravascular brachytherapy with catheter-based systems, AAPM Task Group 60 has recommended measurements that should be made to characterize the sources. Beta emitters, including Sr-90/Y-90 are ideal for intravascular brachytherapy, but problems arise in measuring dose distributions in the high dose gradient region at short distances from the source. In this paper, measurements of radial and orthogonal dose distributions and dose profiles for a Sr-90/Y-90 source train using polyacrylamide gel (PAG) dosimetry and a high-field 4.7 Tesla MRI scanner are presented and compared with measurements made with two types of radiochromic film, MD-55 and HD-810. For the PAG system, the dose distributions were determined with in-plane resolutions of 0.4 mm and 0.2 mm. The measurements of absorbed dose distributions both orthogonal and parallel to the source axis show good agreement between the PAG and radiochromic film. The absolute dose at a radial distance of 2 mm in the central 32 mm of a line parallel to the axis was measured. For the PAG the measured absorbed dose was 1.25% lower, for MD-55 4% higher and for the HD-810 1.6% higher when compared with the value given by the source calibration. These results confirm that both absorbed dose and dose distributions for high gradient vascular brachytherapy sources can be measured using PAG but the disadvantages of gel manufacture and the need for access to a high resolution scanner suggests that the use of radiochromic film is the method of choice.
引用
收藏
页码:824 / 831
页数:8
相关论文
共 29 条
[1]   A high-precision, high-resolution and fast dosimetry system for beta sources applied in cardiovascular brachytherapy [J].
Bambynek, M ;
Flühs, D ;
Quast, U ;
Wegener, D ;
Soares, CG .
MEDICAL PHYSICS, 2000, 27 (04) :662-667
[2]   High resolution polymer gel dosimetry by parameter selective MR-microimaging on a whole body scanner at 3 T [J].
Berg, A ;
Ertl, A ;
Moser, E .
MEDICAL PHYSICS, 2001, 28 (05) :833-843
[3]  
BONNETT DE, 1999, P 1 INT WORKSH RAD T, P142
[4]   The reproducibility of polyacrylamide gel dosimetry applied to stereotactic conformal radiotherapy [J].
Cosgrove, VP ;
Murphy, PS ;
McJury, M ;
Adams, EJ ;
Warrington, AP ;
Leach, MO ;
Webb, S .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (05) :1195-1210
[5]   Three-dimensional dosimetry using polymer gel and magnetic resonance imaging applied to the verification of conformal radiation therapy in head-and-neck cancer [J].
De Deene, Y ;
De Wagter, C ;
Van Duyse, B ;
Derycke, S ;
De Neve, W ;
Achten, E .
RADIOTHERAPY AND ONCOLOGY, 1998, 48 (03) :283-291
[6]  
De Deene Y, 2000, PHYS MED BIOL, V45, P859, DOI 10.1088/0031-9155/45/4/304
[7]  
De Deene YH, 2001, PHYS MED BIOL, V46, P2801, DOI 10.1088/0031-9155/46/11/304
[8]   SIGNAL, NOISE, AND CONTRAST IN NUCLEAR MAGNETIC-RESONANCE (NMR) IMAGING [J].
EDELSTEIN, WA ;
BOTTOMLEY, PA ;
HART, HR ;
SMITH, LS .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1983, 7 (03) :391-401
[9]   High-resolution dose profile studies based on MR Imaging with polymer BANG™ gels in stereotactic radiation techniques [J].
Ertl, A ;
Berg, A ;
Zehetmayer, M ;
Frigo, P .
MAGNETIC RESONANCE IMAGING, 2000, 18 (03) :343-349
[10]   An investigation into the use of polymer gel dosimetry in low dose rate brachytherapy [J].
Farajollahi, AR ;
Bonnett, DE ;
Ratcliffe, AJ ;
Aukett, RJ ;
Mills, JA .
BRITISH JOURNAL OF RADIOLOGY, 1999, 72 (863) :1085-1092