Raman spectroscopic measurement of tablet-to-tablet coating variability

被引:77
作者
Romero-Torres, S
Pérez-Ramos, JD
Morris, KR
Grant, ER
机构
[1] Purdue Univ, Dept Ind & Phys Pharm, W Lafayette, IN 47907 USA
[2] Purdue Univ, Dept Chem, W Lafayette, IN 47907 USA
关键词
Raman spectroscopy; tablet coating; process analysis; process control; multivariate classification;
D O I
10.1016/j.jpba.2005.01.007
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We report new results suggesting the feasibility of Raman spectrometry as a tool by which to examine the variability of tablet coatings. Our experiments feature a probe that can operate with a revolving laser focus to average content and coating non-uniformity. Raman spectral changes are correlated with tablet exposure times in a pan coater by means of partial least squares (PLS) multivariate analysis. Statistical models are found to be improved by pre-processing schemes that emphasize spectral changes while minimizing the effects of background light scattering and fluorescence. These pre-processing techniques include multiplicative scatter correction (MSC) and standard normal variate (SNV) transformation, used in concert with Savitzky-Golay second derivative smoothing (SGSD). The two approaches give comparable results yielding R-2 values for PLS calibration and cross-calibrated prediction variance regression of 0.999 and 0.997, respectively. Correlation results and model residual values demonstrate that Raman spectroscopy serves sensitively to reflect the coating thickness of the tablets studied. (c) 2005 Elsevier B.V.. All rights reserved.
引用
收藏
页码:270 / 274
页数:5
相关论文
共 14 条
[1]   Quantitative analysis of film coating in a fluidized bed process by in line NIR spectrometry and multivariate batch calibration [J].
Andersson, M ;
Folestad, S ;
Gottfries, J ;
Johansson, MO ;
Josefson, M ;
Wahlund, KG .
ANALYTICAL CHEMISTRY, 2000, 72 (09) :2099-2108
[2]   Monitoring of a film coating process for tablets using near infrared reflectance spectrometry [J].
Andersson, M ;
Josefson, M ;
Langkilde, FW ;
Wahlund, KG .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1999, 20 (1-2) :27-37
[3]   STANDARD NORMAL VARIATE TRANSFORMATION AND DE-TRENDING OF NEAR-INFRARED DIFFUSE REFLECTANCE SPECTRA [J].
BARNES, RJ ;
DHANOA, MS ;
LISTER, SJ .
APPLIED SPECTROSCOPY, 1989, 43 (05) :772-777
[4]  
Brereton R. C., 2002, CHEMOMETRICS DATA AN
[5]  
Dhanoa M., 1994, Journal of Near Infrared Spectroscopy, V2, P43, DOI [10.1255/jnirs.30, DOI 10.1255/JNIRS.30]
[6]  
Esbensen K. H., 2002, MULTIVARIATE DATA AN
[7]   PARTIAL LEAST-SQUARES REGRESSION - A TUTORIAL [J].
GELADI, P ;
KOWALSKI, BR .
ANALYTICA CHIMICA ACTA, 1986, 185 :1-17
[8]   LINEARIZATION AND SCATTER-CORRECTION FOR NEAR-INFRARED REFLECTANCE SPECTRA OF MEAT [J].
GELADI, P ;
MACDOUGALL, D ;
MARTENS, H .
APPLIED SPECTROSCOPY, 1985, 39 (03) :491-500
[9]   Near-infrared spectroscopic monitoring of the film coating process [J].
Kirsch, JD ;
Drennen, JK .
PHARMACEUTICAL RESEARCH, 1996, 13 (02) :234-237
[10]   DETERMINATION OF FILM-COATED TABLET PARAMETERS BY NEAR-INFRARED SPECTROSCOPY [J].
KIRSCH, JD ;
DRENNEN, JK .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1995, 13 (10) :1273-1281