First-in-man transcatheter aortic valve implantation of a 20-mm edwards SAPIEN XT valve: One step forward for the treatment of patients with severe aortic stenosis and small aortic annulus
被引:13
|作者:
Rodes-Cabau, Josep
[1
]
DeLarochelliere, Robert
[1
]
Dumont, Eric
[1
]
机构:
[1] Univ Laval, Quebec Heart & Lung Inst, Quebec City, PQ G1V 4G5, Canada
关键词:
transcatheter aortic valve implantation;
aortic stenosis;
small aortic annulus;
ROOT RUPTURE;
MISMATCH;
D O I:
10.1002/ccd.23198
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
We present the case of an 85-year-old woman diagnosed with severe aortic stenosis, porcelain aorta, and a small aortic annulus (17.3 mm), who underwent successful transfemoral transcatheter aortic valve implantation (TAVI) with a 20-mm Edwards SAPIEN XT valve using the NovaFlex+ delivery system. At 1-month follow-up the patient was in NYHA functional class I, and Doppler echocardiography showed a mean residual gradient of 15 mm Hg and trivial paravalvular aortic regurgitation. This case, which shows for the first time the feasibility of TAVI with a 20-mm valve, opens a new avenue for the challenging treatment of patients with aortic stenosis and a small aortic annulus. (c) 2011 Wiley Periodicals, Inc.