Fast lifetime and amplitude determination in luminescence exponential decays

被引:14
作者
Lopez-Ruiz, N. [1 ]
Hernandez-Belanger, D. [1 ]
Carvajal, M. A. [1 ]
Capitan-Vallvey, L. F. [2 ]
Palma, A. J. [1 ]
Martinez-Olmos, A. [1 ]
机构
[1] Univ Granada, Dept Elect & Comp Technol, CITIC UGR, ECsens, E-18071 Granada, Spain
[2] Univ Granada, Dept Analyt Chem, ECsens, E-18071 Granada, Spain
关键词
Luminescence; Lifetime measurement; Oxygen sensor; Color detector; OXYGEN SENSORS; FLUORESCENCE; TEMPERATURE;
D O I
10.1016/j.snb.2015.04.077
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this work a novel algorithm for the fitting of an exponential decay luminescence curve is proposed. It is based on the integration of the luminescence intensity over two or three temporal windows in different states of the sample: steady state with and without optical excitation and during the whole luminescence intensity decay. These measurements allow to obtain the amplitude of the signal, the offset and the lifetime respectively, not only for single exponential decays, but also for multiple exponential behavior, if averaged values are considered. The proposed approach does not require a very fast instrumentation, since the width of the integration windows are fixed to values much higher than the expected lifetime of the sample, which represents a clear advantage over other window-based methods for the determination of luminescence lifetime. In addition, no previous knowledge of the expected lifetime is necessary. Specifically, this algorithm is applied for the fitting of the luminescence curve of an oxygen sensor based on PtOEP, and implemented on a very reduced and simple electronics, consisting basically on a digital color detector of 12-bits resolution that allow to define integration times high enough for measurements of the intensity using this approach. Good results are achieved, proving the benefits of the method. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:595 / 602
页数:8
相关论文
共 48 条
[1]  
[Anonymous], 2004, Optical Sensors: Industrial, Environmental and Diagnostic Applications
[2]   PHASE PLANE AND GUGGENHEIM METHODS FOR TREATMENT OF KINETIC DATA [J].
BACON, JR ;
DEMAS, JN .
ANALYTICAL CHEMISTRY, 1983, 55 (04) :653-656
[3]   Dual fluorescence sensor for trace oxygen and temperature with unmatched range and sensitivity [J].
Baleizao, Carlos ;
Nagl, Stefan ;
Schaeferling, Michael ;
Berberan-Santos, Mario N. ;
Wolfbeis, Otto S. .
ANALYTICAL CHEMISTRY, 2008, 80 (16) :6449-6457
[4]   AN ERROR ANALYSIS OF THE RAPID LIFETIME DETERMINATION METHOD FOR THE EVALUATION OF SINGLE EXPONENTIAL DECAYS [J].
BALLEW, RM ;
DEMAS, JN .
ANALYTICAL CHEMISTRY, 1989, 61 (01) :30-33
[5]   Absorption and emission spectroscopic characterization of platinum-octaethyl-porphyrin (PtOEP) [J].
Bansal, A. K. ;
Holzer, W. ;
Penzkofer, A. ;
Tsuboi, Taiju .
CHEMICAL PHYSICS, 2006, 330 (1-2) :118-129
[6]   Fabrication of luminescent nanostructures by electron-beam direct writing of PMMA resist [J].
Barrios, C. A. ;
Carrasco, S. ;
Canalejas-Tejero, V. ;
Lopez-Romero, D. ;
Navarro-Villoslada, F. ;
Moreno-Bondi, M. C. ;
Fierro, J. L. G. ;
Capel-Sanchez, M. C. .
MATERIALS LETTERS, 2012, 88 :93-96
[7]   Studies on the fabrication and characterization of optical sensor coatings for aerodynamic applications [J].
Bharathidasan, T. ;
Rikhari, Bhavana ;
Prasannan, Dijo ;
Dinesh Kumar, V. ;
Chakradhar, R.P.S. ;
Basu, Bharathibai J. .
Journal of Applied Sciences, 2012, 12 (16) :1646-1650
[8]   Fluorescence detections and directions [J].
Birch, D. J. S. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2011, 22 (05)
[9]  
Borisov SM, 2010, INFORM MIDEM, V40, P291
[10]   Oxygen-sensing film coated photodetectors for portable instrumentation [J].
Capitan-Vallvey, L. F. ;
Asensio, L. J. ;
Lopez-Gonzalez, J. ;
Fernandez-Ramos, M. D. ;
Palma, A. J. .
ANALYTICA CHIMICA ACTA, 2007, 583 (01) :166-173