Reproducibility of the New Goldmann AT900D Digital Tonometer

被引:4
作者
Morales-Fernandez, Laura [1 ]
Martinez-de-la-Casa, Jose M. [1 ]
Garcia-Feijoo, Julian [1 ]
Saenz-Frances, Federico [1 ]
Santos, Enrique [1 ]
Garcia-Sanchez, Julian [1 ]
机构
[1] Univ Complutense, Dept Oftalmol, Hosp Clin San Carlos, Inst Invest Oftalmol Ramon Castroviejo, Madrid 28240, Spain
关键词
glaucoma; tonometer; Goldmann; intraocular pressure; reproducibility; DYNAMIC CONTOUR TONOMETRY; CENTRAL CORNEAL THICKNESS; APPLANATION TONOMETRY; INTRAOCULAR-PRESSURE; ACCURACY;
D O I
10.1097/IJG.0b013e31820bd1e9
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To examine the reproducibility of the new digital Goldmann applanation tonometer AT 900D (DGT) for measuring intraocular pressure and to compare its readings with those provided by the conventional analog Goldmann tonometer (AGT). Methods: In 3 different sessions (S1, S2, and S3), 3 intraocular pressure measurements were made in 15 eyes of 15 healthy subjects using both the DGT and AGT. The order of use of the tonometers was random and the data were collected by a different examiner to the one taking the measurements. Data were compared by determining coefficients of variation for each session and representing the differences detected as Bland-Altman plots. Results: The mean differences recorded (AGT-DGT) were 0.1, 0.3, and -0.1 for S1, S2, and S3, respectively. None of the differences were significant and none were >= 3 mm Hg. Coefficients of variation were lower in the 3 sessions for the DGT, the difference being significant for S2 and S3 (4.87% vs. 3.99%, P = 0.400; 4.84% vs. 3.55%, P = 0.028; and 5.04% vs. 3.50%, P = 0.025 in S1, S2, and S3 respectively). Conclusions: The new digital Goldmann tonometer shows a high reproducibility and excellent correlation of measures with those of the AGT.
引用
收藏
页码:186 / 188
页数:3
相关论文
共 18 条
[1]   The effect of repeated applanation on subsequent IOP measurements [J].
AlMubrad, Turki M. ;
Ogbuehi, Kelechi C. .
CLINICAL AND EXPERIMENTAL OPTOMETRY, 2008, 91 (06) :524-529
[2]  
Bland J, 1986, LANCET, V1, P308
[3]   Digit preference in Goldmann applanation tonometry: The hedgehog effect [J].
Buller, AJ ;
Chatzinikolas, K ;
Giannopoulos, N ;
Nair, J ;
Kelleher, SH ;
Raj, SD ;
Roy, K ;
Naylor, G .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2005, 140 (03) :527-529
[4]   Assessment of true intraocular pressure: The gap between theory and practical data [J].
Chihara, Etsuo .
SURVEY OF OPHTHALMOLOGY, 2008, 53 (03) :203-218
[5]   Accuracy of the new ICare rebound tonometer vs. other portable tonometers in healthy eyes [J].
García-Resúa, C ;
González-Meijome, JM ;
Gilino, J ;
Yebra-Pimentel, E .
OPTOMETRY AND VISION SCIENCE, 2006, 83 (02) :102-107
[6]   Agreement between two Goldmann type applanation tonometers [J].
George, Ronnie ;
Arvind, Hemamalini ;
Baskaran, M. ;
Ramesh, S. Ve ;
Raju, Prema ;
Vijaya, Lingam .
INDIAN JOURNAL OF OPHTHALMOLOGY, 2008, 56 (06) :516-517
[7]   GOLDMANN APPLANATION TONOMETRY IN PATIENTS WITH REGULAR CORNEAL ASTIGMATISM [J].
HOLLADAY, JT ;
ALLISON, ME ;
PRAGER, TC .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1983, 96 (01) :90-93
[8]   Incorporating corneal pachymetry into the management of glaucoma [J].
Iester, Michele ;
Mete, Maurizio ;
Figus, Michele ;
Frezzotti, Paolo .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (09) :1623-1628
[9]  
IMBERT A, 1885, ARCH OPHTHAL PARIS, V5, P358
[10]   The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry [J].
Kotecha, A ;
White, ET ;
Shewry, JM ;
Garway-Heath, DF .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2005, 89 (12) :1572-1575