Application of triphasic pulses with adjustable phase amplitude ratio (PAR) for cochlear ECAP recording: I. Amplitude growth functions

被引:11
作者
Bahmer, Andreas [1 ]
Baumann, Uwe [1 ]
机构
[1] Univ Frankfurt Main, Clin Otolaryngol, D-60590 Frankfurt, Germany
关键词
ECAP; Biphasic pulses; Precision triphasic pulses; Polarity; Pulse shape; Artifact reduction; Amplitude growth function; COMPOUND ACTION-POTENTIALS; NERVE ACTION-POTENTIALS; BRAIN-STEM RESPONSES; ELECTRICAL-STIMULATION; STRATEGIES; ARTIFACT; ABR;
D O I
10.1016/j.jneumeth.2011.12.005
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This study describes the use of triphasic electrical stimulation pulses with an adjustable phase amplitude ratio (PAR) for the reduction of electrical stimulus artifacts. It is hypothesized that the setting of a certain PAR can facilitate a nearly artifact-free recording of electrically evoked compound action potentials (ECAP) in the cochlea. Artifact reduction with triphasic pulses using single epochs is expected to prevent latency or polarity effects, which are seen in standard forward masking or alternating polarity strategies. Although the application of a third phase is already implemented in implants manufactured by MED-EL (Zierhofer, 2003) and Cochlear (Sydney, Nucleus 5 System: van Dijk et al. (2007)) for the reduction of stimulation artifacts generated with these stimulators in ECAP measurements, an elaborate systematic evaluation of PAR for artifact reduction has not yet been conducted (compare evaluation for one subject Schoesser et al. (2001)). In the present paper, the effect of PAR variation on human ECAP recording and the feasibility of amplitude growth function recording with triphasic pulses and an optimized PAR are evaluated. Measurements were accomplished in five subjects, whereby more detailed test series were carried out in one subject. All subjects were implanted with devices from the company MED-EL, Innsbruck. A comparison of PAR optimized triphasic pulses was carried out against two other measurement techniques (biphasic alternating polarity stimulation and biphasic stimulation according to Miller) for apical, middle, and basal electrodes. ECAP thresholds were estimated by means of amplitude growth functions. However, recording of ECAP with triphasic pulses showed drawbacks: additional artifacts depending on stimulation and/or recording parameters are introduced, the ratio between the additional artifact and improved detectability of neural responses is dependent on PAR, and response thresholds obtained with triphasic pulses - although similar in shape - are in most cases substantially higher compared to thresholds measured with the Miller method. Higher thresholds most probably occur because the triphasic pulse patterns seem to less effectively stimulate neural structures compared to biphasic pulses since measured response thresholds are higher. For certain electrode groups threshold profiles obtained with triphasic pulses were found to be similar compared to stimulation with biphasic pulses. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:202 / 211
页数:10
相关论文
共 23 条
[1]   ELECTRICALLY EVOKED AUDITORY BRAIN-STEM RESPONSE - REFRACTORY PROPERTIES AND STRENGTH-DURATION FUNCTIONS [J].
ABBAS, PJ ;
BROWN, CJ .
HEARING RESEARCH, 1991, 51 (01) :139-148
[2]   Recording of electrically evoked auditory brainstem responses (E-ABR) with an integrated stimulus generator in Matlab [J].
Bahmer, Andreas ;
Peter, Otto ;
Baumann, Uwe .
JOURNAL OF NEUROSCIENCE METHODS, 2008, 173 (02) :306-314
[3]   Recording and analysis of electrically evoked compound action potentials (ECAPs) with MED-EL cochlear implants and different artifact reduction strategies in Mat lab [J].
Bahmer, Andreas ;
Peter, Otto ;
Baumann, Uwe .
JOURNAL OF NEUROSCIENCE METHODS, 2010, 191 (01) :66-74
[4]   Recording of electrically evoked auditory brainstem responses after electrical stimulation with biphasic, triphasic and precision triphasic pulses [J].
Bahmer, Andreas ;
Polak, Marek ;
Baumann, Uwe .
HEARING RESEARCH, 2010, 259 (1-2) :75-85
[5]  
Battmer R, 2005, 4 INT S WORKSH OBJ M, P21
[6]   ELECTRICALLY EVOKED WHOLE-NERVE ACTION-POTENTIALS - PARAMETRIC DATA FROM THE CAT [J].
BROWN, CJ ;
ABBAS, PJ .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1990, 88 (05) :2205-2210
[7]   ELECTRICALLY EVOKED WHOLE-NERVE ACTION-POTENTIALS - DATA FROM HUMAN COCHLEAR IMPLANT USERS [J].
BROWN, CJ ;
ABBAS, PJ ;
GANTZ, B .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1990, 88 (03) :1385-1391
[8]   ACOUSTICALLY DERIVED AUDITORY-NERVE ACTION-POTENTIAL EVOKED BY ELECTRICAL-STIMULATION - AN ESTIMATION OF THE WAVEFORM OF SINGLE UNIT CONTRIBUTION [J].
DESAUVAGE, RC ;
CAZALS, Y ;
ERRE, JP ;
ARAN, JM .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1983, 73 (02) :616-627
[9]  
Eddington DK, 2004, N01DC21001 NIH
[10]  
Kiefer J, 2001, AUDIOLOGY, V40, P32