Dual articulation retentive acetabular liners and wear: surface analysis of 40 retrieved polyethylene implants

被引:54
作者
Adam, P [1 ]
Farizon, F [1 ]
Fessy, MH [1 ]
机构
[1] CHU St Etienne, Ctr Orthoped & Traumatol, Hop Bellevue, F-42055 St Etienne, France
来源
REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR | 2005年 / 91卷 / 07期
关键词
replacement hip arthroplasty; polyethylene; wear;
D O I
10.1016/S0035-1040(05)84466-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose of the study The original concept of dual articulation retentive liners has been implemented in routine clinical practice for total hip arthroplasty since 1976. The retentive liner improves stability and the dual mobility favors preservation of joint motion. In this study, we analyzed wear of the two concave and convex surfaces in 40 retrieved polyethylene implants in order to search for factors favoring wear. Material and methods Forty polyethylene inserts were examined. These implants had been removed after mean implantation of eight years. Mean patient age at implantation was 48 years. The inserts were removed because of prosthesis infection or mechanical failure. Macroscopic examination was followed by a surface analysis with direct measurement of alterations of the curvature radii. Internal concavity was measured in the three dimensions using a 4-mm stylus (BNH 706). External convexity was measured by lateral projection. The estimated error was 5 urn for both measurement methods. Manufacturers' tolerances for these implants were approximately 50 pm. Linear and volumetric wear was determined by comparing the measured dimensions with the theoretical dimensions of new inserts. Results Macroscopically, all of the pieces studied had lost their initial striation on the convex surface; 40% of the pieces displayed visible wear of the retentive collar. Mean annual wear was 9 mu m (SD 9 mu m) for the convex surface and 73 mu m (SD 69 mu m) for the concave surface. Mean total wear, taken as the sum of the wears on the convex and concave surfaces was 82 pm (SD 72 pm). Volumetric wear was 28.9 mm(3)/yr for the convex surface (SD 27.6) and 25.5 for the concave surface (SD 23.2) giving a mean annual total volumetric wear of 54.3 mm(3)/yr (SD 39.6). Discussion Total wear observed for these 40 dual articulation liners which had functioned in vivo was to the same order as reported for the metal-polyethylene bearing with 22.2 mm femoral heads. Considering the wear on both the convex and concave surface, the dual articulation was not associated with increased wear compared with the conventional metal-polyethylene bearings, but with the advantages of retention and greater stability. Conclusion Use of a dual articulation acetabular liner is an attractive solution when a metal-polyethylene bearing is considered. The benefit in terms of joint stability does not increase wear.
引用
收藏
页码:627 / 636
页数:10
相关论文
共 16 条
[1]  
Anderson M J, 1994, J Arthroplasty, V9, P17, DOI 10.1016/0883-5403(94)90133-3
[2]  
Aubriot J H, 1993, Acta Orthop Belg, V59 Suppl 1, P267
[3]  
Bousquet G, 1985, J Orthop Surg Tech, V1, P15
[4]  
Devane PA, 1995, CLIN ORTHOP RELAT R, V319, P303
[5]   Results with a cementless alumina-coated cup with dual mobility - A twelve-year follow-up study [J].
Farizon, F ;
de Lavison, R ;
Azoulai, JJ ;
Bousquet, G .
INTERNATIONAL ORTHOPAEDICS, 1998, 22 (04) :219-224
[6]  
GRIGORIS P, 1994, CLIN ORTHOP RELAT R, P148
[7]   POLYETHYLENE AND METAL DEBRIS GENERATED BY NON-ARTICULATING SURFACES OF MODULAR ACETABULAR COMPONENTS [J].
HUK, OL ;
MANJULA, B ;
BETTS, F ;
RIMNAC, CM ;
LIEBERMAN, JR ;
HUO, MH ;
SALVATI, EA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (04) :568-574
[8]  
Katayama K, 2001, J Orthop Sci, V6, P503, DOI 10.1007/s007760100005
[9]   Wear of bipolar hip prostheses [J].
Kusaba, A ;
Kuroki, Y .
JOURNAL OF ARTHROPLASTY, 1998, 13 (06) :668-673
[10]  
LECLERCQ S, 1995, REV CHIR ORTHOP, V81, P389