Two-Center Experience in the Endovascular Treatment of Ruptured and Unruptured Intracranial Aneurysms Using the WEB Device: A Retrospective Analysis

被引:55
作者
Popielski, J. [1 ]
Berlis, A. [2 ]
Weber, W. [1 ]
Fischer, S. [1 ]
机构
[1] Knappschaftskrankenhaus Bochum Langendreer, Univ Klin, Inst Diagnost & Intervent Radiol, Neuroradiol,Nukl Med, Schornau 23-25, D-44892 Bochum, Germany
[2] Klinikum Augsburg, Klin Diagnost Radiol & Neuroradiol, Augsburg, Germany
关键词
CEREBRAL ANEURYSMS; WIDE-NECK; BIFURCATION ANEURYSMS; FOLLOW-UP; METAANALYSIS; COMPLICATIONS; COILING; MULTICENTER; OCCLUSION; OUTCOMES;
D O I
10.3174/ajnr.A5413
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: The safety and efficacy of the Woven EndoBridge (WEB) device for the treatment of cerebral aneurysms has been investigated in several studies. Our objective was to report the experience of 2 neurovascular centers with the WEB device in the treatment of broad-based intracranial aneurysms, including the technical feasibility and safety as well as short- and midterm angiographic and clinical follow-up-results. MATERIALS AND METHODS: We performed a retrospective analysis of all ruptured and unruptured aneurysms treated with a WEB device (WEB Single-Layer and Single-Layer Sphere) between August 2014 and February 2017. Primary outcome measures included the feasibility of implantation and the angiographic outcome. Secondary outcome measures included the clinical outcome at discharge and procedural complications. RESULTS: One hundred two aneurysms in 101 patients, including 37 (36.3%) ruptured aneurysms, were treated with the WEB device. Implantation was successful in 98 (96.1%) aneurysms. Additional devices (stents/coils) were necessary in 15.3% (15/98) of aneurysms. Procedural complications occurred in 4.9% (5/102). Of these, 4 were thromboembolic events and 1 was an intraprocedural rupture. Angiographic follow-up at 3 and 12 months was available for 79.6% (78/98) and 50.0% (49/98) of all aneurysms to date, respectively, showing a sufficient aneurysm occlusion in 80.7% (63/78) at 3 months and 77.6% (38/49) at 12 months. Delayed aneurysm ruptures have not been observed during the follow-up period to date. CONCLUSIONS: The WEB device offers a safe and effective treatment option for broad-based intracranial aneurysms without the need for dual antiplatelet therapy.
引用
收藏
页码:111 / 117
页数:7
相关论文
共 28 条
[1]  
Adeeb N, 2017, NEUROSURGERY, DOI [10. 1093/neuros/nyx310 28595331, DOI 10.1093/NEUROS/NYX31028595331]
[2]   Endovascular Treatment of Intracranial Aneurysms with the WEB Device: A Systematic Review of Clinical Outcomes [J].
Armoiry, X. ;
Turjman, F. ;
Hartmann, D. J. ;
Sivan-Hoffmann, R. ;
Riva, R. ;
Labeyrie, P. E. ;
Aulagner, G. ;
Gory, B. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2016, 37 (05) :868-872
[3]   Efficacy and Safety of the Woven EndoBridge (WEB) Device for the Treatment of Intracranial Aneurysms: A Systematic Review and Meta-Analysis [J].
Asnafi, S. ;
Rouchaud, A. ;
Pierot, L. ;
Brinjikji, W. ;
Murad, M. H. ;
Kallmes, D. F. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2016, 37 (12) :2287-2292
[4]   Rescue Treatment of Thromboembolic Complications during Endovascular Treatment of Cerebral Aneurysms: A Meta-Analysis [J].
Brinjikji, W. ;
Morales-Valero, S. F. ;
Murad, M. H. ;
Cloft, H. J. ;
Kallmes, D. F. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2015, 36 (01) :121-125
[5]   Single-Layer WEBs: lntrasaccular Flow Disrupters for Aneurysm Treatment-Feasibility Results from a European Study [J].
Caroff, J. ;
Mihalea, C. ;
Klisch, J. ;
Strasilla, C. ;
Berlis, A. ;
Patankar, T. ;
Weber, W. ;
Behme, D. ;
Jacobsen, E. A. ;
Liebig, T. ;
Prothmann, S. ;
Cognard, C. ;
Finkenzeller, T. ;
Moret, J. ;
Spelle, L. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2015, 36 (10) :1942-1946
[6]   Woven Endobridge (WEB) Device for endovascular treatment of ruptured intracranial wide-neck aneurysms: a single-center experience [J].
Caroff, Jildaz ;
Mihalea, Cristian ;
Dargento, Francesco ;
Neki, Hiroaki ;
Ikka, Leon ;
Benachour, Nidhal ;
Moret, Jacques ;
Spelle, Laurent .
NEURORADIOLOGY, 2014, 56 (09) :755-761
[7]   Initial and mid-term results from 108 consecutive patients with cerebral aneurysms treated with the WEB device [J].
Clajus, Christin ;
Strasilla, Christoph ;
Fiebig, Tom ;
Sychra, Vojtech ;
Fiorella, David ;
Klisch, Joachim .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2017, 9 (04) :411-417
[8]   Clinical outcome after surgical clipping or endovascular coiling for cerebral aneurysms: a pragmatic meta- analysis of randomized and non- randomized trials with short- and long- term follow- up [J].
Delgado, Alberto Falk ;
Andersson, Tommy ;
Delgado, Anna Falk .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2017, 9 (03) :264-+
[9]  
Fiorella D, 2017, J NEUROINTERV SURG, DOI [10. 1136/neurintsurg-2016-012841 28096478, DOI 10.1136/NEURINTSURG-2016-01284128096478]
[10]   Interobserver variability in the assessment of aneurysm occlusion with the WEB aneurysm embolization system [J].
Fiorella, David ;
Arthur, Adam ;
Byrne, James ;
Pierot, Laurent ;
Molyneux, Andy ;
Duckwiler, Gary ;
McCarthy, Thomas ;
Strother, Charles .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2015, 7 (08) :591-595