Treatment precision of image-guided liver SBRT using implanted fiducial markers depends on marker-tumour distance

被引:94
作者
Seppenwoolde, Y. [1 ]
Wunderink, W. [1 ]
Wunderink-van Veen, S. R. [2 ,3 ]
Storchi, P. [1 ]
Romero, A. Mendez [1 ]
Heijmen, B. J. M. [1 ]
机构
[1] Erasmus MC, Dept Radiat Oncol, Daniel den Hoed Canc Ctr, NL-3075 EA Rotterdam, Netherlands
[2] Erasmus Univ, NL-3076 RE Rotterdam, Netherlands
[3] Delft Univ Technol, GEON Res Grp Behav Econ, NL-3076 RE Rotterdam, Netherlands
关键词
STEREOTACTIC BODY RADIOTHERAPY; BEAM COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; GEOMETRICAL UNCERTAINTIES; ABDOMINAL COMPRESSION; SETUP ERRORS; ORGAN MOTION; ACCURACY; CANCER; REDUCTION;
D O I
10.1088/0031-9155/56/17/001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study is to assess the accuracy of day-to-day predictions of liver tumour position using implanted gold markers as surrogates and to compare the method with alternative set-up strategies, i.e. no correction, vertebrae and 3D diaphragm-based set-up. Twenty patients undergoing stereotactic body radiation therapy (SBRT) with abdominal compression for primary or metastatic liver cancer were analysed. We determined the day-today correlation between gold marker and tumour positions in contrast-enhanced CT scans acquired at treatment preparation and before each treatment session. The influence of marker-tumour distance on the accuracy of prediction was estimated by introducing a method extension of the set-up error paradigm. The distance between gold markers and the centre of the tumour varied between 5 and 96 mm. Marker-guidance was superior to guiding treatment using other surrogates, although both the random and systematic components of the prediction error SD depended on the tumour-marker distance. For a marker-tumour distance of 4 cm, we observed sigma = 1.3 mm and Sigma = 1.6 mm. The 3D position of the diaphragm dome was the second best predictor. In conclusion, the tumour position can be predicted accurately using implanted markers, but marker-guided set-up accuracy decreases with increasing distance between implanted markers and the tumour.
引用
收藏
页码:5445 / 5468
页数:24
相关论文
共 34 条
[1]   Evaluating the influence of setup uncertainties on treatment planning for focal liver tumors [J].
Balter, JM ;
Brock, KK ;
Lam, KL ;
Tatro, D ;
Dawson, LA ;
McShan, D ;
Ten Haken, RK .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02) :610-614
[2]   Determination of ventilatory liver movement via radiographic evaluation of diaphragm position [J].
Balter, JM ;
Dawson, LA ;
Kazanjian, S ;
McGinn, C ;
Brock, KK ;
Lawrence, T ;
Ten Haken, R .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (01) :267-270
[3]   A novel method for estimating SBRT delivered dose with beam's-eye-view images [J].
Berbeco, Ross I. ;
Hacker, Fred ;
Zatwarnicki, Chris ;
Park, Sang-June ;
Ionascu, Dan ;
O'Farrell, Desmond ;
Mamon, Harvey J. .
MEDICAL PHYSICS, 2008, 35 (07) :3225-3231
[4]   Clinical feasibility of using an epid in cine mode for image-guided verification of stereotactic body radiotherapy [J].
Berbeco, Ross I. ;
Hacker, Fred ;
Ionascu, Dan ;
Mamon, Harvey J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (01) :258-266
[5]   MAXIMIZING SETUP ACCURACY USING PORTAL IMAGES AS APPLIED TO A CONFORMAL BOOST TECHNIQUE FOR PROSTATIC-CANCER [J].
BIJHOLD, J ;
LEBESQUE, JV ;
HART, AAM ;
VIJLBRIEF, RE .
RADIOTHERAPY AND ONCOLOGY, 1992, 24 (04) :261-271
[6]  
Case R, 2007, RADIOTHER ONCOL, V84, pS38
[7]   INTER- AND INTRAFRACTION VARIABILITY IN LIVER POSITION IN NON-BREATH-HOLD STEREOTACTIC BODY RADIOTHERAPY [J].
Case, Robert B. ;
Sonke, Jan-Jakob ;
Moseley, Douglas J. ;
Kim, John ;
Brock, Kristy K. ;
Dawson, Laura A. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (01) :302-308
[8]   Needle tract implantation after percutaneous interventional procedures in hepatocellular carcinomas: Lessons learned from a 10-year experience [J].
Chang, Samuel ;
Kim, Seong Hyun ;
Lim, Hyo K. ;
Kim, Seung Hoon ;
Lee, Won Jae ;
Choi, Dongil ;
Kim, Young-Sun ;
Rhim, Hyunchul .
KOREAN JOURNAL OF RADIOLOGY, 2008, 9 (03) :268-274
[9]   Accuracy of daily image guidance for hypofractionated liver radiotherapy with active breathing control [J].
Dawson, LA ;
Eccles, C ;
Bissonnette, JP ;
Brock, KK .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (04) :1247-1252
[10]   A protocol for the reduction of systematic patient setup errors with minimal portal imaging workload [J].
De Boer, HCJ ;
Heijmen, BJM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 50 (05) :1350-1365