A new dead-time determination method for gamma-ray detectors using attenuation law

被引:3
作者
Akyurek, T. [1 ]
机构
[1] Marmara Univ, Fac Art & Sci, Dept Phys, TR-34722 Istanbul, Turkey
关键词
dead-time; non-paralyzing model; paralyzing model; DMRAL method;
D O I
10.1016/j.net.2021.06.038
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
This study presents a new dead-time measurement method using the gamma attenuation law and generalized dead-time models for nuclear gamma-ray detectors. The dead-time of the Nal(TI) detection system was obtained to validate the new dead-time determination method using very thin lead and polyethylene absorbers. Non-paralyzing dead-time was found to be 8.39 mu s, and paralyzing dead-time was found to be 8.35 mu s using lead absorber for Nal(TI) scintillator detection system. These dead-time values are consistent with the previously reported dead-time values for scintillator detection systems. The gamma build-up factor's contribution to the dead-time was neglected because a very thin material was used. (C) 2021 Korean Nuclear Society, Published by Elsevier Korea LLC.
引用
收藏
页码:4093 / 4097
页数:5
相关论文