On the registration of low energy alpha particle with modified GafChromic EBT2 radiochromic film

被引:3
作者
El-Naggar, Hala I. [1 ]
Ghanim, E. H. [2 ]
El Ghazaly, M. [3 ]
Salama, T. T. [3 ]
机构
[1] Benha Univ, Dept Phys, Fac Sci, Banha 13518, Egypt
[2] Beni Suef Univ, Fac Ind Educ, Dept Basic Sci, Bani Suwayf, Egypt
[3] Zagazig Univ, Dept Phys, Fac Sci, Zagazig 44519, Egypt
关键词
GafChromic EBT2 radiochromic film; Modified EBT2-M radiochromic film; Alpha particle; Flatbed scanner; 8-Bit RGB image; Absorbance; RGB Color value; DOSE-RESPONSE; DOSIMETRY; DEPENDENCE; OPTIMIZATION;
D O I
10.1016/j.radphyschem.2021.109852
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The objectives of the current work are to investigate the modification of the self-developed GafChromic EBT2 radiochromic film, to register 2 MeV alpha particles, by eliminating the polystyrene layer of 50 mu m and the adhesive layer of 15 mu m. Such modification enables 2 MeV alpha particles to reach and interact directly by the diacetylene monomers active layer of thickness 30 mu m. The modified EBT2 radiochromic film, EBT2-M was irradiated by alpha-particles of 2 MeV starting from 1 h to 300 h in a clean dark room; one day later the irradiated EBT2 films were scanned with a flatbed scanner, and the absorbance spectra were measured by UV-VIS spectrophotometer. The obtained results reveal that the darkness of the irradiated EBT2-M film increases with the increasing of the exposure time to 2 MeV alpha particles, but for a prolonged exposure time the darkness is saturated. The UV-VIS absorbance spectra at 502 nm, 525 nm, 582 nm, 602 nm, and 631 nm are correlated with exposure time to 2 MeV alpha particles, which is, in turn, the absorbed dose by an exponential growth function that saturates at a prolonged exposure time, when diacetylene monomers are completely polymerized. The color channel level Red, Green, and Blue resulting from split RGB image was determined and investigated as a function in the exposure time to the alpha particle of 2 MeV, the color scale level decreases exponentially with the increase in exposure time. The reported results of the grey, red and green color scale have similar absorbance whereas the blue color channel shows insignificant behavior. The presented results pave the way to use the modified EBT2-M radiochromic film to mapping the relative dose delivered by low-energy alpha particles.
引用
收藏
页数:6
相关论文
共 24 条
  • [1] Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies
    Arjomandy, Bijan
    Tailor, Ramesh
    Anand, Aman
    Sahoo, Narayan
    Gillin, Michael
    Prado, Karl
    Vicic, Milos
    [J]. MEDICAL PHYSICS, 2010, 37 (05) : 1942 - 1947
  • [2] Investigation of energy dependance for EBT3 response to irradiation with alpha beams
    Basset, A.
    Koumeir, C.
    Chiavassa, S.
    Delpon, G.
    Le Deroff, C.
    Frelin, A. -M.
    Guertin, A.
    Haddad, F.
    Metivier, V.
    Michel, N.
    Poirier, F.
    Servagent, N.
    Varmenot, N.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2019, 454 : 56 - 60
  • [3] Dosimetric response of radiochromic films to protons of low energies in the Bragg peak region
    Battaglia, M. C.
    Schardt, D.
    Espino, J. M.
    Gallardo, M. I.
    Cortes-Giraldo, M. A.
    Quesada, J. M.
    Lallena, A. M.
    Miras, H.
    Guirado, D.
    [J]. PHYSICAL REVIEW ACCELERATORS AND BEAMS, 2016, 19 (06):
  • [4] Dose and absorption spectra response of EBT2 Gafchromic film to high energy x-rays
    Butson, M. J.
    Cheung, T.
    Yu, P. K. N.
    Alnawaf, H.
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2009, 32 (04) : 196 - 202
  • [5] Measurement of energy dependence for XRCT radiochromic film
    Butson, Martin J.
    Cheung, Tsang
    Yu, Peter K. N.
    [J]. MEDICAL PHYSICS, 2006, 33 (08) : 2923 - 2925
  • [6] Energy response of the new EBT2 radiochromic film to x-ray radiation
    Butson, Martin J.
    Yu, Peter K. N.
    Cheung, Tsang
    Alnawaf, Hani
    [J]. RADIATION MEASUREMENTS, 2010, 45 (07) : 836 - 839
  • [7] Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp X-ray range
    Butson, MJ
    Cheung, T
    Yu, PKN
    [J]. APPLIED RADIATION AND ISOTOPES, 2006, 64 (01) : 60 - 62
  • [8] Visible absorption properties of radiation exposed XR type-T radiochromic film
    Butson, MJ
    Cheung, T
    Yu, PKN
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (19) : N347 - N351
  • [9] Radiochromic film for medical radiation dosimetry
    Butson, MJ
    Yu, PKN
    Cheung, T
    Metcalfe, P
    [J]. MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2003, 41 (3-5) : 61 - 120
  • [10] Precise radiochromic film dosimetry using a flat-bed document scanner
    Devic, S
    Seuntjens, J
    Sham, E
    Podgorsak, EB
    Schmidtlein, CR
    Kirov, AS
    Soares, CG
    [J]. MEDICAL PHYSICS, 2005, 32 (07) : 2245 - 2253