Quantification of fractional flow reserve based on angiographic image data

被引:18
|
作者
Wong, Jerry T. [1 ]
Le, Huy [1 ]
Suh, William M. [2 ]
Chalyan, David A. [1 ]
Mehraien, Toufan [1 ]
Kern, Morton J. [2 ]
Kassab, Ghassan S. [3 ]
Molloi, Sabee [1 ]
机构
[1] Univ Calif Irvine, Dept Radiol Sci, Irvine, CA 92697 USA
[2] Univ Calif Irvine, Div Cardiol, Dept Med, Irvine, CA 92868 USA
[3] Indiana Purdue Univ, Dept Biomed Engn, Indianapolis, IN 46202 USA
来源
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING | 2012年 / 28卷 / 01期
关键词
Angiography; Blood flow; Blood volume; Regional blood flow; Stenosis; CORONARY BLOOD-FLOW; DIGITAL-SUBTRACTION-ANGIOGRAPHY; PRESSURE SLOPE INDEX; IN-VIVO VALIDATION; ARTERIAL-PRESSURE; SCALING LAWS; MAXIMUM VASODILATION; HYPEREMIC FLOW; LUMEN VOLUME; HEART-RATE;
D O I
10.1007/s10554-010-9767-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Coronary angiography provides excellent visualization of coronary arteries, but has limitations in assessing the clinical significance of a coronary stenosis. Fractional flow reserve (FFR) has been shown to be reliable in discerning stenoses responsible for inducible ischemia. The purpose of this study is to validate a technique for FFR quantification using angiographic image data. The study was carried out on 10 anesthetized, closed-chest swine using angioplasty balloon catheters to produce partial occlusion. Angiography based FFR was calculated from an angiographically measured ratio of coronary blood flow to arterial lumen volume. Pressure-based FFR was measured from a ratio of distal coronary pressure to aortic pressure. Pressure-wire measurements of FFR (FFR (P) ) correlated linearly with angiographic volume-derived measurements of FFR (FFR (V) ) according to the equation: FFR (P) = 0.41 FFR (V) + 0.52 (P-value < 0.001). The correlation coefficient and standard error of estimate were 0.85 and 0.07, respectively. This is the first study to provide an angiographic method to quantify FFR in swine. Angiographic FFR can potentially provide an assessment of the physiological severity of a coronary stenosis during routine diagnostic cardiac catheterization without a need to cross a stenosis with a pressure-wire.
引用
收藏
页码:13 / 22
页数:10
相关论文
共 50 条
  • [31] Clinical and angiographic predictors of persistently ischemic fractional flow reserve after percutaneous revascularization
    Agarwal, Shiv Kumar
    Kasula, Srikanth
    Almomani, Ahmed
    Hacioglu, Yalcin
    Ahmed, Zubair
    Uretsky, Barry F.
    Hakeem, Abdul
    AMERICAN HEART JOURNAL, 2017, 184 : 10 - 16
  • [32] Angiographic Severity Does Not Correlate with Fractional Flow Reserve in Heavily Calcified Coronary Arteries
    Johnson, Paul M.
    Madamanchi, Chaitanya
    Sharalaya, Zarina M.
    Iqbal, Zahra
    Gehi, Anil K.
    Kaul, Prashant
    Stouffer, George A.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2017, 89 (02) : 226 - 232
  • [33] Diagnostic accuracy of two-dimensional coronary angiographic derived fractional flow reserve
    Sadeghipour, P.
    Babakhani, H.
    Abdi, S.
    Ghasemi, M.
    Moosavi, J.
    Sadeghian, M.
    Shafe, O.
    Mohebbi, B.
    EUROPEAN HEART JOURNAL, 2020, 41 : 1268 - 1268
  • [34] Clinical and angiographic predictors of discordance between myocardial perfusion scintigraphy and fractional flow reserve
    Mendes, L. C. M.
    Silva, M. V.
    Ambiel, M. L. B. R.
    Lacerda Filho, S. L.
    Reis, H. R.
    Araujo, E. C.
    Barbosa, C. J. D. G.
    EUROPEAN HEART JOURNAL, 2018, 39 : 977 - 977
  • [35] Estimating blood flow velocity in angiographic image data
    Hentschke, Clemens M.
    Serowy, Steffen
    Janiga, Gabor
    Rose, Georg
    Toennies, Klaus D.
    MEDICAL IMAGING 2011: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND MODELING, 2011, 7964
  • [36] FRACTIONAL FLOW RESERVE
    Garcia, J. A.
    CARDIOLOGY, 2014, 128 : 33 - 33
  • [37] Fractional flow reserve
    Yadava O.P.
    Casselman F.
    Indian Journal of Thoracic and Cardiovascular Surgery, 2018, 34 (4) : 533 - 533
  • [38] The imortance of hyperemic contrast velocity assessment on image-based fractional flow reserve calculation
    Tar, B.
    Uveges, A.
    Koszegi, Z.
    EUROPEAN HEART JOURNAL, 2020, 41 : 318 - 318
  • [39] Non-invasive Angiographic-based Fractional Flow Reserve: Technical Development, Clinical Implications, and Future Perspectives
    Joyce Peper
    Michiel L. Bots
    Tim Leiner
    Martin J. Swaans
    Current Medical Science, 2023, 43 : 423 - 433
  • [40] Non-invasive Angiographic-based Fractional Flow Reserve: Technical Development, Clinical Implications, and Future Perspectives
    Peper, Joyce
    Bots, Michiel L.
    Leiner, Tim
    Swaans, Martin J.
    CURRENT MEDICAL SCIENCE, 2023, 43 (03) : 423 - 433