Recurrent dislocation of a posterior-stabilized prosthesis: a series of three cases

被引:9
作者
Gidwani, S [1 ]
Langkamer, VG [1 ]
机构
[1] Southmead Gen Hosp, Avon Orthopaed Ctr, Bristol, Avon, England
关键词
dislocation; knee arthroplasty; complications; prosthesis design;
D O I
10.1016/S0968-0160(01)00102-8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report three cases of recurrent posterior dislocation of primary posterior-stabilized total knee replacements. In all of the cases, reduction was not possible under sedation alone, and general anaesthesia was needed to allow disengagement of the components and reduction of the dislocation. In addition, it was noted in all cases that the ability of the patients to flex their knee replacement excessively contributed to the instability. The design of the prosthesis, and issues of operative technique are discussed with reference to this serious complication. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:317 / 320
页数:4
相关论文
共 10 条
[1]  
BARGREN JH, 1980, CLIN ORTHOP RELAT R, P271
[2]  
Galinat B J, 1988, J Arthroplasty, V3, P363, DOI 10.1016/S0883-5403(88)80039-1
[3]  
GEBHARD JS, 1990, CLIN ORTHOP RELAT R, P225
[4]   TOTAL CONDYLAR KNEE PROSTHESIS - REPORT OF 220 CASES [J].
INSALL, J ;
SCOTT, WN ;
RANAWAT, CS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1979, 61 (02) :173-180
[5]  
KADOYA Y, 2000, P COMB C BRIT JAP OR
[6]  
LONBARDI AV, 1993, J ARTHROPLASTY, V8, P633
[7]  
Mihalko WM, 1999, CLIN ORTHOP RELAT R, P243
[8]  
Ochsner J L Jr, 1996, Am J Orthop (Belle Mead NJ), V25, P310
[9]  
SHARKEY PF, 1992, CLIN ORTHOP RELAT R, P128
[10]   Dislocation of total knee arthroplasty - A report of 6 cases with 2 patterns of instability [J].
Wang, CJ ;
Wang, HE .
ACTA ORTHOPAEDICA SCANDINAVICA, 1997, 68 (03) :282-285