Total nucleated cell differential for blood and bone marrow using a single tube in a five-color flow cytometer

被引:46
作者
Bjornsson, Sven [1 ]
Wahlstrom, Saga [1 ]
Norstrom, Eva [1 ]
Bernevi, Ingela [1 ]
O'Neill, Ulla [1 ]
Johansson, Eva [1 ]
Runstrom, Hakan [1 ]
Simonsson, Per [1 ]
机构
[1] Univ Hosp MAS, Dept Clin Chem, S-20502 Maimo, Sweden
关键词
leukocyte differential; total nucleated count; peripheral blood; bone marrow; cell cycle; ploidy; nuclear stain; DRAQ5;
D O I
10.1002/cyto.b.20382
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Flow cytometry allows the use of several antibodies in addition to light scatter, and most flow cytometers will provide at least seven measurements on each cell passing through the laser beam. A skilled microscopist will classify at least 14 cell classes in bone marrow or blood. Our goal was to use the seven parameters available in our flow cytometer to provide a reliable differential count using only one tube. Methods: Peripheral blood samples were analyzed on the Beckman Coulter LH750 cell counter, and the flagging and messages from the cell counter were used to select normal or pathological samples. Samples without flags (N = 50), with >2% erythroblasts (N = 80), or with "Blast" or "Verify diff" flags (N = 54) were investigated. We used a lyse-no-wash method to ensure minimal loss of fragile cells with live gating on DRAQ5-positive cells to acquire only nucleated cells. The FL-1 to FL-4 channels were used for the antibodies CD36-FITC, CD203-PE, CD138-PE, CD45-ECD, CD16-Pcy5, and CD56-Pcy5. FL-5 was used for the DNA-stain DRAQ5. Results: Using live gate acquisition on DRAQ5, we were able to classify total nucleated cells into 10 classes. We were unable to identify megakaryocytes, but platelets could be studied by rerunning the sample after dilution and gating on DRAQ5-negative CD36-posive events. Validation against digitized microscopy and cell counter showed linear correlations within each cell class with correlation coefficients that seem reasonable for cellular classification. The lowest correlation was found for basophil granulocytes. Flow cytometry detected twice as many immature neutrophils compared to microscopy. Conclusions: We have designed a one-tube immunophenotyping panel for classification of total nucleated cells and platelets in blood or bone marrow. The seven parameters available in one single tube in our cytometer seem to be enough for reliable differential count even in difficult pathological samples. The analytical imprecision of the flow cytometer differential was much lower than that obtained with microscopy or cell counter differentials. (C) 2007 clinical Cytometry Society
引用
收藏
页码:91 / 103
页数:13
相关论文
共 40 条
  • [21] RUMKE CL, 1985, BLOOD CELLS, V11, P311
  • [22] Schlenke P, 1998, CYTOMETRY, V33, P310, DOI 10.1002/(SICI)1097-0320(19981101)33:3<310::AID-CYTO4>3.0.CO
  • [23] 2-K
  • [24] Smith PJ, 2000, CYTOMETRY, V40, P280, DOI 10.1002/1097-0320(20000801)40:4<280::AID-CYTO4>3.0.CO
  • [25] 2-7
  • [26] Differential counting of blood leukocytes using automated microscopy and a decision support system based on artificial neural networks -: evaluation of DiffMasterTM Octavia
    Swolin, B
    Simonsson, P
    Backman, S
    Löfqvist, I
    Bredin, I
    Johnsson, M
    [J]. CLINICAL AND LABORATORY HAEMATOLOGY, 2003, 25 (03): : 139 - 147
  • [27] A RAPID SAMPLE PREPARATION TECHNIQUE FOR FLOW CYTOMETRIC ANALYSIS OF IMMUNOFLUORESCENCE ALLOWING ABSOLUTE ENUMERATION OF CELL SUBPOPULATIONS
    TERSTAPPEN, LWMM
    MEINERS, H
    LOKEN, MR
    [J]. JOURNAL OF IMMUNOLOGICAL METHODS, 1989, 123 (01) : 103 - 112
  • [28] TERSTAPPEN LWMM, 1990, LEUKEMIA, V4, P657
  • [29] TERSTAPPEN LWMM, 1991, BLOOD CELLS, V17, P585
  • [30] TERSTAPPEN LWMM, 1992, BLOOD CELLS, V18, P311