Comparison of the OcclutechA® FigullaA® Septal Occluder and AmplatzerA® Septal Occluder for Atrial Septal Defect Device Closure

被引:27
作者
Roymanee, Supaporn [1 ]
Promphan, Worakan [2 ]
Tonklang, Nakharin [1 ]
Wongwaitaweewong, Kanjarut [1 ]
机构
[1] Prince Songkla Univ, Fac Med, Dept Pediat, Pediat Cardiol Unit, Hat Yai, Songkhla, Thailand
[2] Rangsit Univ, Coll Med, Pediat Cardiac Ctr, Queen Sirikit Natl Inst Child Hlth, Bangkok, Thailand
关键词
ASD device closure; ASD percutaneous closure; Occlutech septal occluder; Amplatzer septal occluder; TERM-FOLLOW-UP; TRANSCATHETER CLOSURE; ASD OCCLUDER; PERCUTANEOUS CLOSURE; EXPERIENCE; CHILDREN; ADULTS;
D O I
10.1007/s00246-015-1103-y
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The Occlutech(A (R)) Figulla(A (R)) septal occluder (OFSO) is a later-generation double-disk device with few reports of its success rates and complications compared with the Amplatzer(A (R)) septal occluder (ASO), which is the worldwide standard device in percutaneous atrial septal defect (ASD) closure. We recruited and compared the results in 149 patients (76.5 % female) who underwent ASD device closure in our center between January 2003 and June 2012. The patients ranged in age from 2.3 to 77.2 years. There were no statistically significant differences between the two groups regarding patient baseline characteristics and procedure variables. The success rate using either device was excellent (ASO 94.4 % and OFSO 97.4 %; p = 0.43). Although the diameter of the ASD and the pulmonary arterial pressure in the OFSO group were slightly higher than in the ASO group, the median fluoroscopic time in the OFSO group was significantly shorter (ASO 13.7 min; OFSO 9.0 min; p < 0.001). The overall median follow-up time was 3.6 years (interquartile range 2.1-9.0 years). There were no significant differences between the major and minor complications when comparing the two devices. Both devices were safe and effective for percutaneous ASD closures. The OFSO had the benefit of a shorter fluoroscopic time.
引用
收藏
页码:935 / 941
页数:7
相关论文
共 24 条
  • [21] Percutaneous Atrial Shunt Closure Using the Novel Occlutech Figulla Device: 6-Month Efficacy and Safety
    Van den Branden, Ben J. L.
    Post, Martijn C.
    Plokker, Herbert W. M.
    Ten Berg, Jurrien M.
    Suttorp, Maarten J.
    [J]. JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2011, 24 (03) : 264 - 270
  • [22] Transcatheter Closure of Atrial Septal Defects in Children, Middle-Aged Adults, and Older Adults: Failure Rates, Early Complications; and Balloon Sizing Effects
    Vijarnsorn, Chodchanok
    Durongpisitkul, Kritvikrom
    Chanthong, Prakul
    Chungsomprasong, Paweena
    Soongswang, Jarupim
    Loahaprasitiporn, Duangmanee
    Nana, Apichart
    [J]. CARDIOLOGY RESEARCH AND PRACTICE, 2012, 2012
  • [23] Percutaneous ASD Closure in a Large Australian Series: Short- and Long-Term Outcomes
    Walters, D. L.
    Boga, T.
    Burstow, D.
    Scalia, G.
    Hourigan, L. A.
    Aroney, C. N.
    [J]. HEART LUNG AND CIRCULATION, 2012, 21 (09) : 572 - 575
  • [24] Transcatheter closure of atrial septal defect without balloon sizing
    Wang, Jou-Kou
    Tsai, Shen-Kou
    Lin, Shu-Man
    Chiu, Shuenn-Nan
    Lin, Ming-Tai
    Wu, Mei-Hwan
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2008, 71 (02) : 214 - 221