Quantitative performance of a quadrupole-orbitrap-MS in targeted LC-MS determinations of small molecules

被引:60
作者
Grund, Baptiste [1 ]
Marvin, Laure [2 ]
Rochat, Bertrand [1 ]
机构
[1] CHU Vaudois, Univ Hosp Lausanne, Quantitat Mass Spectrometry Facil, CH-1011 Lausanne, Switzerland
[2] Serv Consommat & Affaires Vet, Chemin Boveresses 155, Epalinges, Switzerland
关键词
Accurate-mass; Drugs; High-resolution mass-spectrometry; LC-MS; Orbitrap; Q-Exactive; Time-of-flight; Plasma; Quantification; Triple-quadrupole; Metabotyping; RESOLUTION MASS-SPECTROMETRY; LIQUID-CHROMATOGRAPHY; TRIPLE QUADRUPOLE; CLINICAL LABORATORIES; MATRIX INTERFERENCE; RESIDUE ANALYSIS; RESOLVING POWER; DRUG DISCOVERY; HUMAN PLASMA; QUANTIFICATION;
D O I
10.1016/j.jpba.2016.02.025
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
High-resolution mass spectrometry (HRMS) has been associated with qualitative and research analysis and QQQ-MS with quantitative and routine analysis. This view is now challenged and for this reason, we have evaluated the quantitative LC-MS performance of a new high-resolution mass spectrometer (HRMS), a Q-orbitrap-MS, and compared the results obtained with a recent triple-quadrupole MS (QQQ-MS). High-resolution full-scan (HR-FS) and MS/MS acquisitions have been tested with real plasma extracts or pure standards. Limits of detection, dynamic range, mass accuracy and false positive or false negative detections have been determined or investigated with protease inhibitors, tyrosine kinase inhibitors, steroids and metanephrines. Our quantitative results show that today's available HRMS are reliable and sensitive quantitative instruments and comparable to QQQ-MS quantitative performance. Taking into account their versatility, user-friendliness and robustness, we believe that HRMS should be seen more and more as key instruments in quantitative LC-MS analyses. In this scenario, most targeted LC-HRMS analyses should be performed by HR-FS recording virtually "all" ions. In addition to absolute quantifications, HR-FS will allow the relative quantifications of hundreds of metabolites in plasma revealing individual's metabolome and exposome. This phenotyping of known metabolites should promote HRMS in clinical environment. A few other LC-HRMS analyses should be performed in single-ion-monitoring or MS/MS mode when increased sensitivity and/or detection selectivity will be necessary. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:48 / 56
页数:9
相关论文
共 34 条
[1]   Analysis and quantification of vitamin D metabolites in serum by ultra-performance liquid chromatography coupled to tandem mass spectrometry and high-resolution mass spectrometry - a method comparison and validation [J].
Bruce, Stephen J. ;
Rochat, Bertrand ;
Beguin, Alexandre ;
Pesse, Benoit ;
Guessous, Idris ;
Boulat, Olivier ;
Henry, Hugues .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2013, 27 (01) :200-206
[2]   Comparison of Different Time of Flight-Mass Spectrometry Modes for Small Molecule Quantitative Analysis [J].
Chindarkar, Nandkishor S. ;
Park, Hyung-Doo ;
Stone, Judith A. ;
Fitzgerald, Robert L. .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2015, 39 (09) :675-685
[3]   Quantitative monitoring of tamoxifen in human plasma extended to 40 metabolites using liquid-chromatography high-resolution mass spectrometry: new investigation capabilities for clinical pharmacology [J].
Dahmane, Elyes ;
Boccard, Julien ;
Csajka, Chantal ;
Rudaz, Serge ;
Decosterd, Laurent ;
Genin, Eric ;
Duretz, Benedicte ;
Bromirski, Maciej ;
Zaman, Khalil ;
Testa, Bernard ;
Rochat, Bertrand .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2014, 406 (11) :2627-2640
[4]   Analytical interference of 4-hydroxy-3-methoxymethamphetamine with the measurement of plasma free normetanephrine by ultra-high pressure liquid chromatography-tandem mass spectrometry [J].
Dunand, Marielle ;
Donzelli, Massimiliano ;
Rickli, Anna ;
Hysek, Cedric M. ;
Liechti, Matthias E. ;
Grouzmann, Eric .
CLINICAL BIOCHEMISTRY, 2014, 47 (12) :1121-1123
[5]   A LC-tandem MS assay for the simultaneous measurement of new antiretroviral agents: Raltegravir, maraviroc, darunavir, and etravirine [J].
Fayet, A. ;
Beguin, A. ;
Zanolari, B. ;
Cruchon, S. ;
Guignard, N. ;
Telenti, A. ;
Cavassini, M. ;
Guenthard, H. F. ;
Buclin, T. ;
Biollaz, J. ;
Rochat, B. ;
Decosterd, L. A. .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2009, 877 (11-12) :1057-1069
[6]   Comparison of the quantitative performance of a Q-Exactive high-resolution mass spectrometer with that of a triple quadrupole tandem mass spectrometer for the analysis of illicit drugs in wastewater [J].
Fedorova, Ganna ;
Randak, Tomas ;
Lindberg, Richard H. ;
Grabic, Roman .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2013, 27 (15) :1751-1762
[7]  
Fung EN, 2013, BIOANALYSIS, V5, P1277, DOI [10.4155/bio.13.81, 10.4155/BIO.13.81]
[8]   Use of high-resolution mass spectrometry to investigate a metabolite interference during liquid chromatography/tandem mass spectrometric quantification of a small molecule in toxicokinetic study samples [J].
Furlong, Michael ;
Bessire, Andrew ;
Song, Wei ;
Huntington, Christopher ;
Groeber, Elizabeth .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2010, 24 (13) :1902-1910
[9]  
Glauser G., 2016, ANAL CHEM
[10]   Therapeutic Drug Monitoring of the new targeted anticancer agents imatinib, nilotinib, dasatinib, sunitinib, sorafenib and lapatinib by LC tandem mass spectrometry [J].
Haouala, A. ;
Zanolari, B. ;
Rochat, B. ;
Montemurro, M. ;
Zaman, K. ;
Duchosal, M. A. ;
Ris, H. B. ;
Leyvraz, S. ;
Widmer, N. ;
Decosterd, L. A. .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2009, 877 (22) :1982-1996