Energy-resolved beam-monitoring system for undulator radiation using scattering from a chemical-vapor-deposition diamond film

被引:2
作者
Kudo, Togo [1 ,2 ]
Sano, Mutsumi [1 ]
Itoga, Toshiro [1 ]
Tajiri, Hiroo [1 ]
Ozaki, Kyosuke [2 ]
Takahashi, Sunao [1 ]
机构
[1] JASRI, 1-1-1 Kouto, Sayo, Hyogo 6795198, Japan
[2] RIKEN SPring 8 Ctr, 1-1-1 Kouto, Sayo, Hyogo 6795148, Japan
关键词
POSITION MONITORS; PHOTON; SPECTRA;
D O I
10.1063/1.5141152
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
With low-emittance synchrotron radiation rings, introducing accurate x-ray beams to a sample is difficult, and ensuring that the direction of the undulator beam is stable is essential. However, measuring the centroid of the undulator photon beam at the beamline front-end (FE) is difficult because the soft x-ray radiation is contaminated by the bending magnets upstream and downstream of the undulator. The x-ray beam position monitors (XBPMs), based on the interaction with the halo of undulator radiation, cannot estimate the centroid of the beam, and they cannot eliminate the effects of the bending magnets. To solve this problem, we have developed an energy-resolved beam-monitoring system for undulator radiation with the scattering from a diamond thin film deposited by chemical vapor deposition (CVD) in this study. An undulator x-ray beam is irradiated onto this film, and its elastic and Compton scattering are observed through a 50 mu m-diameter pinhole. A beam spot is detected through a pinhole camera system using a direct-detection-type charge-coupled device camera with energy resolution. The peak of the first-order harmonic of the undulator radiation is selectively visualized to measure the centroid of the undulator radiation, as well as the spectrum. The proposed system using a CVD diamond film can provide accurate position information for a photon beam exiting the FE.
引用
收藏
页数:5
相关论文
共 17 条
[1]   Blade-type X-ray beam position monitors for SPring-8 undulator beamlines [J].
Aoyagi, H ;
Kudo, T ;
Kitamura, H .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 467 :252-255
[2]   Characterization of beryllium and CVD diamond for synchrotron radiation beamline windows and x-ray beam monitor [J].
Goto, S. ;
Takahashi, S. ;
Kudo, T. ;
Yabashi, M. ;
Tamasaku, K. ;
Nishino, Y. ;
Ishikawa, T. .
ADVANCES IN X-RAY/EUV OPTICS AND COMPONENTS II, 2007, 6705
[3]   A compact X-ray free-electron laser emitting in the sub-angstrom region [J].
Ishikawa, Tetsuya ;
Aoyagi, Hideki ;
Asaka, Takao ;
Asano, Yoshihiro ;
Azumi, Noriyoshi ;
Bizen, Teruhiko ;
Ego, Hiroyasu ;
Fukami, Kenji ;
Fukui, Toru ;
Furukawa, Yukito ;
Goto, Shunji ;
Hanaki, Hirofumi ;
Hara, Toru ;
Hasegawa, Teruaki ;
Hatsui, Takaki ;
Higashiya, Atsushi ;
Hirono, Toko ;
Hosoda, Naoyasu ;
Ishii, Miho ;
Inagaki, Takahiro ;
Inubushi, Yuichi ;
Itoga, Toshiro ;
Joti, Yasumasa ;
Kago, Masahiro ;
Kameshima, Takashi ;
Kimura, Hiroaki ;
Kirihara, Yoichi ;
Kiyomichi, Akio ;
Kobayashi, Toshiaki ;
Kondo, Chikara ;
Kudo, Togo ;
Maesaka, Hirokazu ;
Marechal, Xavier M. ;
Masuda, Takemasa ;
Matsubara, Shinichi ;
Matsumoto, Takahiro ;
Matsushita, Tomohiro ;
Matsui, Sakuo ;
Nagasono, Mitsuru ;
Nariyama, Nobuteru ;
Ohashi, Haruhiko ;
Ohata, Toru ;
Ohshima, Takashi ;
Ono, Shun ;
Otake, Yuji ;
Saji, Choji ;
Sakurai, Tatsuyuki ;
Sato, Takahiro ;
Sawada, Kei ;
Seike, Takamitsu .
NATURE PHOTONICS, 2012, 6 (08) :540-544
[4]   Synchrotron radiation x-ray beam profile monitor using chemical vapor deposition diamond film [J].
Kudo, Togo ;
Takahashi, Sunao ;
Nariyama, Nobuteru ;
Hirono, Toko ;
Tachibana, Takeshi ;
Kitamura, Hideo .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (12)
[5]   Using direct illumination CCDs as high-resolution area detectors for X-ray scattering [J].
Livet, F ;
Bley, F ;
Mainville, J ;
Caudron, R ;
Mochrie, SGJ ;
Geissler, E ;
Dolino, G ;
Abernathy, D ;
Grübel, G ;
Sutton, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 451 (03) :596-609
[6]   IMPLEMENTATION OF THE DOPPLER BROADENING OF A COMPTON-SCATTERED PHOTON INTO THE EGS4 CODE [J].
NAMITO, Y ;
BAN, S ;
HIRAYAMA, H .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 349 (2-3) :489-494
[7]   Beamline for surface and interface structures at SPring-8 [J].
Sakata, O ;
Furukawa, Y ;
Goto, S ;
Mochizuki, T ;
Uruga, T ;
Takeshita, K ;
Ohashi, H ;
Ohata, T ;
Matsushita, T ;
Takahashi, S ;
Tajiri, H ;
Ishikawa, T ;
Nakamura, M ;
Ito, M ;
Sumitani, K ;
Takahashi, T ;
Shimura, T ;
Saito, A ;
Takahashi, M .
SURFACE REVIEW AND LETTERS, 2003, 10 (2-3) :543-547
[8]   Smart X-ray beam position monitor system using artificial-intelligence methods for the Advanced Photon Source insertion-device beamlines [J].
Shu, DM ;
Ding, H ;
Barraza, J ;
Kuzay, TM ;
Haeffner, D ;
Ramanathan, M .
JOURNAL OF SYNCHROTRON RADIATION, 1998, 5 :632-635
[9]   New x-ray pink-beam profile monitor system for the SPring-8 beamline front-end [J].
Takahashi, Sunao ;
Kudo, Togo ;
Sano, Mutsumi ;
Watanabe, Atsuo ;
Tajiri, Hiroo .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (08)
[10]  
Tanaka H., 2016, IEEE SYS MAN CYBERN, P2867