Development of a spherically focused phased array transducer for ultrasonic image-guided hyperthermia

被引:26
作者
Liu, Jingfei [1 ]
Foiret, Josquin [1 ]
Stephens, Douglas N. [1 ]
Le Baron, Olivier [2 ]
Ferrara, Katherine W. [1 ]
机构
[1] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
[2] Imasonic, F-70190 Voray Sur Lognon, France
关键词
phased array transducer; focused ultrasound therapy; ultrasonic image-guided hyperthermia; ultrasound hyperthermia; MALIGNANT BRAIN-TUMORS; THERMAL THERAPY; MILD HYPERTHERMIA; ANIMAL-MODELS; SURGERY; SYSTEM; CANCER; FEASIBILITY; PROSTATE; TISSUE;
D O I
10.1088/0031-9155/61/14/5275
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A 1.5 MHz prolate spheroidal therapeutic array with 128 circular elements was designed to accommodate standard imaging arrays for ultrasonic image-guided hyperthermia. The implementation of this dual-array system integrates real-time therapeutic and imaging functions with a single ultrasound system (Vantage 256, Verasonics). To facilitate applications involving small animal imaging and therapy the array was designed to have a beam depth of field smaller than 3.5 mm and to electronically steer over distances greater than 1 cm in both the axial and lateral directions. In order to achieve the required f number of 0.69, 1-3 piezocomposite modules were mated within the transducer housing. The performance of the prototype array was experimentally evaluated with excellent agreement with numerical simulation. A focal volume (2.70 mm (axial) x 0.65 mm (transverse) x 0.35 mm (transverse)) defined by the -6 dB focal intensity was obtained to address the dimensions needed for small animal therapy. An electronic beam steering range defined by the -3 dB focal peak intensity (17 mm (axial) x 14 mm (transverse) x 12 mm (transverse)) and -8 dB lateral grating lobes (24 mm (axial) x 18 mm (transverse) x 16 mm (transverse)) was achieved. The combined testing of imaging and therapeutic functions confirmed well-controlled local heating generation and imaging in a tissue mimicking phantom. This dual-array implementation offers a practical means to achieve hyperthermia and ablation in small animal models and can be incorporated within protocols for ultrasound-mediated drug delivery.
引用
收藏
页码:5275 / 5296
页数:22
相关论文
共 65 条
[1]   MEANS FOR COMPUTING THE KIRCHHOFF SURFACE INTEGRAL FOR A DISK RADIATOR AS A SINGLE INTEGRAL WITH FIXED LIMITS [J].
ARCHERHALL, JA ;
BASHTER, AI ;
HAZELWOOD, AJ .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1979, 65 (06) :1568-1570
[2]   Dual-Frequency Ultrasound Imaging and Therapeutic Bilaminar Array Using Frequency Selective Isolation Layer [J].
Azuma, Takashi ;
Ogihara, Makoto ;
Kubota, Jun ;
Sasaki, Akira ;
Umemura, Shin-ichiro ;
Furuhata, Hiroshi .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2010, 57 (05) :1211-1224
[3]   Adaptive Transthoracic Refocusing of Dual-Mode Ultrasound Arrays [J].
Ballard, John R. ;
Casper, Andrew J. ;
Wan, Yayun ;
Ebbini, Emad S. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2010, 57 (01) :93-102
[4]   Localised hyperthermia in rodent models using an MRI-compatible high-intensity focused ultrasound system [J].
Bing, Chenchen ;
Nofiele, Joris ;
Staruch, Robert ;
Ladouceur-Wodzak, Michelle ;
Chatzinoff, Yonatan ;
Ranjan, Ashish ;
Chopra, Rajiv .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2015, 31 (08) :813-822
[5]   FOCUSING OF HIGH-INTENSITY ULTRASOUND THROUGH THE RIB CAGE USING A THERAPEUTIC RANDOM PHASED ARRAY [J].
Bobkova, Svetlana ;
Gavrilov, Leonid ;
Khokhlova, Vera ;
Shaw, Adam ;
Hand, Jeffrey .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2010, 36 (06) :888-906
[6]  
BRITT RH, 1983, MED INSTRUM, V17, P172
[7]   Real-Time Implementation of a Dual-Mode Ultrasound Array System: In Vivo Results [J].
Casper, Andrew J. ;
Liu, Dalong ;
Ballard, John R. ;
Ebbini, Emad S. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2013, 60 (10) :2751-2759
[8]   Comparison of acoustic fields radiated from piezoceramic and piezocomposite focused radiators [J].
Cathignol, D ;
Sapozhnikov, OA ;
Theillère, Y .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1999, 105 (05) :2612-2617
[9]   In vivo feasibility of image-guided transvaginal focused ultrasound therapy for the treatment of intracavitary fibroids [J].
Chan, AH ;
Fujimoto, VY ;
Moore, DE ;
Held, RT ;
Paun, M ;
Vaezy, S .
FERTILITY AND STERILITY, 2004, 82 (03) :723-730
[10]   Method for MRI-guided conformal thermal therapy of prostate with planar transurethral ultrasound heating applicators [J].
Chopra, R ;
Burtnyk, M ;
Haider, MA ;
Bronskill, MJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (21) :4957-4975