Breast cancer staging: Combined digital breast tomosynthesis and automated breast ultrasound versus magnetic resonance imaging

被引:17
作者
Girometti, Rossano [1 ]
Tomkova, Ludmila [1 ]
Cereser, Lorenzo [1 ]
Zuiani, Chiara [1 ]
机构
[1] Univ Udine, Univ Hosp, Inst Radiol, Dept Med, Ple S Maria Misericordia 15`, I-33100 Udine, UD, Italy
关键词
Breast cancer; Digital breast; Tomosynthesis; Automated breast volume scanner; Magnetic resonance imaging; HAND-HELD ULTRASOUND; VOLUME SCANNER ABVS; PREOPERATIVE ASSESSMENT; DIAGNOSTIC-ACCURACY; MAMMOGRAPHY; LESIONS; MRI; ULTRASONOGRAPHY; SIZE;
D O I
10.1016/j.ejrad.2018.09.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate whether combined Digital breast tomosynthesis and Automated breast volume scanner (DBT-ABVS) are comparable to Magnetic resonance imaging (MRI) in staging breast cancer. Methods: We retrospectively included seventy-three patients with histologically proven breast cancer who underwent preoperative DBT, ABVS and 1.5 T MRI in the period July 2015-July 2016. Two radiologists in consensus recorded the number, site and Breast imaging-reporting and data system (BI-RADS) category of breast findings during two independent reading strategies, i.e. DBT-ABVS vs. MRI. Using histology or 1-year follow up as the standard of reference, we calculated the accuracy for cancer of both imaging strategies. Bland-Altman analysis was used to evaluate the agreement between MRI vs. DBT or ABVS in cancer size assessment. Results: Patients showed a total of 160 lesions (108 malignant and 52 benign). Malignant lesions were unifocal, multifocal, multicentric and biltateral in 53, 15, 4 and 1 cases, respectively. Diagnostic accuracy of DBT-ABVS vs. MRI was comparable for all cancers (90.0% [95% C.I. 84.3-94.2] vs. 93.8% [95% C.I. 88.8-97.0], respectively). DBT-ABVS showed lower sensitivity and positive predictive values for additional disease (76.5% [95% C.I. 58.8-89.3] vs. 91.7% [95% C.I. 84.6-96.1], and 78.8% [95% C.I. 61.0-91.0] vs 93.4% [95% C.I. 86.9-97.3], respectively). Compared to MRI, ABVS + DBT missed 6 lesions, including two invasive cancers and one extensive intravascular invasion associated to ductal carcinoma in situ. Bland-Altman analysis showed ABVS to agree with MRI at a higher extent than DBT in assessing cancer size. Conclusions: Though less performing than MRI, DBT-ABVS showed acceptable diagnostic accuracy in staging breast cancer. This strategy might be used if MRI is unavailable or unfeasible.
引用
收藏
页码:188 / 195
页数:8
相关论文
共 40 条
[1]   Diagnostic accuracy of mammography, clinical examination, US, and MR imaging in preoperative assessment of breast cancer [J].
Berg, WA ;
Gutierrez, L ;
NessAiver, MS ;
Carter, WB ;
Bhargavan, M ;
Lewis, RS ;
Ioffe, OB .
RADIOLOGY, 2004, 233 (03) :830-849
[2]   Characterization of Breast Masses in Digital Breast Tomosynthesis and Digital Mammograms: An Observer Performance Study [J].
Chan, Heang-Ping ;
Helvie, Mark A. ;
Hadjiiski, Lubomir ;
Jeffries, Deborah O. ;
Klein, Katherine A. ;
Neal, Colleen H. ;
Noroozian, Mitra ;
Paramagul, Chintana ;
Roubidoux, Marilyn A. .
ACADEMIC RADIOLOGY, 2017, 24 (11) :1372-1379
[3]   Automated breast ultrasound system (ABUS): reproducibility of mass localization, size measurement, and characterization on serial examinations [J].
Chang, Jung Min ;
Cha, Joo Hee ;
Park, Jeong Seon ;
Kim, Seung Ja ;
Moon, Woo Kyung .
ACTA RADIOLOGICA, 2015, 56 (10) :1163-1170
[4]   Comparison of Automated Breast Volume Scanning and Hand-Held Ultrasound in the Detection of Breast Cancer: an Analysis of 5,566 Patient Evaluations [J].
Choi, Woo Jung ;
Cha, Joo Hee ;
Kim, Hak Hee ;
Shin, Hee Jung ;
Kim, Hyunji ;
Chae, Eun Young ;
Hong, Min Ji .
ASIAN PACIFIC JOURNAL OF CANCER PREVENTION, 2014, 15 (21) :9101-9105
[5]   Assessment of disease extent on contrast-enhanced MRI in breast cancer detected at digital breast tomosynthesis versus digital mammography alone [J].
Chudgar, A. V. ;
Conant, E. F. ;
Weinstein, P. ;
Keller, M. ;
Synnestvedt, M. ;
Yamartino, P. ;
McDonald, E. S. .
CLINICAL RADIOLOGY, 2017, 72 (07) :573-579
[6]  
DOrsi C., 2013, ACR BIRADS. atlas, breast imaging reporting and data sys-tem, V5th
[7]   Breast tomosynthesis: Accuracy of tumor measurement compared with digital mammography and ultrasonography [J].
Fornvik, Daniel ;
Zackrisson, Sophia ;
Ljungberg, Otto ;
Svahn, Tony ;
Timberg, Pontus ;
Tingberg, Anders ;
Andersson, Ingvar .
ACTA RADIOLOGICA, 2010, 51 (03) :240-247
[8]   Understanding Bland Altman analysis [J].
Giavarina, Davide .
BIOCHEMIA MEDICA, 2015, 25 (02) :141-151
[9]   Automated breast volume scanner (ABVS) in assessing breast cancer size: A comparison with conventional ultrasound and magnetic resonance imaging [J].
Girometti, Rossano ;
Zanotel, Martina ;
Londero, Viviana ;
Linda, Anna ;
Lorenzon, Michele ;
Zuiani, Chiara .
EUROPEAN RADIOLOGY, 2018, 28 (03) :1000-1008
[10]   Interobserver reliability of automated breast volume scanner (ABVS) interpretation and agreement of ABVS findings with hand held breast ultrasound (HHUS), mammography and pathology results [J].
Golatta, Michael ;
Franz, Dorothea ;
Harcos, Aba ;
Junkermann, Hans ;
Rauch, Geraldine ;
Scharf, Alexander ;
Schuetz, Florian ;
Sohn, Christof ;
Heil, Joerg .
EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (08) :E332-E336