Pathological examination of an explanted phakic posterior chamber intraocular lens

被引:16
作者
Visessook, N [1 ]
Peng, Q [1 ]
Apple, DJ [1 ]
Gerl, R [1 ]
Schmickler, S [1 ]
Schoderbek, RJ [1 ]
Guindi, A [1 ]
机构
[1] Med Univ S Carolina, Dept Ophthalmol, Storm Eye Inst, Ctr Res Ocular Therapeut & Biodevices, Charleston, SC 29425 USA
关键词
D O I
10.1016/S0886-3350(99)80129-1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To report the clinicopathological correlation of an explanted phakic posterior chamber intraocular lens (PPC IOL) and to study the conformation of this lens implanted into human eyes obtained postmortem. Setting: Center for Research on Ocular Therapeutics and Biodevices, Department of Ophthalmology, Storm Eye Institute, Medical University of South Carolina, Charlest ton, South Carolina, USA. Methods: Three silicone PPC IOLs were explanted. One lens, explanted from a 38-year-old woman with complicated cataract, was submitted for pathological analysis. in addition to gross and Scanning electron microscopy (SEM), the lens was studied after experimental implantation in human eyes obtained postmortem using frontal, posterior, and side-view techniques. Results: Although gross and SEM showed that the IOL was well polished, examination of the lens in human cadaver eyes showed it was oversized and poorly fixated It was relatively bulky in its anterior-posterior dimension and revealed evidence of significant contact with the iris and crystalline lens. Conclusions: This study illustrates many pitfalls to be avoided in the design of a plate PPC IOL. The lens in this report was too large, and instead of ciliary sulcus fixation, it showed poor fixation through the zonules onto the posterior face of the pars plicata. Present and future PPC IOLs should be submitted for similar preclinical studies to clarify the type and site of fixation.
引用
收藏
页码:216 / 222
页数:7
相关论文
共 16 条
[1]  
APPLE DJ, 1990, OPHTHALMOLOGY, V97, P810
[2]   Silicone oil adhesion to intraocular lenses: An experimental study comparing various biomaterials [J].
Apple, DJ ;
Isaacs, RT ;
Kent, DG ;
Martinez, LM ;
Kim, S ;
Thomas, SG ;
Basti, S ;
Barker, D ;
Peng, Q .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1997, 23 (04) :536-544
[3]  
APPLE DJ, 1998, OCULAR PATHOLOGY CLI, P205
[4]  
APPLE DJ, 1995, SURV OPHTHALMOL, V40, P279
[5]   SIDE-VIEW ANALYSIS OF THE LENS .1. THE CRYSTALLINE LENS AND THE EVACUATED BAG [J].
ASSIA, EI ;
APPLE, DJ .
ARCHIVES OF OPHTHALMOLOGY, 1992, 110 (01) :89-93
[6]   SIDE-VIEW ANALYSIS OF THE LENS .2. POSITIONING OF INTRAOCULAR LENSES [J].
ASSIA, EI ;
APPLE, DJ .
ARCHIVES OF OPHTHALMOLOGY, 1992, 110 (01) :94-98
[7]   A modified preparation technique for closed system ocular surgery of human eyes obtained postmortem - An improved research and teaching tool [J].
Auffarth, GU ;
Wesendahl, TA ;
Solomon, KD ;
Brown, SJ ;
Apple, DJ .
OPHTHALMOLOGY, 1996, 103 (06) :977-982
[8]  
Baikoff G, 1990, Refract Corneal Surg, V6, P383
[9]   PHAKIC ANTERIOR-CHAMBER INTRAOCULAR LENSES [J].
BAIKOFF, G .
INTERNATIONAL OPHTHALMOLOGY CLINICS, 1991, 31 (01) :75-86
[10]  
ERTURK H, 1995, J REFRACT SURG, V11, P388