Evaluation of exposure factors of dual-energy contrast-enhanced mammography to optimize radiation dose with improved image quality

被引:1
|
作者
Niroshani, Sachila [1 ,2 ]
Nakamura, Tokiko [1 ,3 ]
Michiru, Nikaidou [1 ]
Negishi, Toru [1 ]
机构
[1] Tokyo Metropolitan Univ, Grad Sch Human Hlth Sci, Dept Radiol Sci, Tokyo, Japan
[2] Gen Sir John Kotelawala Def Univ, Fac Allied Hlth Sci, Dept Radiog & Radiotherapy, Werahera, Sri Lanka
[3] Juntendo Univ, Dept Radiol, Shizuoka Hosp, Shizuoka, Japan
关键词
Dual-energy contrast-enhanced mammography; low energy; high energy; radiation dose; DIGITAL MAMMOGRAPHY; SPECTRAL MAMMOGRAPHY; CESM;
D O I
10.1177/20584601221117251
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Dual-energy contrast-enhanced mammography (DECEM) is an advanced breast imaging technique of digital mammography. Purpose: To assess the total radiation dose received from complete DECEM using different combinations of exposure parameters for low- and high-energy images. Materials and methods: A dedicated phantom with three different concentrations of iodine inserts was used. Each iodine insert was 10 mm in diameter and concentration of 1.0 mgI/cm(3), 2.0 mgI/cm(3), and 4.0 mgI/cm(3). The phantom was exposed at varying kVp levels. Mean glandular dose (MGD) was estimated. Contrast to noise ratio (CNR) and figure of merit (FOM) of the iodine inserts were used to assess the image quality. Results: The optimum CNR of the recombined images was obtained by using 28 kVp + 49 kVp tube voltage combination for 50 mm thickness, 50% fibroglandular phantom only with a 26% dose increase compared to the highest voltages (32 kVp + 49 kVp) that can be used for low energy (LE) and high energy (HE) imaging. The CNR value was increased with increasing iodine concentration (R-2 > 0.99). Conclusion: The use of as low as possible tube voltage for the LE imaging of standard 50% fibroglandular-50% adipose, 50 mm thickness breast while using the highest tube voltage for HE imaging has reduced the MGD while keeping optimum image quality.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Glandular dose in contrast-enhanced dual-energy mammography
    Cossu, E.
    Castellani, F.
    Fiaschetti, V.
    Mariateresa, M.
    Tiziana, R.
    Angelo, T.
    CANCER RESEARCH, 2019, 79 (04)
  • [2] Dual-energy contrast-enhanced mammography
    Travieso Aja, M. M.
    Rodriguez Rodriguez, M.
    Mayon Hernandez, S.
    Vega Benitez, V.
    Luzardo, O. P.
    RADIOLOGIA, 2014, 56 (05): : 390 - 399
  • [3] Dose and image quality measurements for contrast-enhanced dual energy mammography systems
    Oduko, J. M.
    Homolka, P.
    Jones, V
    Whitwam, D.
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [4] Contrast-enhanced dual-energy digital mammography in the evaluation of breast cancer
    Awad, Fatma Mohamed
    Ismaeil, Hanaa
    EGYPTIAN JOURNAL OF RADIOLOGY AND NUCLEAR MEDICINE, 2014, 45 (02): : 591 - 595
  • [5] Dual-energy contrast-enhanced breast tomosynthesis: optimization of beam quality for dose and image quality
    Samei, Ehsan
    Saunders, Robert S., Jr.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (19): : 6359 - 6378
  • [6] Dual-energy contrast-enhanced spectral mammography (CESM)
    Daniaux, Martin
    De Zordo, Tobias
    Santner, Wolfram
    Amort, Birgit
    Koppelstaetter, Florian
    Jaschke, Werner
    Dromain, Clarisse
    Oberaigner, Willi
    Hubalek, Michael
    Marth, Christian
    ARCHIVES OF GYNECOLOGY AND OBSTETRICS, 2015, 292 (04) : 739 - 747
  • [7] Motion artifacts in dual-energy contrast-enhanced mammography
    Allec, Nicholas
    Abbaszadeh, Shiva
    Lewin, John M.
    Karim, Karim S.
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [8] Dual-energy contrast-enhanced spectral mammography (CESM)
    Martin Daniaux
    Tobias De Zordo
    Wolfram Santner
    Birgit Amort
    Florian Koppelstätter
    Werner Jaschke
    Clarisse Dromain
    Willi Oberaigner
    Michael Hubalek
    Christian Marth
    Archives of Gynecology and Obstetrics, 2015, 292 : 739 - 747
  • [9] Dual-energy contrast-enhanced digital mammography: Examination protocol
    Barra, F. R.
    Ribeiro, A. C.
    Mathieu, O. D.
    Rodrigues, A. C.
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2014, 95 (03) : 351 - 352
  • [10] Dual-energy contrast-enhanced digital subtraction mammography: Feasibility
    Lewin, JM
    Isaacs, PK
    Vance, V
    Larke, FJ
    RADIOLOGY, 2003, 229 (01) : 261 - 268