Flat-Panel CT Arthrography Feasibility Study and Comparison to Multidetector CT Arthrography

被引:17
作者
Guggenberger, Roman [1 ]
Fischer, Michael Alexander [1 ]
Hodler, Juerg [1 ]
Pfammatter, Thomas [1 ]
Andreisek, Gustav [1 ]
机构
[1] Univ Zurich Hosp, Dept Radiol, Musculoskeletal Imaging Res Grp, CH-8091 Zurich, Switzerland
关键词
flat-panel CT; arthrography; cone-beam CT; COMPUTED-TOMOGRAPHY ARTHROGRAPHY; TRABECULAR STRUCTURE-ANALYSIS; DIAGNOSTIC-ACCURACY; IMAGE QUALITY; DETECTOR CT; VOLUME CT; NOISE; MDCT; AGREEMENT;
D O I
10.1097/RLI.0b013e3182419082
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: To show the feasibility of flat-panel computed tomography (FPCT) arthrography and quantitatively and qualitatively compare different FPCT protocols with standard multidetector computed tomography (MDCT). Materials and Methods: First, a phantom simulating joint space with increasing iodine concentrations was scanned using a standard MDCT and 3 different FPCT protocols. Quantitative analyses were performed by measuring CT numbers of iodine dilutions, radiation dose, and image noise as well as signal-to-noise ratio and contrast-to-noise ratio. Second, FPCT arthrographies of 4 animal joint specimens were performed and analyzed qualitatively by 2 independent readers who evaluated image artifacts, image noise, overall image quality and anatomic depiction of bone, cartilage, and soft tissue. Kappa values were calculated for inter-reader agreement. Pearson's correlation coefficient (r) and Wilcoxon signed-ranks test with Bonferroni corrections for multiple comparisons were used to compare MDCT and FPCT. Results: In phantoms, all CT scans showed a linear correlation between increasing iodine concentrations and mean HU values of contrast media and radiation dose, respectively (r = 0.98-0.99, P < 0.01). Dose-length product remained constant for MDCT scans. Signal-to-noise ratio for phantom water linearly decreased in all FPCT scans with increasing iodine concentrations. Contrast-to-noise ratio curves showed reduced slope at iodine concentrations higher than 75 mg/mL. FPCT arthrography after intra-articular administration of 5 to 6 mL of a 25% dilution of iopromide (Ultravist 300 mg/mL, Bayer HealthCare, Berlin, Germany) was successfully performed in all 4 animal joint specimens. Kappa values for inter-reader agreement of qualitative image analyses were 0.62 to 0.91. Image and depiction quality of 20-s FPCT scans were similar or superior compared with standard MDCT (P < 0.005). Conclusion: FPCT arthrography is feasible and may allow similar image quality compared with standard MDCT arthrography.
引用
收藏
页码:312 / 318
页数:7
相关论文
共 50 条
[21]   Imaging of osseous and cartilaginous intraarticular bodies in the knee: Comparison of MR imaging and MR arthrography with CT and CT arthrography in cadavers [J].
Brossmann, J ;
Preidler, KW ;
Daenen, B ;
Pedowitz, RA ;
Andresen, R ;
Clopton, P ;
Trudell, D ;
Pathria, M ;
Resnick, D .
RADIOLOGY, 1996, 200 (02) :509-517
[22]   The role of multidetector CT arthrography in the investigation of suspected intra-articular hip pathology [J].
Christie-Large, M. ;
Tapp, M. J. F. ;
Theivendran, K. ;
James, S. L. J. .
BRITISH JOURNAL OF RADIOLOGY, 2010, 83 (994) :861-867
[23]   Clinical Application of Flat-Panel CT in the Angio Suite [J].
Hausegger, K. A. ;
Fuerstner, M. ;
Hauser, M. ;
Smetana, F. ;
Kau, T. .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2011, 183 (12) :1116-1122
[24]   Evaluation of glenoid labral tears: comparison between dual-energy CT arthrography and MR arthrography of the shoulder [J].
Giovanni Foti ;
William Mantovani ;
Matteo Catania ;
Paolo Avanzi ;
Simone Caia ;
Claudio Zorzi ;
Giovanni Carbognin .
La radiologia medica, 2020, 125 :39-47
[25]   Evaluation of glenoid labral tears: comparison between dual-energy CT arthrography and MR arthrography of the shoulder [J].
Foti, Giovanni ;
Mantovani, William ;
Catania, Matteo ;
Avanzi, Paolo ;
Caia, Simone ;
Zorzi, Claudio ;
Carbognin, Giovanni .
RADIOLOGIA MEDICA, 2020, 125 (01) :39-47
[26]   MR and CT Arthrography of the Wrist [J].
Cerezal, Luis ;
de Dios Berna-Mestre, Juan ;
Canga, Ana ;
Llopis, Eva ;
Rolon, Alejandro ;
Martin-Oliva, Xavier ;
del Pinal, Francisco .
SEMINARS IN MUSCULOSKELETAL RADIOLOGY, 2012, 16 (01) :27-41
[27]   Comparing the Image Quality of a Mobile Flat-Panel Computed Tomography and a Multidetector Computed Tomography A Phantom Study [J].
Neubauer, Jakob ;
Voigt, Johannes M. ;
Lang, Hannah ;
Scheuer, Carsten ;
Goerke, Sebastian M. ;
Langer, Mathias ;
Fiebich, Martin ;
Kotter, Elmar .
INVESTIGATIVE RADIOLOGY, 2014, 49 (07) :491-497
[28]   Flat-panel conebeam CT in the clinic: history and current state [J].
Fahrig, Rebecca ;
Jaffray, David A. ;
Sechopoulos, Ioannis ;
Stayman, J. Webster .
JOURNAL OF MEDICAL IMAGING, 2021, 8 (05)
[29]   Motion gated small animal imaging with a flat-panel CT [J].
Grasruck, M. ;
Bartling, S. ;
Dinkel, J. ;
Kiessling, F. ;
Semmler, W. ;
Stierstorfer, K. ;
Schmidt, B. .
MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
[30]   Comparison of Diagnostic Accuracy of 3.0-T MR Arthrography and CT Arthrography in Intraarticular Hip Pathology [J].
Jeong, Ji Won ;
Park, Ji Seon ;
Ryu, Kyung Nam ;
Cho, Yoon-Je .
INVESTIGATIVE MAGNETIC RESONANCE IMAGING, 2024, 28 (03) :122-131