Performance Evaluation of the Sysmex CS-5100 Automated Coagulation Analyzer

被引:7
作者
Chen, Liming [1 ]
Chen, Yu [1 ]
机构
[1] Zhejiang Univ, Dept Lab Med, Coll Med, Affiliated Hosp 1, Hangzhou 310003, Zhejiang, Peoples R China
关键词
evaluation; coagulation tests; performance; Sysmex CS-5100; comparison; D-DIMER;
D O I
10.7754/Clin.Lab.2014.141124
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Coagulation testing is widely applied clinically, and laboratories increasingly demand automated coagulation analyzers with short turn-around times and high-throughput. The purpose of this study was to evaluate the performance of the Sysmex CS-5100 automated coagulation analyzer for routine use in a clinical laboratory. Methods: The prothrombin time (PT), international normalized ratio (INR), activated partial thromboplastin time (APTT), fibrinogen (Fbg), and D-dimer were compared between the Sysmex CS-5100 and Sysmex CA-7000 analyzers, and the imprecision, comparison, throughput, STAT function, and performance for abnormal samples were measured in each. Results: The within-run and between-run coefficients of variation (CV) for the PT, APTT, INR, and D-dimer analyses showed excellent results both in the normal and pathologic ranges. The correlation coefficients between the Sysmex CS-5100 and Sysmex CA-7000 were highly correlated. The throughput of the Sysmex CS-5100 was faster than that of the Sysmex CA-7000. There was no interference at all by total bilirubin concentrations and triglyceride concentrations in the Sysmex CS-5100 analyzer. Conclusions: We demonstrated that the Sysmex CS-5100 performs with satisfactory imprecision and is well suited for coagulation analysis in laboratories processing large sample numbers and icteric and lipemic samples.
引用
收藏
页码:653 / 660
页数:8
相关论文
共 22 条
[1]   ADAPTATION OF SYNTHETIC PEPTIDE SUBSTRATE-BASED ASSAYS ON A DISCRETE ANALYZER [J].
AIACH, M ;
LEON, M ;
MICHAUD, A ;
CAPRON, L .
SEMINARS IN THROMBOSIS AND HEMOSTASIS, 1983, 9 (03) :206-216
[2]  
AMIRAL J, 1984, CLIN CHEM, V30, P1512
[3]  
[Anonymous], 1992, FED REG
[4]  
[Anonymous], 1999, WHO TECHNICAL REPORT
[5]   Evaluation and performance characteristics of the automated coagulation analyzer ACL TOP [J].
Appert-Flory, Anny ;
Fischer, Florence ;
Jambou, Didier ;
Toulon, Pierre .
THROMBOSIS RESEARCH, 2007, 120 (05) :733-743
[6]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[7]  
CLAUSS A., 1957, ACTA HAEMATOL, V17, P237
[8]  
CLSI, 2004, EV PREC PERF QUANT M
[9]   Evaluation of the automated coagulation analyzer Sysmex® CA-7000 [J].
Dorn-Beineke, A ;
Dempfle, CE ;
Bertsch, T ;
Wisser, H .
THROMBOSIS RESEARCH, 2005, 116 (02) :171-179
[10]   Evaluation of the automated coagulation analyzer Sysmex® CA-7000 [J].
Fischer, F ;
Appert-Flory, A ;
Jambou, D ;
Toulon, P .
THROMBOSIS RESEARCH, 2006, 117 (06) :721-729