Macular Fluid Reduces Reproducibility of Choroidal Thickness Measurements on Enhanced Depth Optical Coherence Tomography

被引:16
|
作者
Wong, Sophia S. [1 ]
Vuong, Vivian S. [1 ]
Cunefare, David [2 ]
Farsiu, Sina [2 ,3 ]
Moshiri, Ala [1 ]
Yiu, Glenn [1 ]
机构
[1] Univ Calif Davis, Dept Ophthalmol, 4860 Y St,Suite 2400, Sacramento, CA 95817 USA
[2] Duke Univ, Dept Biomed Engn, Durham, NC 27706 USA
[3] Duke Univ, Dept Ophthalmol, Durham, NC USA
关键词
RETINAL VEIN OCCLUSION; CENTRAL SEROUS CHORIORETINOPATHY; GROWTH-FACTOR THERAPY; DIABETIC-RETINOPATHY; PANRETINAL PHOTOCOAGULATION; NORMAL EYES; REPEATABILITY; DEGENERATION; EDEMA;
D O I
10.1016/j.ajo.2017.10.005
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To determine if different types of retinal fluid in the central macula affect the reproducibility of choroidal thickness (CT) measurements on enhanced depth imaging optical coherence tomography (EDI-OCT). DESIGN: Retrospective reliability analysis. METHODS: EDI-OCT images were obtained and the choroidal-scleral junction was analyzed through semiautomated segmentation. CT was measured at the fovea and averaged across the central 3-mm horizontal segment. Demographic data, central macular thickness, and type of fluid present were recorded. Intragrader and intergrader repeatability were assessed using the intraclass correlation coefficient (ICC) and coefficient of repeatability (CR). RESULTS: Of 124 eyes analyzed, 60 (48.4%) had diabetic macular edema, 32 (25.8%) had neovascular age-related macular degeneration, and 32 (25.8%) had other causes of fluid. Intergrader ICC (CR) was 0.95 (74.1 mu m) and 0.96 (63.9 mu m) for subfoveal and average CT, respectively. CR was similar across various causes of retinal fluid, but was worst for subretinal fluid compared to intraretinal or sub-retinal pigment epithelial fluid. CR also worsened with increasing choroidal thickness, but was not affected by retinal thickness. Intragrader repeatability was generally greater than intergrader values, and followed the same trend. CONCLUSIONS: The presence of macular fluid reduces CT measurement reproducibility, particularly in eyes with subretinal fluid and greater choroidal thickness. A difference of 74.1 mu m in subfoveal CT or 63.9 pm in average CT may be necessary to detect true clinical change in eyes with macular fluid. (C) 2017 Elsevier Inc. All rights reserved.
引用
收藏
页码:108 / 114
页数:7
相关论文
共 50 条
  • [31] A Simplified Method to Measure Choroidal Thickness Using Adaptive Compensation in Enhanced Depth Imaging Optical Coherence Tomography
    Gupta, Preeti
    Sidhartha, Elizabeth
    Girard, Michael J. A.
    Mari, Jean Martial
    Wong, Tien-Yin
    Cheng, Ching-Yu
    PLOS ONE, 2014, 9 (05):
  • [32] Macular Choroidal Thickness and Volume in Eyes With Angioid Streaks Measured by Swept Source Optical Coherence Tomography
    Ellabban, Abdallah A.
    Tsujikawa, Akitaka
    Matsumoto, Akiko
    Ogino, Ken
    Hangai, Masanori
    Ooto, Sotaro
    Yamashiro, Kenji
    Akiba, Masahiro
    Yoshimura, Nagahisa
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2012, 153 (06) : 1133 - 1143
  • [33] REPEATABILITY OF CHOROIDAL THICKNESS MEASUREMENTS ASSESSED WITH SWEPT-SOURCE OPTICAL COHERENCE TOMOGRAPHY IN HEALTHY AND DIABETIC INDIVIDUALS
    Abadia, Beatriz
    Calvo, Pilar
    Bartol-Puyal, Francisco
    Verdes, Guayente
    Sunen, Ines
    Ferreras, Antonio
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2019, 39 (04): : 786 - 793
  • [34] Macular choroidal thickness in normal Egyptians measured by swept source optical coherence tomography
    Moussa, Magdy
    Sabry, Dalia
    Soliman, Wael
    BMC OPHTHALMOLOGY, 2016, 16
  • [35] Enhanced Depth Imaging Optical Coherence Tomography of Small Choroidal Melanoma Comparison With Choroidal Nevus
    Shields, Carol L.
    Kaliki, Swathi
    Rojanaporn, Duangnate
    Ferenczy, Sandor R.
    Shields, Jerry A.
    ARCHIVES OF OPHTHALMOLOGY, 2012, 130 (07) : 850 - 856
  • [36] Choroidal thickness measurements with optical coherence tomography in branch retinal vein occlusion
    Coban-Karatas, Muge
    Altan-Yaycioglu, Rana
    Ulas, Burak
    Sizmaz, Selcuk
    Canan, Handan
    Sariturk, Cagla
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2016, 9 (05) : 725 - 729
  • [37] Repeatability and Reproducibility of Retinal Thickness Measurements in Diabetic Patients with Spectral Domain Optical Coherence Tomography
    Fiore, Tito
    Androudi, Sofia
    Iaccheri, Barbara
    Lupidi, Marco
    Fabrizio, Giansanti
    Fruttini, Daniela
    Biondi, Laura
    Cagini, Carlo
    CURRENT EYE RESEARCH, 2013, 38 (06) : 674 - 679
  • [38] Reproducibility of Retinal Thickness Measurements on Normal and Pathologic Eyes by Different Optical Coherence Tomography Instruments
    Giani, Andrea
    Cigada, Mario
    Choudhry, Netan
    Deiro, Antonio Peroglio
    Oldani, Marta
    Pellegrini, Marco
    Invernizzi, Alessandro
    Duca, Piergiorgio
    Miller, Joan W.
    Staurenghi, Giovanni
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2010, 150 (06) : 815 - 824
  • [39] Choroidal thickness measurement by enhanced depth imaging and swept-source optical coherence tomography in central serous chorioretinopathy
    Ferdiriva Hamzah
    Ari Shinojima
    Ryusaburo Mori
    Mitsuko Yuzawa
    BMC Ophthalmology, 14
  • [40] Macular Thickness Measurements in Healthy Eyes Using Six Different Optical Coherence Tomography Instruments
    Wolf-Schnurrbusch, Ute E. K.
    Ceklic, Lala
    Brinkmann, Christian K.
    Iliev, Milko E.
    Frey, Manuel
    Rothenbuehler, Simon P.
    Enzmann, Volker
    Wolf, Sebastian
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2009, 50 (07) : 3432 - 3437