Study on the light output non-uniformity of PbWO4 crystal

被引:0
|
作者
Shao, M [1 ]
Xu, T
Liu, MC
Wang, CR
Ruan, LJ
Chen, HF
Pan, XH
Wang, CZ
机构
[1] Univ Sci & Technol China, Dept Modern Phys, Hefei 230026, Peoples R China
[2] Hefei Univ Technol, Coll Mech & Auto Engn, Hefei 230009, Peoples R China
来源
HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION | 2001年 / 25卷 / 06期
关键词
lead-tungstate crystal; scintillator; non-uniformity; Na-22; source; surface roughness;
D O I
暂无
中图分类号
O57 [原子核物理学、高能物理学];
学科分类号
070202 ;
摘要
This paper presents the measurement on the light output non-uniformity of full size PbWO4 crystal, which will be used in the electromagnetic calorimeter (ECAL) of CMS detector. It makes use of the Na-22 radiactive source and the coincident method. With the help of convolution fitting method, it is possible to make fine measurement on the light output non-uniformity of the crystal. The effect of depolishing one of the lateral faces of the crystal on the non-uniformity has been studied. The slope of the non-uniformity curve decreases with the increasing of the roughness. When the roughness reaches R-x = (1.49 +/- 0.27) mum the non-uniformity curve meets the requirement of CMS-ECAL.
引用
收藏
页码:555 / 560
页数:6
相关论文
共 7 条
  • [1] AUFFRAY E, 2000, 014 CMS
  • [2] *CERN LHCC, 1997, 9733 CERNLHCC, P307
  • [3] DEPASS P, 1998, 068 CMS, P68
  • [4] HILLERMANNS H, 1998, 005 CMS
  • [5] Shao M, 2001, HIGH ENERG PHYS NUC, V25, P50
  • [6] WU J, 1998, NUCL INSTRUM METH A, V419, P154
  • [7] XU T, 2000, IN PRESS NUCL INST A