Monte-Carlo simulation of a compact gamma-ray detector using wavelength-shifting fibers coupled to a YAP scintillation crystal

被引:1
作者
Zhu Jie [1 ,2 ]
Ma Hong-Guang [1 ]
Ma Wen-Yan [1 ]
Zeng Hui [2 ]
Wang Zhao-Min [2 ]
Xu Zi-Zong [2 ]
机构
[1] Second Artillery Engn Inst, Xian 710025, Peoples R China
[2] Univ Sci & Technol China, Dept Modern Phys, Hefei 230026, Peoples R China
关键词
Monte-Carlo simulation; WSF; spatial resolution;
D O I
10.1088/1674-1137/32/5/010
中图分类号
O57 [原子核物理学、高能物理学];
学科分类号
070202 ;
摘要
The production and transportation of fluorescent light produced in wavelength-shifting fibers (WSFs) coupled to YAP scintillation crystal is simulated using the GEANT4 codes. An advantage of the wavelength-shifting fiber readout technique over a direct readout with a position-sensitive photo-sensor is the reduced requirement for position sensitive photomultiplier tube photocathode area. With this gamma-ray detector, the gamma camera is small and flexible and has larger effective field of view and low cost. Simulation results show that a) a mean 12 of photons per 59.5 keV gamma ray interaction is produced in the WSF located nearest to the incident gamma ray, and a spatial resolution of 3.6 mm FWHM is obtained, b) a mean 27 of photons per 140 keV gamma ray interaction is produced and a spatial resolution of 3.1 mm FWHM is obtained. Results demonstrate the feasibility of this concept of a compact gamma-ray detector based on wavelength-shifting fibers readout. However, since the very low photoelectron levels, it is very important to use a photon counting device with good single photo-electron response to readout the WSFs.
引用
收藏
页码:377 / 380
页数:4
相关论文
共 10 条
[1]   SCINTILLATION PROPERTIES OF YAP-CE [J].
BACCARO, S ;
BLAZEK, K ;
DENOTARISTEFANI, F ;
MALY, P ;
MARES, JA ;
PANI, R ;
PELLEGRINI, R ;
SOLURI, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 361 (1-2) :209-215
[2]   TOWARD A NUCLEAR-MEDICINE WITH SUB-MILLIMITER SPATIAL-RESOLUTION [J].
BARONE, LH ;
BLAZEK, K ;
BOLLINI, D ;
DELGUERRA, A ;
DENOTARISTEFANI, F ;
DEVINCENTIS, G ;
DIDOMENICO, G ;
GALLI, M ;
GIGANTI, M ;
MALY, P ;
PANI, R ;
PELLEGRINI, R ;
PERGOLA, A ;
PIFFANELLI, A ;
SCOPINARO, F ;
SOLURI, A ;
VITTORI, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 360 (1-2) :302-306
[3]   X-Y position readout of a YAP:Ce crystal matrix using wavelength-shifting fibers [J].
Del Guerra, A ;
Damiani, C ;
Zavattini, G ;
Di Domenico, G ;
Belcari, N ;
Vaiano, A .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (04) :1108-1113
[4]   Optimization of compact gamma cameras for breast imaging [J].
Garibaldi, F ;
Cisbani, E ;
Cusanno, F ;
Iommi, R ;
Urciuoli, GM ;
Pani, R ;
Pellegrini, R ;
Scafe', R ;
Indovina, L ;
Cinti, MN ;
Trotta, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 471 (1-2) :222-228
[5]  
*HAM PHOT TUB HAM, 1993, POS SENS PHOT TUB CR, V2486
[6]  
*KUR SCINT MAT, SPEC MAN
[7]   Development of a compact wavelength-shifting fibre gamma camera [J].
Soares, AJ ;
Cullum, I ;
Miller, DJ ;
Royle, GJ ;
Speller, RD .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (03) :1058-1064
[8]  
Zeng Haining, 2001, Acta Photonica Sinica, V30, P1321
[9]  
ZENG HN, 2000, HEP NP, V24, P165
[10]  
Zhu Jie, 2006, Acta Photonica Sinica, V35, P1497