High-intensity focused ultrasound;
Ocular pathology;
Glaucoma;
Ciliary bodies;
Conformal cyclocoagulation;
Medical device;
High frequency;
Third harmonic;
THERAPEUTIC ULTRASOUND;
INDUCED LESIONS;
TRABECULECTOMY;
ABLATION;
OUTCOMES;
GEL;
D O I:
10.1016/j.ultrasmedbio.2011.01.017
中图分类号:
O42 [声学];
学科分类号:
070206 ;
082403 ;
摘要:
This study examined the feasibility of high-intensity focused ultrasound (HIFU) for glaucoma treatment with conformal coagulation of the ciliary bodies (CB). A miniaturized high frequency (21 MHz) device was developed, based on the geometry of the eye and adapted to the anatomy of the rabbit eyeball. Six line-focus lesions were distributed along a circle and produced by six cylindrical transducers. To be conformal, the numerical model predicted an intensity of 6.9 W/cm(2), with exposure duration of 3 s ON (powered per sector). In vivo experiments were conducted on two rabbits. A significant intraocular pressure reduction was noted (-45% and -31%). Histology demonstrated conformal and homogeneous coagulation of the CB without side effects. (E-mail: thomas.charrel@inserm.fr) (C) 2011 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:742 / 754