Implant design and intraosseous stability of immediately placed implants: a human cadaver study

被引:108
作者
Akkocaoglu, M
Uysal, S
Tekdemir, I
Akca, K
Cehreli, MC
机构
[1] Hacettepe Univ, Dept Oral Surg, Fac Dent, Ankara, Turkey
[2] Hacettepe Univ, Dept Oral Diag & Radiol, Fac Dent, Ankara, Turkey
[3] Ankara Univ, Dept Anat, Fac Med, TR-06100 Ankara, Turkey
[4] Hacettepe Univ, Dept Prosthodont, Fac Dent, Ankara, Turkey
关键词
immediate placement; implant design; implant diameter; installation torque; mandible; removal torque; resonance frequency analysis;
D O I
10.1111/j.1600-0501.2004.01099.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: The objective of this study was to explore effects of implant macrodesign and diameter on initial intraosseous stability and interface mechanical properties of immediately placed implants. Material and method: Mandibular premolars of four fresh-frozen human cadavers were extracted. 04.1/4.8 mm ITI(R) TE(R), 04.1 and 4.8 mm solid screw synOcta(R) ITI(R) implants were placed into freshly prepared extraction sockets. Resonance frequency analysis was conducted to quantify primary implant stability quotient (ISQ). Installation torque value (ITV) and removal torque value (RTV) of the implants were measured using a custom-made strain-gauged torque wrench connected to a data acquisition system at a sample rate of 10,000 Hz. The vertical defect depth around the collar of each implant was measured directly by an endodontic spreader. The bone-implant contact was determined in digitalized images of periapical radiographs and expressed as percentage bone contact. Results: The ISQ values of the TE(R) implant was higher than the 04.1 mm implant (P < 0.01), and comparable with the 04.8 mm implants (P > 0.05). ITVs and RTVs of TE(R) and 04.8 mm implants were higher than the 04.1 mm implant, although the differences between groups were statistically insignificant (P > 0.05). The vertical defect depths around all types of implants were similar. In the radiographic analyses, percentage bone-implant contact of the TE(R) and 04.8 mm implants were comparable at the marginal bone region and both were higher than that of the 04.1 mm ITI9(R) implant. Nonparametric correlations between groups revealed a significant correlation between ITV and RTV (r=0.838; P < 0.001), but not between ISQ values and ITVs and RTVs (P > 0.05). Conclusion: Immediately placed ITI(R) TE(R) implant leads to initial intraosseous stability and interface mechanical properties comparable with a wide diameter implant.
引用
收藏
页码:202 / 209
页数:8
相关论文
共 30 条
[1]  
Balshi TJ, 2002, INT J ORAL MAX IMPL, V17, P587
[2]  
Barewal RM, 2003, INT J ORAL MAX IMPL, V18, P641
[3]   Tissue integration of non-submerged implants 1-year results of a prospective study with 100 ITI hollow-cylinder and hollow-screw implants [J].
Buser, D. ;
Weber, H-P ;
Lang, N. P. .
CLINICAL ORAL IMPLANTS RESEARCH, 1990, 1 (01) :33-40
[4]   Surgical procedures in partially edentulous patients with ITI implants [J].
Buser, D ;
von Arx, T .
CLINICAL ORAL IMPLANTS RESEARCH, 2000, 11 :83-100
[5]   Dynamic fatigue resistance of implant-abutment junction in an internally notched morse-taper oral implant:: influence of abutment design [J].
Çehreli, MC ;
Akça, K ;
Iplikçioglu, H ;
Sahin, S .
CLINICAL ORAL IMPLANTS RESEARCH, 2004, 15 (04) :459-465
[6]  
Cooper LF, 2002, INT J ORAL MAX IMPL, V17, P517
[7]  
Geris L., 2003, Computer Methods in Biomechanics and Biomedical Engineering, V6, P277, DOI 10.1080/10255840310001634412
[8]  
Glauser Roland, 2003, Clin Implant Dent Relat Res, V5 Suppl 1, P47, DOI 10.1111/j.1708-8208.2003.tb00015.x
[9]  
GOAZ PW, 1994, ORAL RADIOLOGY PRINC, P151
[10]   Assessing the implant/bone interface by using natural frequency analysis [J].
Huang, HM ;
Pan, LC ;
Lee, SY ;
Chiu, CL ;
Fan, KH ;
Ho, KN .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 2000, 90 (03) :285-291