Accuracy of pelvic radiotherapy: Prospective analysis of 90 patients in a randomised trial of blocked versus standard radiotherapy

被引:30
作者
Huddart, RA
Nahum, A
Neal, A
McLean, M
Dearnaley, DP
Law, M
Dyer, J
Tait, D
机构
[1] ROYAL MARSDEN HOSP,NHS TRUST,DEPT RADIOTHERAPY,SUTTON SM2 5PT,SURREY,ENGLAND
[2] ROYAL MARSDEN HOSP,NHS TRUST,DEPT ONCOL,SUTTON SM2 5PT,SURREY,ENGLAND
[3] ROYAL MARSDEN HOSP,NHS TRUST,DEPT PHYS,SUTTON SM2 5PT,SURREY,ENGLAND
[4] ROYAL MARSDEN HOSP,NHS TRUST,DEPT STAT,SUTTON SM2 5PT,SURREY,ENGLAND
关键词
conformal radiotherapy; pelvis; treatment errors; multivariate analysis;
D O I
10.1016/0167-8140(96)01717-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The aim of this study was to assess the accuracy of pelvic radiotherapy during a trial of blocked radiotherapy at the Royal Marsden Hospital, UK. Prospective evaluation was performed on 90 patients receiving CT planned pelvic radiotherapy using weekly anterior-posterior and lateral portal films. Field placement errors (FPEs) were calculated by comparing field centres of each film with a designated point of interest. Data was evaluated to calculate the overall treatment simulator differences, the number of error free treatments, and mean treatment-simulator position and to evaluate the role of systematic versus random errors. Age, weight, disease site, position of treatment, fractionation, blocked versus conventional techniques were assessed for their effect on treatment accuracy. The mean absolute error between treatment and simulator films was anterior right-left (ARL) 0.25 cm, anterior superior-inferior (ASI) 0.32 cm, lateral anterior-posterior (LAP) 0.42 cm, and lateral superior-inferior (LSI) 0.28 cm. On average the field centre was displaced by 0.66 cm (standard deviation, S.D. = 0.34) from that intended. On each treatment day 29% of anterior films and 45% of lateral films had at least one 0.5 cm error. Overall 59% of treatments had at least one 0.5 cm error and 9% a 1.0 cm error, The field centre was more than 0.5 cm from the position intended in 66% of treatments and over 1 cm for 14% of treatments. Analysis of variance showed that both random and systematic errors occurred in all directions, Though random errors were of similar magnitude in all direction (variance sigma(2) = 0.06-0.09 cm(2)); systematic errors showed a 4-fold variation being greatest in the LAP direction (sigma(2) = 0.19 cm(2)) and least the ARL direction (sigma(2) = 0.048 cm(2)). No factor consistently predicted for worse outcome in all directions. Hypofractionated treatments were less accurate in the LSI direction (P > 0.05). Systematic errors were associated in the ARL direction with hypofractionation (P < 0.01) and, in the LSI direction with weight (P < 0.03) and age (P < 0.05). We conclude that significant random and systematic errors can occur during pelvic radiotherapy especially in the LAP direction. These results suggest that in the absence of a customised immobilisation device, to cover 95% of errors, margins of 0.6 cm for RL and SI directions and 0.9 cm for AP direction should be allowed between the planning and clinical target volumes. However, ideally, each centre should determine their own margin requirements according to local clinical practice.
引用
收藏
页码:19 / 29
页数:11
相关论文
共 33 条
  • [1] Armitage P, 1987, Statistical methods in medical research, V2nd
  • [2] ONLINE REPOSITIONING DURING TREATMENT OF THE PROSTATE - A STUDY OF POTENTIAL LIMITS AND GAINS
    BALTER, JM
    CHEN, GTY
    PELIZZARI, CA
    KRISHNASAMY, S
    RUBIN, S
    VIJAYAKUMAR, S
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 (01): : 137 - 143
  • [3] A VERIFICATION PROCEDURE TO IMPROVE PATIENT SET-UP ACCURACY USING PORTAL IMAGES
    BEL, A
    VANHERK, M
    BARTELINK, H
    LEBESQUE, JV
    [J]. RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) : 253 - 260
  • [4] TRANSFER ERRORS OF PLANNING CT TO SIMULATOR - A POSSIBLE SOURCE OF SETUP INACCURACIES
    BEL, A
    BARTELINK, H
    VIJLBRIEF, RE
    LEBESQUE, JV
    [J]. RADIOTHERAPY AND ONCOLOGY, 1994, 31 (02) : 176 - 180
  • [5] MAXIMIZING SETUP ACCURACY USING PORTAL IMAGES AS APPLIED TO A CONFORMAL BOOST TECHNIQUE FOR PROSTATIC-CANCER
    BIJHOLD, J
    LEBESQUE, JV
    HART, AAM
    VIJLBRIEF, RE
    [J]. RADIOTHERAPY AND ONCOLOGY, 1992, 24 (04) : 261 - 271
  • [6] RADIATION-FIELD EDGE-DETECTION IN PORTAL IMAGES
    BIJHOLD, J
    GILHUIJS, KGA
    VANHERK, M
    MEERTENS, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (12) : 1705 - 1710
  • [7] QUALITY ASSURANCE USING PORTAL IMAGING - THE ACCURACY OF PATIENT POSITIONING IN IRRADIATION OF BREAST-CANCER
    CREUTZBERG, CL
    ALTHOF, VGM
    HUIZENGA, H
    VISSER, AG
    LEVENDAG, PC
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (03): : 529 - 539
  • [8] SIZES AND SOURCES OF FIELD PLACEMENT ERROR IN ROUTINE IRRADIATION FOR PROSTATE-CANCER
    DUNSCOMBE, P
    LOOSE, S
    LESZCZYNSKI, K
    [J]. RADIOTHERAPY AND ONCOLOGY, 1993, 26 (02) : 174 - 176
  • [9] TIME TREND OF PATIENT SETUP DEVIATIONS DURING PELVIC IRRADIATION USING ELECTRONIC PORTAL IMAGING
    ELGAYED, AAH
    BEL, A
    VIJLBRIEF, R
    BARTELINK, H
    LEBESQUE, JV
    [J]. RADIOTHERAPY AND ONCOLOGY, 1993, 26 (02) : 162 - 171
  • [10] IMAGE COMPARISON TECHNIQUES FOR USE WITH MEGAVOLTAGE IMAGING-SYSTEMS
    EVANS, PM
    GILDERSLEVE, JQ
    MORTON, EJ
    SWINDELL, W
    COLES, R
    FERRARO, M
    RAWLINGS, C
    XIAO, ZR
    DYER, J
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (776) : 701 - 709