Validity of single-channel masseteric electromyography by using an ultraminiature wearable electromyographic device for diagnosis of sleep bruxism

被引:30
作者
Maeda, Masana [1 ]
Yamaguchi, Taihiko [1 ]
Mikami, Saki [2 ]
Yachida, Wataru [3 ]
Saito, Taishi [1 ]
Sakuma, Toshimitsu [1 ]
Nakamura, Hironobu [1 ]
Saito, Miku [1 ]
Mizuno, Mariko [1 ]
Yamada, Kyoko [1 ]
Satoh, Kaoru [3 ]
机构
[1] Hokkaido Univ, Grad Sch Dent Med, Dept Crown & Bridge Prosthodont, Sapporo, Hokkaido, Japan
[2] Hokkaido Univ Hosp, Ctr Adv Oral Med, Dept Temporomandibular Disorders, Sapporo, Hokkaido, Japan
[3] Hokkaido Univ Hosp, Dept Crown & Bridge Prosthodont, Sapporo, Hokkaido, Japan
关键词
Sleep bruxism; Polysomnography; Single-channel electromyogram; Sensitivity and specificity; Cut-off value; JAW-MUSCLE-ACTIVITY;
D O I
10.1016/j.jpor.2019.04.003
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The purpose of this study was to clarify the validity of assessment of sleep bruxism (SB) by using single-channel electromyogram (EMG) and a cut-off value with optimum sensitivity and specificity. Methods: The subjects were twenty volunteers with clinical diagnosis of SB. Assessment by masseteric EMG data only by using a wearable EMG was compared with reference standard assessment by masseteric EMG data using polysomnography with audio-visual recording (PSG-AV). From EMG activities recorded by single-channel EMG, bursts of more than two times the baseline amplitude with a duration of 0.25 s or more were selected by a burst unit (EMG-burst-all). Furthermore, from EMG-burst-all, bursts that were more than 5-20% of the maximum voluntary contraction value (EMG-burst-5%, EMG-burst-10%, EMG-burst-20%) were selected. By an episode unit, phasic, tonic, and mixed episodes were selected by single-channel EMG (EMG-episodes). Among the EMG-episodes, further, reference standard episodes of SB (PSG-episodes) were selected by PSG-AV assessment. Results: Sixteen subjects were diagnosed as bruxers based on PSG-AV (PSG-episodes/h >2). By a burst unit and an episode unit, there were significant correlations between assessment variables of SB by single-channel EMG and PSG-AV except for EMG-burst-20%/h. When the cut-off value in EMG-episodes/h was 5.5/h for sleep bruxers, both sensitivity and specificity were 100%. As for variables by a burst unit, EMG-burst-all/h and EMG-burst-5% had higher values of sensitivity and specificity. Conclusions: The results suggested that single-channel EMG is valid for diagnosis of SB if a cut-off value that is appropriate for single-channel EMG is used. (C) 2019 Japan Prosthodontic Society. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:90 / 97
页数:8
相关论文
共 22 条
[1]  
AASM, 2005, SLEEP REL BRUX INT C, P189
[2]  
Abe S, 2009, INT J PROSTHODONT, V22, P342
[3]  
[Anonymous], 2014, INT CLASSIFICATION S
[4]  
[Anonymous], 2016, AASM MAN SCOR SLEEP
[5]   Sleep bruxism; an overview of an oromandibular sleep movement disorder [J].
Bader, G ;
Lavigne, G .
SLEEP MEDICINE REVIEWS, 2000, 4 (01) :27-43
[6]   Diagnostic accuracy of sleep bruxism scoring in absence of audio-video recording: a pilot study [J].
Carra, Maria Clotilde ;
Nelly Huynh ;
Lavigne, Gilles J. .
SLEEP AND BREATHING, 2015, 19 (01) :183-190
[7]  
Carra Maria Clotilde, 2012, Dent Clin North Am, V56, P387, DOI 10.1016/j.cden.2012.01.003
[8]   Detection of sleep bruxism: comparison between an electromyographic and electrocardiographic portable holter and polysomnography [J].
Castroflorio, T. ;
Deregibus, A. ;
Bargellini, A. ;
Debernardi, C. ;
Manfredini, D. .
JOURNAL OF ORAL REHABILITATION, 2014, 41 (03) :163-169
[9]  
Dreyer P., 2015, J Dent Sleep Med, V02, P147
[10]   Oro-facial activities in sleep bruxism patients and in normal subjects: a controlled polygraphic and audio-video study [J].
Dutra, K. M. C. ;
Pereira, F. J., Jr. ;
Rompre, P. H. ;
Huynh, N. ;
Fleming, N. ;
Lavigne, G. J. .
JOURNAL OF ORAL REHABILITATION, 2009, 36 (02) :86-92