Measurement of the modulation transfer function of digital X-ray detectors with an opaque edge-test device

被引:25
|
作者
Illers, H
Buhr, E
Günther-Kohfahl, S
Neitzel, U
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
[2] Philips Med Syst, D-22335 Hamburg, Germany
关键词
D O I
10.1093/rpd/nch506
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
A variant of the edge method for the determination of the pre-sampled modulation transfer function (MTF) of digital X-ray imaging devices has been developed and accepted as the standard method in the novel DQE measurement standard IEC 62220-1. An opaque tungsten edge-test device accomplishes the ideal step-like profile of the incident X rays, The edge spread function is measured over a large region across the edge transition that enables, an accurate MTF measurement including the 'low-frequency drop'. The method has been applied to different state-of-the-art X-ray imaging detectors, a computed radiography, a CsI-based indirect and an Se-based direct flat-panel detector. The MTF measurement results will be presented. In contrast to the opaque edge device, the commonly used semi-transparent edge-test devices produce scatter radiation that deteriorates the incident X-ray profile, which leads to a systematic overestimation of the MTF.
引用
收藏
页码:214 / 219
页数:6
相关论文
共 50 条
  • [31] A method for determining the information capacity of x-ray imaging scintillator detectors by means of luminescence and modulation transfer function measurements
    I. Kandarakis
    D. Cavouras
    E. Kanellopoulos
    C. D. Nomicos
    G. S. Panayiotakis
    Medical & Biological Engineering & Computing, 1999, 37 : 25 - 30
  • [32] The Effect of Oblique X-Ray Incidence on the Design of X-Ray Detectors for Digital Tomosynthesis
    Acciavatti, R.
    Maidment, A.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [33] MODULATION TRANSFER-FUNCTION OF A DIGITAL DENTAL-X-RAY SYSTEM
    CHEN, SK
    HOLLENDER, L
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1994, 77 (03): : 308 - 313
  • [34] Quality control measurements for digital x-ray detectors
    Marshall, N. W.
    Mackenzie, A.
    Honey, I. D.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (04): : 979 - 999
  • [35] Digital x-ray detectors for radiography: Technical innovations
    Dobbins, JT
    RADIOLOGY, 1999, 213P : 149 - 149
  • [36] A comparative study of digital x-ray software and detectors
    Prompalit, Sakunrat
    Sinsuebphon, Nattawut
    Thanasupsombat, Chalinee
    Thongvigitmanee, Saowapak
    13TH BIOMEDICAL ENGINEERING INTERNATIONAL CONFERENCE (BMEICON 2021), 2018,
  • [37] MODULATION TRANSFER-FUNCTION EVALUATION OF LINEAR SOLID-STATE X-RAY-SENSITIVE DETECTORS USING EDGE TECHNIQUES
    ZHU, YM
    KAFTANDJIAN, V
    PEIX, G
    BABOT, D
    APPLIED OPTICS, 1995, 34 (22): : 4937 - 4943
  • [38] A digital medical X-ray diagnostic device
    E. A. Babichev
    S. E. Baru
    A. I. Volobuev
    V. V. Gusev
    G. M. Kolachev
    Yu. A. Nemkov
    V. A. Neustroev
    E. L. Nekhanevich
    V. V. Porosev
    G. A. Savinov
    V. A. Sidorov
    Yu. G. Ukraintsev
    A. G. Khabakhpashev
    L. I. Shekhtman
    Yu. B. Yurchenko
    Biomedical Engineering, 1997, 31 (1) : 14 - 18
  • [39] Contrast transfer function measurement of X-ray solid state linear detectors using bar/space pattern methods
    Kaftandjian, V
    Zhu, YM
    Peix, G
    Babot, D
    NDT & E INTERNATIONAL, 1996, 29 (01) : 3 - 11
  • [40] Measurement of the modulation transfer function of an X-ray microscope based on multiple Fourier orders analysis of a Siemens star
    Oton, Joaquin
    Sorzano, Carlos Oscar S.
    Marabini, Roberto
    Pereiro, Eva
    Carazo, Jose M.
    OPTICS EXPRESS, 2015, 23 (08): : 9567 - 9572