Measurement of the modulation transfer function of digital X-ray detectors with an opaque edge-test device

被引:25
|
作者
Illers, H
Buhr, E
Günther-Kohfahl, S
Neitzel, U
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
[2] Philips Med Syst, D-22335 Hamburg, Germany
关键词
D O I
10.1093/rpd/nch506
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
A variant of the edge method for the determination of the pre-sampled modulation transfer function (MTF) of digital X-ray imaging devices has been developed and accepted as the standard method in the novel DQE measurement standard IEC 62220-1. An opaque tungsten edge-test device accomplishes the ideal step-like profile of the incident X rays, The edge spread function is measured over a large region across the edge transition that enables, an accurate MTF measurement including the 'low-frequency drop'. The method has been applied to different state-of-the-art X-ray imaging detectors, a computed radiography, a CsI-based indirect and an Se-based direct flat-panel detector. The MTF measurement results will be presented. In contrast to the opaque edge device, the commonly used semi-transparent edge-test devices produce scatter radiation that deteriorates the incident X-ray profile, which leads to a systematic overestimation of the MTF.
引用
收藏
页码:214 / 219
页数:6
相关论文
共 50 条
  • [21] Multifunctional digital x-ray image detectors
    Zhlobinskaya S.V.
    Morgun O.N.
    Nemchenko K.E.
    Biomedical Engineering, 2008, 42 (5) : 256 - 260
  • [22] Modulation transfer function (MTF) measurement of CdZnTe detectors for digital radiography and computed tomography
    Giakos, GC
    Vedantham, S
    Suryanarayanan, S
    Chowdhury, S
    Guntupalli, R
    Odogba, J
    Vega-Lozada, V
    Pillai, B
    Dasgupta, A
    WHERE INSTRUMENTATION IS GOING - CONFERENCE PROCEEDINGS, VOLS 1 AND 2, 1998, : 456 - 461
  • [23] Simple method for modulation transfer function determination of digital imaging detectors from edge images
    Buhr, E
    Günther-Kohfahl, S
    Neitzel, U
    MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 877 - 884
  • [24] Measurement of the Energy Resolution of Silicon X-Ray Detectors Using Absorption Edge Spectra
    S. M. Osadchii
    A. A. Petukhov
    V. B. Dunin
    Measurement Techniques, 2019, 62 : 465 - 469
  • [25] Measurement of the Energy Resolution of Silicon X-Ray Detectors Using Absorption Edge Spectra
    Osadchii, S. M.
    Petukhov, A. A.
    Dunin, V. B.
    MEASUREMENT TECHNIQUES, 2019, 62 (05) : 465 - 469
  • [26] Modulation transfer function for a digital micromirror device
    Hsieh, Mei-Li
    OPTICAL ENGINEERING, 2006, 45 (03)
  • [27] Microfabricated transition-edge x-ray detectors
    Hilton, GC
    Martinis, JM
    Irwin, KD
    Bergren, NF
    Wollman, DA
    Huber, ME
    Deiker, S
    Nam, SW
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2001, 11 (01) : 739 - 742
  • [28] AN X-RAY OPAQUE GRID FOR ACCURATE MEASUREMENT FROM RADIOGRAPHS
    PHILIPSON, JAM
    STEPHENSON, CH
    MOULD, AJ
    DEW, DR
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1982, 64 (02): : 250 - 250
  • [29] Wavefront transformation and the modulation transfer function of x-ray multilayer mirrors
    Artioukov, IA
    Fechtchenko, RM
    Vinogradov, AV
    JOURNAL OF OPTICS A-PURE AND APPLIED OPTICS, 2002, 4 (03): : 233 - 236
  • [30] A method for determining the information capacity of x-ray imaging scintillator detectors by means of luminescence and modulation transfer function measurements
    Kandarakis, I
    Cavouras, D
    Kanellopoulos, E
    Nomicos, CD
    Panayiotakis, GS
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1999, 37 (01) : 25 - 30