Accuracy of Blood Glucose Meters for Self-Monitoring Affects Glucose Control and Hypoglycemia Rate in Children and Adolescents with Type 1 Diabetes

被引:17
|
作者
Boettcher, Claudia [1 ]
Dost, Axel [2 ]
Wudy, Stefan A. [1 ]
Flechtner-Mors, Marion [3 ]
Borkenstein, Martin [4 ]
Schiel, Ralf [5 ]
Weitzel, Dieter [6 ]
Bechtold-Dalla Pozza, Susanne [7 ]
Wolf, Johannes [8 ]
Holl, Reinhard W. [3 ]
机构
[1] Univ Giessen, Div Paediat Endocrinol & Diabetol, Ctr Child & Adolescent Med, D-35392 Giessen, Germany
[2] Univ Hosp Jena, Dept Paediat, Jena, Germany
[3] Univ Ulm, Inst Epidemiol & Med Biometry, D-89069 Ulm, Germany
[4] Med Univ Graz, Dept Paediat, Graz, Austria
[5] MEDIGREIF Inselklin Heringsdorf, Dept Diabet & Metab Dis, Ostseebad Heringsdorf, Germany
[6] Childrens & Adolescents Outpatient Ctr, German Clin Diagnost, Wiesbaden, Germany
[7] Univ Munich, Paediat Endocrinol & Diabetol, Munich, Germany
[8] St Vincent Childrens Hosp, Paderborn, Germany
关键词
COGNITIVE FUNCTION; GLYCEMIC CONTROL; SYSTEMS; INSULIN; COMPLICATIONS; MULTICENTER; IMPACT;
D O I
10.1089/dia.2014.0262
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims/Hypothesis: This study investigated the accuracy of blood glucose meters for self-monitoring and its influence on glycated hemoglobin (HbA(1c)) levels and the frequency of hypoglycemic coma. Materials and Methods: Self-measured and simultaneously obtained laboratory blood glucose values from 9,163 patients with type 1 diabetes <18 years of age in the German/Austrian Diabetes Prospective Documentation Initiative registry were analyzed by investigating their compliance with the International Organization for Standardization (ISO) criteria (versions 2003 and 2013) and by error grid analyses. Regression models elucidated effects on glucose control and hypoglycemia rates. Results: Depending on the respective subgroup (defined by sex, age, duration of diabetes, mode of insulin therapy), 78.7-94.7% of the self-monitoring of blood glucose (SMBG) values met the old and 79.7-88.6% met the new ISO criteria. In Clarke and Parkes error grid analyses, the percentages of SMBG values in Zone A ranged between 92.8% and 94.6% (Clarke) and between 92.2% and 95.0% (Parkes). The patient group with SMBG devices measuring "far too low" (compared with the laboratory-obtained glucose levels) presented with a higher HbA(1c) level than those measuring "far too high," "too high," "identical/almost identical," or "too low" (based on quintiles of deviation). Performing "far too high" was associated with the highest rate of hypoglycemic coma in comparison with the other deviation quintiles. Conclusions: This study showed that current SMBG devices fulfilled neither the previous nor the new ISO criteria. Large deviations of the SMBG values from the "true" glucose levels resulted in higher HbA(1c) levels and markedly increased rates of hypoglycemic events.
引用
收藏
页码:275 / 282
页数:8
相关论文
共 50 条
  • [1] Blood-glucose self-monitoring in children and adolescents with type 1 diabetes mellitus - Glucose meters and their accuracy
    Schiel, R.
    Kramer, G.
    Beltschikow, W.
    DIABETOLOGIE UND STOFFWECHSEL, 2007, 2 (02) : 102 - 107
  • [2] Determinants of Self-monitoring of Blood Glucose in Iranian Children and Adolescents with Type 1 Diabetes
    Fayyaz, Farimah
    Khosravi, Sepehr
    Mosallanejad, Asieh
    Tabatabaei-Malazy, Ozra
    Nazari, Seyed Saeed Hashemi
    Shaghaghi, Maede
    INTERNATIONAL JOURNAL OF ENDOCRINOLOGY AND METABOLISM, 2023, 21 (04)
  • [3] Self-monitoring of blood glucose in treatment of type 2 diabetes
    Szymborska-Kajanek, Aleksandra
    Psurek, Anna
    Hese, Robert
    Strojek, Krzysztof
    DIABETES RESEARCH AND CLINICAL PRACTICE, 2009, 86 : S49 - S52
  • [4] Self-monitoring of blood glucose in type 2 diabetes
    Lowe, Julia
    AUSTRALIAN PRESCRIBER, 2010, 33 (05) : 138 - 140
  • [5] Self-monitoring of blood glucose in diabetes: is it worth it?
    O'Kane, Maurice J.
    Pickup, John
    ANNALS OF CLINICAL BIOCHEMISTRY, 2009, 46 : 273 - 282
  • [6] Flash glucose monitoring in type 1 diabetes: A comparison with self-monitoring blood glucose
    Babaya, Naru
    Noso, Shinsuke
    Hiromine, Yoshihisa
    Taketomo, Yasunori
    Niwano, Fumimaru
    Yoshida, Sawa
    Yasutake, Sara
    Kawabata, Yumiko
    Ikegami, Hiroshi
    JOURNAL OF DIABETES INVESTIGATION, 2020, 11 (05) : 1222 - 1229
  • [7] A history of blood glucose meters and their role in self-monitoring of diabetes mellitus
    Clarke, S. E.
    Foster, J. R.
    BRITISH JOURNAL OF BIOMEDICAL SCIENCE, 2012, 69 (02) : 83 - 93
  • [8] Comparing Real-Time Continuous Glucose Monitoring to Self-Monitoring of Blood Glucose: Advantages and Limitations for Children and Adolescents With Type 1 Diabetes
    Aljuhani, Reyof
    Adas, Merna
    Alnaami, Raghad
    Alshehri, Reham
    Alqarni, Rahaf
    NoorSaeed, Sundus
    Al-Agha, Abdulmoein
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (01)
  • [9] Self-monitoring of blood glucose for type 2 diabetes
    Bloomgarden, Zachary
    Ning, Guang
    JOURNAL OF DIABETES, 2015, 7 (05) : 593 - 594
  • [10] Systematic Review of Self-Monitoring of Blood Glucose in Patients With Type 2 Diabetes
    Chircop, James
    Sheffield, David
    Kotera, Yasuhiro
    NURSING RESEARCH, 2021, 70 (06) : 487 - 497