Quality assurance methodology for Varian RapidArc treatment plans

被引:31
作者
Iftimia, Ileana [1 ,2 ]
Cirino, Eileen T. [1 ]
Xiong, Li [1 ]
Mower, Herbert W. [1 ,2 ]
机构
[1] Lahey Clin Fdn, Dept Radiat Oncol, Burlington, MA 01805 USA
[2] Tufts Univ, Sch Med, Boston, MA 02111 USA
关键词
patient-specific RapidArc QA; MapCHECK/MapPHAN 5 2D diode array; Exradin A16 ion chamber; QUASAR phantom; RadCalc MU calculation; MODULATED ARC THERAPY; CLINICAL IMPLEMENTATION; IMAT;
D O I
10.1120/jacmp.v11i4.3164
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
With the commercial introduction of the Varian RapidArc, a new modality for treatment planning and delivery, the need has arisen for consistent and efficient techniques for performing patient-specific quality assurance (QA) tests. In this paper we present our methodology for a RapidArc treatment plan QA procedure. For our measurements we used a 2D diode array (MapCHECK) embedded at 5 cm water equivalent depth in MapPHAN 5 phantom and an Exradin A16 ion chamber placed in six different positions in a cylindrical homogeneous phantom (QUASAR). We also checked the MUs for the RapidArc plans by using independent software (RadCalc). The agreement between Eclipse calculations and MapCHECK/MapPHAN 5 measurements was evaluated using both absolute distance-to-agreement (DTA) and gamma index with 10% dose threshold (TH), 3% dose difference (DD), and 3 mm DTA. The average agreement was 94.4% for the DTA approach and 96.3% for the gamma index approach. In high-dose areas, the discrepancy between calculations and ion chamber measurements using the QUASAR phantom was within 4.5% for prostate cases. For the RadCalc calculations, we used the average SSD along the arc; however, for some patients the agreement for the MUs obtained with RadCalc versus Eclipse was inadequate (discrepancy > 5%). In these cases, the plan was divided into partial arc plans so that RadCalc could perform a better estimation of the MUs. The discrepancy was further reduced to within similar to 4% using this approach. Regardless of the variation in prescribed dose and location of the treated areas, we obtained very good results for all patients studied in this paper.
引用
收藏
页码:130 / 143
页数:14
相关论文
共 21 条
[1]   Portal Imaging for Volumetric Modulated Arc Therapy [J].
Burman, C. ;
Li, J. ;
Chan, M. .
MEDICAL PHYSICS, 2009, 36 (06)
[2]   Dosimetry audit for a multi-centre IMRT head and neck trial [J].
Clark, Catharine H. ;
Hansen, Vibeke Nordmark ;
Chantler, Hannah ;
Edwards, Craig ;
James, Hayley V. ;
Webster, Gareth ;
Miles, Elizabeth A. ;
Urbano, M. Teresa Guerrero ;
Bhide, Shree A. ;
Bidmead, A. Margaret ;
Nutting, Christoper M. .
RADIOTHERAPY AND ONCOLOGY, 2009, 93 (01) :102-108
[3]   Whole abdominopelvic radiotherapy (WAPRT) using intensity-modulated arc therapy (IMAT): First clinical experience [J].
Duthoy, W ;
De Gersem, W ;
Vergote, K ;
Coghe, M ;
Boterberg, T ;
De Deene, Y ;
De Wagter, C ;
Van Belle, S ;
De Neve, W .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 57 (04) :1019-1032
[4]   Clinical implementation of intensity-modulated arc therapy (IMAT) for rectal cancer [J].
Duthoy, W ;
De Gersem, W ;
Vergote, K ;
Boterberg, T ;
Derie, C ;
Smeets, P ;
De Wagter, C ;
De Neve, W .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (03) :794-806
[5]   Task Group 142 report: Quality assurance of medical accelerators [J].
Klein, Eric E. ;
Hanley, Joseph ;
Bayouth, John ;
Yin, Fang-Fang ;
Simon, William ;
Dresser, Sean ;
Serago, Christopher ;
Aguirre, Francisco ;
Ma, Lijun ;
Arjomandy, Bijan ;
Liu, Chihray ;
Sandin, Carlos ;
Holmes, Todd .
MEDICAL PHYSICS, 2009, 36 (09) :4197-4212
[6]   Dosimetric verification of RapidArc treatment delivery [J].
Korreman, Stine ;
Medin, Joakim ;
Kjaer-Kristoffersen, Flemming .
ACTA ONCOLOGICA, 2009, 48 (02) :185-191
[7]  
KUTCHER JG, 1994, MED PHYS, V21, P581
[8]   Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy [J].
Letourneau, Daniel ;
Publicover, Julia ;
Kozelka, Jakub ;
Moseley, Douglas J. ;
Jaffray, David A. .
MEDICAL PHYSICS, 2009, 36 (05) :1813-1821
[9]   RapidArc - Commissioning and QA procedures [J].
Ling, C. ;
Zhang, P. ;
Archambault, Y. ;
Bocanek, J. ;
Tang, G. ;
LoSasso, T. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01) :S594-S594
[10]   A technique for the quantitative evaluation of dose distributions [J].
Low, DA ;
Harms, WB ;
Mutic, S ;
Purdy, JA .
MEDICAL PHYSICS, 1998, 25 (05) :656-661