A qualitative and quantitative analysis of radiation dose and image quality of computed tomography images using adaptive statistical iterative reconstruction

被引:15
作者
Hussain, Fahad Ahmed [1 ]
Mail, Noor [1 ,3 ]
Shamy, Abdulrahman M. [1 ,2 ]
Alghamdi, Suliman [1 ,3 ]
Saoudi, Abdelhamid [1 ,3 ]
机构
[1] King Abdul Aziz Med City, Dept Oncol, Jeddah 21321, Mekkah, Saudi Arabia
[2] King Saud bin Abdulaziz Univ Hlth Sci, Jeddah, Mekkah, Saudi Arabia
[3] King Abdullah Int Med Res Ctr, Jeddah, Saudi Arabia
关键词
computerized tomography; ERUS; EUS; multidetector computerized tomography; transrectal; X-ray; CT; FEASIBILITY; EXPERIENCE; REDUCTION; ALGORITHM; CHEST;
D O I
10.1120/jacmp.v17i3.5903
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Image quality is a key issue in radiology, particularly in a clinical setting where it is important to achieve accurate diagnoses while minimizing radiation dose. Some computed tomography (CT) manufacturers have introduced algorithms that claim significant dose reduction. In this study, we assessed CT image quality produced by two reconstruction algorithms provided with GE Healthcare's Discovery 690 Elite positron emission tomography (PET) CT scanner. Image quality was measured for images obtained at various doses with both conventional filtered back-projection (FBP) and adaptive statistical iterative reconstruction (ASIR) algorithms. A standard CT dose index (CTDI) phantom and a pencil ionization chamber were used to measure the CT dose at 120 kVp and an exposure of 260 mAs. Image quality was assessed using two phantoms. CT images of both phantoms were acquired at tube voltage (kV) of 120 with exposures ranging from 25 mAs to 400 mAs. Images were reconstructed using FBP and ASIR ranging from 10% to 100%, then analyzed for noise, low-contrast detectability, contrast-to-noise ratio (CNR), and modulation transfer function (MTF). Noise was 4.6 HU in water phantom images acquired at 260 mAs/FBP 120 kV and 130 mAs/50% ASIR 120 kV. The large objects (frequency < 7 lp/cm) retained fairly acceptable image quality at 130 mAs/50% ASIR, compared to 260 mAs/FBP. The application of ASIR for small objects (frequency > 7 lp/cm) showed poor visibility compared to FBP at 260 mAs and even worse for images acquired at less than 130 mAs. ASIR blending more than 50% at low dose tends to reduce contrast of small objects (frequency > 7 lp/cm). We concluded that dose reduction and ASIR should be applied with close attention if the objects to be detected or diagnosed are small (frequency > 7 lp/cm). Further investigations are required to correlate the small objects (frequency > 7 lp/cm) to patient anatomy and clinical diagnosis.
引用
收藏
页码:419 / 432
页数:14
相关论文
共 15 条
[1]  
Droege RT, 1982, MED PHYS, V9
[2]   Reducing the Radiation Dose for CT Colonography Using Adaptive Statistical Iterative Reconstruction: A Pilot Study [J].
Flicek, Kristina T. ;
Hara, Amy K. ;
Silva, Alvin C. ;
Wu, Qing ;
Peter, Mary B. ;
Johnson, C. Daniel .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (01) :126-131
[3]   Iterative Reconstruction Technique for Reducing Body Radiation Dose at CT: Feasibility Study [J].
Hara, Amy K. ;
Paden, Robert G. ;
Silva, Alvin C. ;
Kujak, Jennifer L. ;
Lawder, Holly J. ;
Pavlicek, William .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 193 (03) :764-771
[4]   Comparing five different iterative reconstruction algorithms for computed tomography in an ROC study [J].
Jensen, Kristin ;
Martinsen, Anne Catrine T. ;
Tingberg, Anders ;
Aalokken, Trond Mogens ;
Fosse, Erik .
EUROPEAN RADIOLOGY, 2014, 24 (12) :2989-3002
[5]  
Kak AC., 1984, Digital Image Processing Techniques
[6]   The influence of bowtie filtration on cone-beam CT image quality [J].
Mail, N. ;
Moseley, D. J. ;
Siewerdsen, J. H. ;
Jaffray, D. A. .
MEDICAL PHYSICS, 2009, 36 (01) :22-32
[7]   Low-Tube-Voltage, High-Tube-Current Multidetector Abdominal CT: Improved Image Quality and Decreased Radiation Dose with Adaptive Statistical Iterative Reconstruction Algorithm-Initial Clinical Experience [J].
Marin, Daniele ;
Nelson, Rendon C. ;
Schindera, Sebastian T. ;
Richard, Samuel ;
Youngblood, Richard S. ;
Yoshizumi, Terry T. ;
Samei, Ehsan .
RADIOLOGY, 2010, 254 (01) :145-153
[8]   The phantom portion of the American College of Radiology (ACR) Computed Tomography (CT) accreditation program: Practical tips, artifact examples, and pitfalls to avoid [J].
McCollough, CH ;
Bruesewitz, MR ;
McNitt-Gray, MF ;
Bush, K ;
Ruckdeschel, T ;
Payne, JT ;
Brink, JA ;
Zeman, RK .
MEDICAL PHYSICS, 2004, 31 (09) :2423-2442
[9]   Upgrade to Iterative Image Reconstruction (IR) in Abdominal MDCT Imaging: A Clinical Study for Detailed Parameter Optimization Beyond Vendor Recommendations Using the Adaptive Statistical Iterative Reconstruction Environment (ASIR) [J].
Mueck, F. G. ;
Koerner, M. ;
Scherr, M. K. ;
Geyer, L. L. ;
Deak, Z. ;
Linsenmaier, U. ;
Reiser, M. ;
Wirth, S. .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2012, 184 (03) :229-238
[10]   Diffuse Lung Disease: CT of the Chest with Adaptive Statistical Iterative Reconstruction Technique [J].
Prakash, Priyanka ;
Kalra, Mannudeep K. ;
Ackman, Jeanne B. ;
Digumarthy, Subba R. ;
Hsieh, Jiang ;
Do, Synho ;
Shepard, Jo-Anne O. ;
Gilman, Matthew D. .
RADIOLOGY, 2010, 256 (01) :261-269