Computation of realistic virtual phantom images for an objective lesion detectability assessment in digital mammography

被引:3
作者
Perez-Ponce, Hector [1 ,2 ]
Daul, Christian [1 ]
Wolf, Didier [1 ]
Noel, Alain [1 ,2 ]
机构
[1] Nancy Univ, Ctr Rech Automat Nancy CRAN UMR 7039, CNRS, F-54516 Vandoeuvre Les Nancy, France
[2] Ctr Alexis Vautrin, Dept Med Phys, F-54500 Vandoeuvre Les Nancy, France
关键词
Digital mammography; X-ray source/digital detector model; Lesion detectability; Image quality factor; Contrast detail curves; CDMAM-phantom image simulation; Contrast-detail analysis; QUALITY CHARACTERIZATION; IMAGING-SYSTEMS; NOISE; DETECTORS; SPECTRUM;
D O I
10.1016/j.medengphy.2011.06.004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Image quality assessment is required for an optimal use of mammographic units. On the one hand, there are objective image quality assessment methods based on the measurement of technical parameters such as modulation transfer function (MTF), noise power spectrum (NPS) or detection quantum efficiency (DQE) describing performances of digital detectors. These parameters are, however, without direct relationship with lesion detectability in clinical practice. On the other hand, there are image quality assessment methods involving time consuming procedures, but presenting a direct relationship with lesion detectability. This contribution describes an X-ray source/digital detector model leading to the simulation of virtual contrast-detail phantom (CDMAM) images. The virtual image computation method requires the acquisition of only few real images and allows for an objective image quality assessment presenting a direct relationship with lesion detectability. The transfer function of the proposed model takes as input physical parameters (MTF* and noise) measured under clinical conditions on mammographic units. As presented in this contribution, MTF* is a modified MTF taking into account the effects due to X-ray scatter in the breast and magnification. Results obtained with the structural similarity index prove that the simulated images are quite realistic in terms of contrast and noise. Tests using contrast detail curves highlight the fact that the simulated and real images lead to very similar data quality in terms of lesion detectability. Finally, various statistical tests show that quality factors computed for both the simulated images and the real images are very close for the two data sets. (C) 2011 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1276 / 1286
页数:11
相关论文
共 34 条
[11]  
GROSJEAN B, 2007, THESIS ECOLE CENTRAL, P275
[12]   A-contrario Detectability of Spots in Textured Backgrounds [J].
Grosjean, Benedicte ;
Moisan, Lionel .
JOURNAL OF MATHEMATICAL IMAGING AND VISION, 2009, 33 (03) :313-337
[13]   Detecting film-screen artifacts in mammography using a model-based approach [J].
Highnam, R ;
Brady, M ;
English, R .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1999, 18 (10) :1016-1024
[14]   Quality assurance in mammography: Artifact analysis [J].
Hogge, JP ;
Palmer, CH ;
Muller, CC ;
Little, ST ;
Smith, DC ;
Fatouros, PP ;
de Paredes, ES .
RADIOGRAPHICS, 1999, 19 (02) :503-522
[15]  
ICRU, 1996, 54 ICRU
[16]  
IEC, 2005, 6222012 IEC
[17]  
Karssemeijer N, 1996, INT CONGR SER, V1119, P155
[18]  
Lane D.M., 2007, HYPERSTAT ONLINE STA
[19]  
Lohninger H., 1999, Teach me data analysis
[20]   A comparison between objective and subjective image quality measurements for a full field digital mammography system [J].
Marshall, N. W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (10) :2441-2463