New principle for measuring arterial blood oxygenation, enabling motion-robust remote monitoring

被引:74
作者
van Gastel, Mark [1 ]
Stuijk, Sander [1 ]
de Haan, Gerard [1 ,2 ]
机构
[1] Eindhoven Univ Technol, Dept Elect Engn, Eindhoven, Netherlands
[2] Philips Res, Philips Innovat Grp, Eindhoven, Netherlands
关键词
PULSE OXIMETRY; NONCONTACT; SATURATION; LIGHT; REFLECTANCE; PPG;
D O I
10.1038/srep38609
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Finger-oximeters are ubiquitously used for patient monitoring in hospitals worldwide. Recently, remote measurement of arterial blood oxygenation (SpO(2)) with a camera has been demonstrated. Both contact and remote measurements, however, require the subject to remain static for accurate SpO2 values. This is due to the use of the common ratio-of-ratios measurement principle that measures the relative pulsatility at different wavelengths. Since the amplitudes are small, they are easily corrupted by motion-induced variations. We introduce a new principle that allows accurate remote measurements even during significant subject motion. We demonstrate the main advantage of the principle, i.e. that the optimal signature remains the same even when the SNR of the PPG signal drops significantly due to motion or limited measurement area. The evaluation uses recordings with breath-holding events, which induce hypoxemia in healthy moving subjects. The events lead to clinically relevant SpO(2) levels in the range 80-100%. The new principle is shown to greatly outperform current remote ratio-of-ratios based methods. The mean-absolute SpO(2)-error (MAE) is about 2 percentage-points during head movements, where the benchmark method shows a MAE of 24 percentage-points. Consequently, we claim ours to be the first method to reliably measure SpO(2) remotely during significant subject motion.
引用
收藏
页数:16
相关论文
共 41 条
[1]   Photoplethysmography and its application in clinical physiological measurement [J].
Allen, John .
PHYSIOLOGICAL MEASUREMENT, 2007, 28 (03) :R1-R39
[2]  
[Anonymous], 2011, PART REQ BAS SAF ESS
[3]  
Aoyagi T., 1974, Abstracts of the 13th annual meeting of the Japanese Society of Medical Electronics and Biological Engineering, P90
[4]   Non-contact estimation of heart rate and oxygen saturation using ambient light [J].
Bal, Ufuk .
BIOMEDICAL OPTICS EXPRESS, 2015, 6 (01) :86-97
[5]   Effects of skin pigmentation on pulse oximeter accuracy at low saturation [J].
Bickler, PE ;
Feiner, JR ;
Severinghaus, JW .
ANESTHESIOLOGY, 2005, 102 (04) :715-719
[6]   Improved motion robustness of remote-PPG by using the blood volume pulse signature [J].
de Haan, G. ;
van Leest, A. .
PHYSIOLOGICAL MEASUREMENT, 2014, 35 (09) :1913-1926
[7]   Robust Pulse Rate From Chrominance-Based rPPG [J].
de Haan, Gerard ;
Jeanne, Vincent .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2013, 60 (10) :2878-2886
[8]  
Eichenfield Lawrence F., 1999, Current Opinion in Pediatrics, V11, P471, DOI 10.1097/00008480-199910000-00017
[9]  
Fanaroff A.A., 2012, Klaus and Fanaroff's care of the high-risk neonate
[10]   A DIFFUSION-THEORY MODEL OF SPATIALLY RESOLVED, STEADY-STATE DIFFUSE REFLECTANCE FOR THE NONINVASIVE DETERMINATION OF TISSUE OPTICAL-PROPERTIES INVIVO [J].
FARRELL, TJ ;
PATTERSON, MS ;
WILSON, B .
MEDICAL PHYSICS, 1992, 19 (04) :879-888