Design and fabrication of a 40-MHz annular army transducer

被引:139
作者
Ketterling, JA [1 ]
Aristizábal, O
Turnbull, DH
Lizzi, FL
机构
[1] Riverside Res Inst, New York, NY USA
[2] NYU, Sch Med, New York, NY USA
[3] Skirball Inst Biomol Med, New York, NY USA
关键词
D O I
10.1109/TUFFC.2005.1428050
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This paper investigates the feasibility of fabricating a five-ring, focused annular array transducer operating at 40 MHz. The active piezoelectric material of the transducer was a 9-mu m thick polyvinylidene fluoride (PVDF) film. One side of the PVDF was metallized with gold and forms the ground plane of the transducer. The array pattern of the transducer and electrical traces to each annulus were formed on a copper-clad polyimide film. The PVDF and polyimide were bonded with a thin layer of epoxy, pressed into a spherically curved shape, then back filled with epoxy. A five-ring transducer with equal area elements and 100-mu m kerfs between annuli was fabricated and tested. The transducer had a total aperture of 6 mm and a geometric focus of 12 mm. The pulse/echo response from a quartz plate located at the geometric focus, two-way insertion loss (IL), complex impedance, electrical crosstalk, and lateral beamwidth all were measured for each annulus. The complex impedance data from each element were used to perform electrical matching, and the measurements were repeated. After impedance matching, f(c) approximate to 36 MHz and -6-dB bandwidths ranged from 31 to 39%. The ILs for the matched annuli ranged from -28 to -38 dB.
引用
收藏
页码:672 / 681
页数:10
相关论文
共 31 条
[1]   High frequency single element and annular array transducers incorporating PVDF [J].
Alves, CHF ;
Snook, K ;
Ritter, T ;
Shung, KK .
MEDICAL IMAGING 2000: ULTRASONIC IMAGING AND SIGNAL PROCESSING, 2000, 3982 :116-121
[2]   AN ANNULAR ARRAY SYSTEM FOR HIGH-RESOLUTION BREAST ECHOGRAPHY [J].
ARDITI, M ;
TAYLOR, WB ;
FOSTER, FS ;
HUNT, JW .
ULTRASONIC IMAGING, 1982, 4 (01) :1-31
[3]   TRANSIENT FIELDS OF CONCAVE ANNULAR ARRAYS [J].
ARDITI, M ;
FOSTER, FS ;
HUNT, JW .
ULTRASONIC IMAGING, 1981, 3 (01) :37-61
[4]   44-MHz LiNbO3 transducers for UBM-guided Doppler ultrasound [J].
Aristizábal, O ;
Turnbull, DH .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2003, 50 (06) :623-630
[5]   Design and fabrication of annular arrays for high-frequency ultrasound [J].
Brown, JA ;
Démoré, CEM ;
Lockwood, GR .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2004, 51 (08) :1010-1017
[6]   A low-cost, high-performance pulse generator for ultrasound imaging [J].
Brown, JA ;
Lockwood, GR .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2002, 49 (06) :848-851
[7]   Design considerations for piezoelectric polymer ultrasound transducers [J].
Brown, LF .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2000, 47 (06) :1377-1396
[8]  
Cannata JM, 2002, ISAF 2002: PROCEEDINGS OF THE 13TH IEEE INTERNATIONAL SYMPOSIUM ON APPLICATIONS OF FERROELECTRICS, P343
[9]   Design of efficient, broadband single-element (20-80 MHz) ultrasonic transducers for medical imaging applications [J].
Cannata, JM ;
Ritter, TA ;
Chen, WH ;
Silverman, RH ;
Shung, KK .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2003, 50 (11) :1548-1557
[10]   Miniature high frequency focused ultrasonic transducers for minimally invasive imaging procedures [J].
Fleischman, A ;
Modi, R ;
Nair, A ;
Talman, J ;
Lockwood, G ;
Roy, S .
SENSORS AND ACTUATORS A-PHYSICAL, 2003, 103 (1-2) :76-82