Monte Carlo and experimental dosimetric study of the mHDR-v2 brachytherapy source

被引:7
作者
Chandola, Rakesh M. [1 ,2 ]
Tiwari, Samit [3 ]
Kowar, Manoj K. [3 ]
Choudhary, Vivek [1 ,2 ]
机构
[1] Pt JNM Med Coll, Dept Radiotherapy, Raipur, Chhattisgarh, India
[2] Dr BRAM Hosp, Raipur, Chhattisgarh, India
[3] Bhilai Inst Technol, Dept Appl Phys, Durg, Chhattisgarh, India
关键词
Dose rate constant; Monte Carlo simulation; radial dose function; 2D anisotropy function; RATE IR-192 SOURCES; AIDED DOSIMETRY; CALIBRATION;
D O I
10.4103/0973-1482.77068
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The conventional treatment planning system (TPS) gives analytical calculations with +/- 15 to 20% dose, which may lead to over exposure of critical organs or under dose of target. It is to obtain dose distribution parameters of nucletron high dose rate (HDR) microselectron v2 (mHDR-v2) (192)Ir brachytherapy source by experiment and by calculated study using Monte Carlo (MC) EGSnrc code, and to find the similarity between them, and with any past study. To validate data, another MC GEANT4 study done in this work on the same source is also presented. Different software of the computer e. g. paint, excel, etc are employed for preparation of figures and graphs. The measured study of the source was done using an in-air ionization chamber, water phantom, and measurement set-up, while the calculated study was done by modeling the set up of the measured study by using the MC EGSnrc and GEANT4. Mean and probability are used in calculation of average values, and calculation of the uncertainties in result and discussion. The measured and calculated values of dose rate constant, radial dose function, and 2D anisotropy function were found to be in agreement with each other as well as with published data. The results of this study can be used as input to TPS.
引用
收藏
页码:421 / 426
页数:6
相关论文
共 21 条
  • [1] Monte Carlo dosimetry of a new 192Ir high dose rate brachytherapy source
    Angelopoulos, A
    Baras, P
    Sakelliou, L
    Karaiskos, P
    Sandilos, P
    [J]. MEDICAL PHYSICS, 2000, 27 (11) : 2521 - 2527
  • [2] [Anonymous], Expression of the Uncertainty of Measurement in Calibration
  • [3] Monte Carlo calculation of dose rate distributions around Ir-192 wires
    Ballester, F
    Hernandez, C
    PerezCalatayud, J
    Lliso, F
    [J]. MEDICAL PHYSICS, 1997, 24 (08) : 1221 - 1228
  • [4] Comparison of calibration procedures for 192IR high-dose-rate brachytherapy sources
    Baltas, D
    Geramani, K
    Ioannidis, GT
    Hierholz, K
    Rogge, B
    Kolotas, C
    Müller-Sievers, K
    Milickovic, N
    Kober, B
    Zamboglou, N
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (03): : 653 - 661
  • [5] Beta versus gamma dosimetry close to Ir-192 brachytherapy sources
    Baltas, D
    Karaiskos, P
    Papagiannis, P
    Sakelliou, L
    Loeffler, E
    Zamboglou, N
    [J]. MEDICAL PHYSICS, 2001, 28 (09) : 1875 - 1882
  • [6] CULLEN DE, 1997, EVALUATED PHOTON DAT
  • [7] Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source
    Daskalov, GM
    Loffler, E
    Williamson, JF
    [J]. MEDICAL PHYSICS, 1998, 25 (11) : 2200 - 2208
  • [8] IAEA, 2002, IAEA-TECDOC-1274, DOI [10.1016/s0167-8140(02)82497-5, DOI 10.1016/S0167-8140(02)82497-5]
  • [9] Monte Carlo and TLD dosimetry of an 192Ir high dose-rate brachytherapy source
    Karaiskos, P
    Angelopoulos, A
    Sakelliou, L
    Sandilos, P
    Antypas, C
    Vlachos, L
    Koutsouveli, E
    [J]. MEDICAL PHYSICS, 1998, 25 (10) : 1975 - 1984
  • [10] Kirov S, 1995, PHYS MED BIOL, V40, P2015