Reproducibility-Repeatability of Choroidal Thickness Calculation Using Optical Coherence Tomography

被引:32
作者
Benavente-Perez, Alexandra [1 ]
Hosking, Sarah L. [2 ,3 ,4 ]
Logan, Nicola S. [2 ]
Bansal, Dheeraj [5 ]
机构
[1] SUNY State Coll Optometry, Dept Biol Sci, New York, NY USA
[2] Aston Univ, Sch Life & Hlth Sci, Ophthalm Res Grp, Birmingham B4 7ET, W Midlands, England
[3] City Univ London, Dept Optometry & Visual Sci, London, England
[4] Univ Melbourne, Dept Ophthalmol, Melbourne, Vic, Australia
[5] InnZ Med AB, Ljungskile, Sweden
关键词
choroidal thickness; optical coherence tomography; in vivo manual measurement; axial length; FIBER LAYER THICKNESS; AXIAL LENGTH; OCULAR BIOMETRY; MYOPIA; CHICKS; CHILDREN; OCT; AGE;
D O I
10.1097/OPX.0b013e3181f3eced
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To evaluate the repeatability and reproducibility of subfoveal choroidal thickness (CT) calculations performed manually using optical coherence tomography (OCT). Methods. The CT was imaged in vivo at each of two visits on 11 healthy volunteers (mean age, 35.72 +/- 13.19 years) using the spectral domain OCT. CT was manually measured after applying ImageJ processing filters on 15 radial subfoveal scans. Each radial scan was spaced 12 degrees from each other and contained 2500 A-scans. The coefficient of variability, coefficient of repeatability (CoR), coefficient of reproducibility, and intraclass correlation coefficient determined the reproducibility and repeatability of the calculation. Axial length (AL) and mean spherical equivalent refractive error were measured with the IOLMaster and an open view autorefractor to study their potential relationship with CT. Results. The within-visit and between-visit coefficient of variability, CoR, coefficient of reproducibility, and intraclass correlation coefficient were 0.80, 2.97% 2.44%, and 99%, respectively. The subfoveal CT correlated significantly with AL (R = -0.60, p = 0.05). Conclusions. The subfoveal CT could be measured manually in vivo using OCT and the readings obtained from the healthy subjects evaluated were repeatable and reproducible. It is proposed that OCT could be a useful instrument to perform in vivo assessment and monitoring of CT changes in retinal disease. The preliminary results suggest a negative correlation between subfoveal CT and AL in such a way that it decreases with increasing AL but not with refractive error. (Optom Vis Sci 2010;87:867-872)
引用
收藏
页码:867 / 872
页数:6
相关论文
共 28 条
[1]   In Vivo Human Choroidal Thickness Measurements: Evidence for Diurnal Fluctuations [J].
Brown, Jamin S. ;
Flitcroft, D. Ian ;
Ying, Gui-shuang ;
Francis, Ellie L. ;
Schmid, Gregor F. ;
Quinn, Graham E. ;
Stone, Richard A. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2009, 50 (01) :5-12
[2]   Determinants of normal retinal nerve fiber layer thickness measured by stratus OCT [J].
Budenz, Donald L. ;
Anderson, Douglas R. ;
Varma, Rohit ;
Schuman, Joel ;
Cantor, Louis ;
Savell, Jonathan ;
Greenfeld, David S. ;
Patella, Vincent Michael ;
Quigley, Harry A. ;
Tielsch, James .
OPHTHALMOLOGY, 2007, 114 (06) :1046-1052
[3]   Temporal relationship of choroidal blood flow and thickness changes during recovery from form deprivation myopia in chicks [J].
Fitzgerald, MEC ;
Wildsoet, CF ;
Reiner, A .
EXPERIMENTAL EYE RESEARCH, 2002, 74 (05) :561-570
[4]   Enhanced Depth Imaging Optical Coherence Tomography of the Choroid in Highly Myopic Eyes [J].
Fujiwara, Takamitsu ;
Imamura, Yutaka ;
Margolis, Ron ;
Slakter, Jason S. ;
Spaide, Richard F. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 148 (03) :445-450
[5]   Myopia: precedents for research in the twenty-first century [J].
Gilmartin, B .
CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2004, 32 (03) :305-324
[6]   REFRACTIVE ERROR, AXIAL LENGTH, AND HEIGHT AS A FUNCTION OF AGE IN YOUNG MYOPES [J].
GOSS, DA ;
COX, VD ;
HERRINLAWSON, GA ;
NIELSEN, ED ;
DOLTON, WA .
OPTOMETRY AND VISION SCIENCE, 1990, 67 (05) :332-338
[7]   COMPARISON OF REFRACTIVE COMPONENTS IN YOUTH-ONSET AND EARLY ADULT-ONSET MYOPIA [J].
GROSVENOR, T ;
SCOTT, R .
OPTOMETRY AND VISION SCIENCE, 1991, 68 (03) :204-209
[8]  
Gwiazda J, 2002, INVEST OPHTH VIS SCI, V43, P314
[9]   Modelling the normal retinal nerve fibre layer thickness as measured by Stratus optical coherence tomography [J].
Hougaard, Jesper Leth ;
Ostenfeld, Carl ;
Heijl, Anders ;
Bengtsson, Boel .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2006, 244 (12) :1607-1614
[10]   Optical coherence tomography: An introduction to the technique and its use [J].
Hrynchak, P ;
Simpson, T .
OPTOMETRY AND VISION SCIENCE, 2000, 77 (07) :347-356