Streak artefacts in Computed Tomography Urography: Comparative evaluation between two different iodinated contrast media

被引:1
作者
Pedote, Pasquale [1 ]
Cozzolino, Monica [1 ]
Nicolardi, Martina [1 ]
Accogli, Rocco Umberto [1 ]
Sardaro, Angela [1 ]
Scardapane, Arnaldo [1 ]
Ianora, Amato Antonio Stabile [1 ]
机构
[1] Univ Bari Aldo Moro, Interdisciplinary Dept Med, Sect Diagnost Imaging, Bari, Italy
关键词
artefacts; contrast media; Iodixanol; Iomeprol; urinary tract; CT; DIAGNOSIS; IOPAMIDOL;
D O I
10.1111/1754-9485.13363
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction The objective of this study was to compare the frequency and entity, in computed tomography (CT) urography, of streak artefacts on the urinary tract generated by two contrast agents with a different iodine concentration and osmolarity. Methods Computed tomography scans including an excretory renal phase, performed on adult subjects in the period May-July 2020, were retrospectively evaluated in consensus by three expert radiologists, to detect any streak artefacts located in the urinary tract. Patients were administered either 1.6 mL/kg of Iodixanol 320 mgI/mL or 1.3 mL/kg of Iomeprol 400 mgI/mL. Results In total, 144 CT scans were analysed, subdivided into two groups administered either Iodixanol (71/144 (49.3%) patients) or Iomeprol (73/144 (50.7%) patients). In 41% cases, no beam hardening artefacts were found; among these, 12/59 (20.3%) patients had received Iodixanol and 47/59 (79.7%) Iomeprol. In the Iodixanol group, the mean contrast density on the renal pelvis was 2565.6 HU and streak artefacts occurred in 59/71 cases (83.1%); in 33/59 (55.9%) cases, the artefacts were marked, and in 26/59 (44.1%) minimal. In the Iomeprol group, the mean contrast density on the renal pelvis was 1666 HU and streak artefacts occurred in 26/73 cases (35.6%); in 7/26 (27%) cases, the artefacts were marked and in 19/26 (73%) minimal. Conclusion The study data demonstrate a significant difference in the attenuation values of iodine urine in the excretory system between the Iodixanol and Iomeprol group. Iodixanol induced a higher frequency and burden of artefacts, compared to Iomeprol.
引用
收藏
页码:940 / 945
页数:6
相关论文
共 45 条
  • [21] CARDIORESPIRATORY RESPONSES ELICITED BY RIGHT ATRIAL INJECTIONS OF IODINATED CONTRAST-MEDIA - COMPARATIVE-EVALUATION OF 4 AGENTS
    MORETTIN, LB
    ZHAN, XX
    SANTAMBROGIO, FB
    SANTAMBROGIO, G
    INVESTIGATIVE RADIOLOGY, 1994, 29 (02) : 201 - 209
  • [22] Comparison of different iodine concentration contrast media in perfusion computed tomography of the brain - Is high iodine concentration useful?
    Kloska, Stephan P.
    Fischer, Tobias
    Nabavi, Darius G.
    Wessling, Johannes
    Dittrich, Ralf
    Fischbach, Roman
    Seidensticker, Peter
    Ringelstein, E. Bernd
    Heindel, Walter
    INVESTIGATIVE RADIOLOGY, 2007, 42 (08) : 564 - 568
  • [23] Low kV Computed Tomography of Parenchymal Abdominal Organs-A Systematic Animal Study of Different Contrast Media Injection Protocols
    Overhoff, Daniel
    Jost, Gregor
    McDermott, Michael
    Weber, Olaf
    Pietsch, Hubertus
    Schoenberg, Stefan O.
    Attenberger, Ulrike
    TOMOGRAPHY, 2021, 7 (04) : 815 - 828
  • [24] Comparative Diagnostic Evaluation with Contrast-Enhanced Ultrasound, Computed Tomography and Magnetic Resonance Imaging in Patients with Pancreatic Cystic Neoplasms
    Sun, Ya
    Yang, Shuo
    Qi, Erpeng
    Liu, Fangyi
    Zhou, Fubo
    Lu, Yuhan
    Liang, Ping
    Ye, Huiyi
    Yu, Xiaoling
    CANCER MANAGEMENT AND RESEARCH, 2020, 12 : 2889 - 2898
  • [25] Differentiation between tuberculosis and leukemia in abdominal and pelvic lymph nodes: evaluation with contrast-enhanced multidetector computed tomography
    Zhang, Ge
    Yang, Zhi-gang
    Yao, Jin
    Deng, Wen
    Zhang, Shuai
    Xu, Hua-yan
    Long, Qi-hua
    CLINICS, 2015, 70 (03) : 162 - 168
  • [26] Multidetector computed tomography (MDCT) evaluation of myocardial viability: intraindividual comparison of monomeric vs. dimeric contrast media in a rabbit model
    Mahnken, Andreas H.
    Jost, Gregor
    Bruners, Philipp
    Sieber, Martin
    Seidensticker, Peter R.
    Guenther, Rolf W.
    Pietsch, Hubertus
    EUROPEAN RADIOLOGY, 2009, 19 (02) : 290 - 297
  • [27] Evaluation of metal artefact techniques with same contrast scale for different commercially available dual-energy computed tomography scanners
    Kawahara, Daisuke
    Ozawa, Shuichi
    Yokomachi, Kazushi
    Higaki, Toru
    Shiinoki, Takehiro
    Fujioka, Chikako
    Saito, Akito
    Miki, Kentaro
    Kimura, Tomoki
    Murakami, Yuji
    Awai, Kazuo
    Nagata, Yasushi
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (02) : 539 - 546
  • [28] Multidetector computed tomography (MDCT) evaluation of myocardial viability: intraindividual comparison of monomeric vs. dimeric contrast media in a rabbit model
    Andreas H. Mahnken
    Gregor Jost
    Philipp Bruners
    Martin Sieber
    Peter R. Seidensticker
    Rolf W. Günther
    Hubertus Pietsch
    European Radiology, 2009, 19 : 290 - 297
  • [29] Comparative evaluation of the accuracy of linear measurements between cone beam computed tomography and 3D microtomography
    Mangione, Francesca
    Meleo, Deborah
    Talocco, Marco
    Pecci, Raffaella
    Pacifici, Luciano
    Bedini, Rossella
    ANNALI DELL ISTITUTO SUPERIORE DI SANITA, 2013, 49 (03): : 261 - 265
  • [30] Comparative Evaluation of Modified Canal Staining and Clearing Technique, Cone-Beam Computed Tomography, Peripheral Quantitative Computed Tomography, Spiral Computed Tomography, and Plain and Contrast Medium-enhanced Digital Radiography in Studying Root Canal Morphology
    Neelakantan, Prasanna
    Subbarao, Chandana
    Subbarao, Chandragiri V.
    JOURNAL OF ENDODONTICS, 2010, 36 (09) : 1547 - 1551