Streak artefacts in Computed Tomography Urography: Comparative evaluation between two different iodinated contrast media

被引:1
|
作者
Pedote, Pasquale [1 ]
Cozzolino, Monica [1 ]
Nicolardi, Martina [1 ]
Accogli, Rocco Umberto [1 ]
Sardaro, Angela [1 ]
Scardapane, Arnaldo [1 ]
Ianora, Amato Antonio Stabile [1 ]
机构
[1] Univ Bari Aldo Moro, Interdisciplinary Dept Med, Sect Diagnost Imaging, Bari, Italy
关键词
artefacts; contrast media; Iodixanol; Iomeprol; urinary tract; CT; DIAGNOSIS; IOPAMIDOL;
D O I
10.1111/1754-9485.13363
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction The objective of this study was to compare the frequency and entity, in computed tomography (CT) urography, of streak artefacts on the urinary tract generated by two contrast agents with a different iodine concentration and osmolarity. Methods Computed tomography scans including an excretory renal phase, performed on adult subjects in the period May-July 2020, were retrospectively evaluated in consensus by three expert radiologists, to detect any streak artefacts located in the urinary tract. Patients were administered either 1.6 mL/kg of Iodixanol 320 mgI/mL or 1.3 mL/kg of Iomeprol 400 mgI/mL. Results In total, 144 CT scans were analysed, subdivided into two groups administered either Iodixanol (71/144 (49.3%) patients) or Iomeprol (73/144 (50.7%) patients). In 41% cases, no beam hardening artefacts were found; among these, 12/59 (20.3%) patients had received Iodixanol and 47/59 (79.7%) Iomeprol. In the Iodixanol group, the mean contrast density on the renal pelvis was 2565.6 HU and streak artefacts occurred in 59/71 cases (83.1%); in 33/59 (55.9%) cases, the artefacts were marked, and in 26/59 (44.1%) minimal. In the Iomeprol group, the mean contrast density on the renal pelvis was 1666 HU and streak artefacts occurred in 26/73 cases (35.6%); in 7/26 (27%) cases, the artefacts were marked and in 19/26 (73%) minimal. Conclusion The study data demonstrate a significant difference in the attenuation values of iodine urine in the excretory system between the Iodixanol and Iomeprol group. Iodixanol induced a higher frequency and burden of artefacts, compared to Iomeprol.
引用
收藏
页码:940 / 945
页数:6
相关论文
共 45 条
  • [1] Impact of different concentration iodinated contrast media on pain and comfort in abdominal computed tomography
    Sun, Yan-Xia
    Shang, Jing
    Cui, Yong
    Zhang, Ke-Jin
    Li, Xiao-Ting
    Li, Dan-Ni
    Gu, Xiao-Lei
    Tian, Xiao-Yu
    Sun, Nan
    Zhang, Lan
    You, Miao-Ning
    Zhang, Xiao-Yan
    Sun, Ying-Shi
    EUROPEAN JOURNAL OF RADIOLOGY, 2024, 179
  • [2] Personalized application of three different concentrations of iodinated contrast media in coronary computed tomography angiography
    Zhang, Meng
    Hao, Panpan
    Jiang, Chenyu
    Hao, Guoxiang
    Li, Bin
    Hu, Peixin
    Chen, Qingjie
    Chen, Yuguo
    Zhang, Aifeng
    Zhang, Yun
    Liu, Yanping
    JOURNAL OF CELLULAR AND MOLECULAR MEDICINE, 2020, 24 (10) : 5446 - 5453
  • [3] Low tube voltage computed tomography urography using low-concentration contrast media: Comparison of image quality in conventional computed tomography urography
    Hwang, Inpyeong
    Cho, Jeong Yeon
    Kim, Sang Youn
    Oh, Seung-June
    Ku, Ja Hyeon
    Lee, Joongyup
    Kim, Seung Hyup
    EUROPEAN JOURNAL OF RADIOLOGY, 2015, 84 (12) : 2454 - 2463
  • [4] B-mode and Doppler ultrasonography in the renal evaluation of dogs undergoing computed tomography after intravenous administration of different iodinated contrast media
    Martin, Claudia Matsunaga
    Kogika, Marcia Mery
    Miyashiro, Samantha Ive
    Brandao de Campos Fonseca-Pinto, Ana Carolina
    PESQUISA VETERINARIA BRASILEIRA, 2017, 37 (07): : 759 - 772
  • [5] Low Tube Voltage Chest Computed Tomography With Enhancement Using Low-Concentration Iodinated Contrast Media: Comparison of 240 mg/mL Versus 300 mg/mL Iodinated Contrast Media
    Chang, Suyon
    Jung, Jung Im
    Beck, Kyongmin S. S.
    CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2023, 74 (01): : 127 - 136
  • [6] Influence of osmolarity of contrast medium and saline flush on computed tomography angiography: Comparison of monomeric and dimeric iodinated contrast media with different iodine concentrations at an identical iodine delivery rate
    Kishimoto, Miori
    Doi, Shoko
    Shimizu, Junichiro
    Lee, Ki-Ja
    Iwasaki, Toshiroh
    Miyake, Yoh-Ichi
    Yamada, Kazutaka
    EUROPEAN JOURNAL OF RADIOLOGY, 2010, 76 (01) : 135 - 139
  • [7] Comparative Evaluation of the Artefacts Index of Dental Materials on Two-Dimensional Cone-beam Computed Tomography
    Yuan, Fusong
    Chen, Litong
    Wang, Xiaofei
    Wang, Yong
    Lyu, Peijun
    Sun, Yuchun
    SCIENTIFIC REPORTS, 2016, 6
  • [8] Effect of iodinated contrast media on serum electrolyte concentrations in patients undergoing routine contrast computed tomography scan procedure
    Sankaran, Sindhu
    Saharia, Gautom Kumar
    Naik, Suprava
    Mangaraj, Manaswini
    INTERNATIONAL JOURNAL OF APPLIED AND BASIC MEDICAL RESEARCH, 2019, 9 (04) : 217 - 220
  • [9] Evaluation of Two Sinogram Interpolation Methods for Metal Artefacts Reduction in Computed Tomography
    Osman, Noor Diyana
    Sobri, Nurul Fathin Mohamad
    Achuthan, Anusha
    Saidun, Halimatul Asma
    Aziz, Mohd Zahri Abdul
    Shuaib, Ibrahim Lutfi
    2018 IEEE-EMBS CONFERENCE ON BIOMEDICAL ENGINEERING AND SCIENCES (IECBES), 2018, : 137 - 139
  • [10] Quantitative distribution of iodinated contrast media in body computed tomography: data from a large reference cohort
    David Zopfs
    Josefine Graffe
    Robert Peter Reimer
    Sebastian Schäfer
    Thorsten Persigehl
    David Maintz
    Jan Borggrefe
    Stefan Haneder
    Simon Lennartz
    Nils Große Hokamp
    European Radiology, 2021, 31 : 2340 - 2348