Undersizing of droplets from a vented nebulizer caused by aerosol heating during transit through an Anderson impactor

被引:53
作者
Finlay, WH [1 ]
Stapleton, KW
机构
[1] Univ Alberta, Dept Mech Engn, Edmonton, AB T6G 2G8, Canada
[2] Pari Aerosol Res Inst, D-81369 Munich, Germany
基金
加拿大自然科学与工程研究理事会;
关键词
D O I
10.1016/S0021-8502(98)00024-X
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
The use of Anderson cascade impaction to measure droplet sizes from a vented nebulizer connected directly to the impactor was examined by performing two sets of impactor measurements. In the first set, the impactor was operated in room temperature air. In the second set, the impactor was immersed in a cooled water bath at the same temperature as the aerosol exiting the nebulizer (10 degrees C). Normal saline was nebulized in five Pari LC Jet + nebulizers driven by a single Pulmo-Aide compressor, operating under ambient conditions of 35% RH and 22 degrees C. Impaction was done once steady temperatures were reached. When the impactor was operated in room temperature air, the air travelling through the impactor warmed from 10 degrees C at the entrance to the impactor to room temperature at the exit of the impactor. This heating resulted in significant droplet shrinkage due to humidification by evaporation from the droplets, since the impactor immersed in the cooled water bath gave an MMAD that was on average 70% larger than was obtained when the impactor was operated in room air (3.4 mu m vs 2.1 mu m). These results emphasize the need for caution when using impactors to measure nebulized hygroscopic aerosols, since even if these aerosols enter the impactor in vapor pressure equilibrium with their surrounding air, significant size changes can occur during transit through the impactor if the temperature of the aerosol differs significantly from that surrounding the impactor. (C) 1998 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:105 / 109
页数:5
相关论文
共 9 条
[1]   DISTORTION OF SIZE DISTRIBUTIONS BY CONDENSATION AND EVAPORATION IN AEROSOL INSTRUMENTS [J].
BISWAS, P ;
JONES, CL ;
FLAGAN, RC .
AEROSOL SCIENCE AND TECHNOLOGY, 1987, 7 (02) :231-246
[2]   Errors in regional lung deposition predictions of nebulized salbutamol sulphate due to neglect or partial inclusion of hygroscopic effects [J].
Finlay, WH ;
Stapleton, KW ;
Zuberbuhler, P .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1997, 149 (01) :63-72
[3]   THE EFFECT ON REGIONAL LUNG DEPOSITION OF COUPLED HEAT AND MASS-TRANSFER BETWEEN HYGROSCOPIC DROPLETS AND THEIR SURROUNDING PHASE [J].
FINLAY, WH ;
STAPLETON, KW .
JOURNAL OF AEROSOL SCIENCE, 1995, 26 (04) :655-670
[4]  
Finlay WH, 1997, PARTICUL SCI TECHNOL, V15, P185
[5]   OPERATING CHARACTERISTICS OF SOME COMPRESSED-AIR NEBULIZERS [J].
MERCER, TT ;
TILLERY, MI ;
CHOW, HY .
AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1968, 29 (01) :66-+
[6]   COMPARISONS OF PLANAR AND TOMOGRAPHIC GAMMA SCINTIGRAPHY TO MEASURE THE PENETRATION INDEX OF INHALED AEROSOLS [J].
PHIPPS, PR ;
GONDA, I ;
BAILEY, DL ;
BORHAM, P ;
BAUTOVICH, G ;
ANDERSON, SD .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1989, 139 (06) :1516-1523
[7]   EVAPORATION OF AQUEOUS AEROSOLS PRODUCED BY JET NEBULIZERS - EFFECTS ON PARTICLE-SIZE AND CONCENTRATION OF SOLUTION IN THE DROPLETS [J].
PHIPPS, PR ;
GONDA, I .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1994, 7 (03) :239-258
[8]  
Stapleton K.W., 1997, MA1 U ALB DEP MECH E
[9]  
STERK PJ, 1984, B EUR PHYSIOPATH RES, V20, P65