Continuous heart rhythm monitoring using mobile photoplethysmography in ambulatory patients

被引:10
作者
Hochstadt, Aviram [1 ,2 ]
Havakuk, Ofer [1 ,2 ]
Chorin, Ehud [1 ,2 ]
Schwartz, Arie Lorin [1 ,2 ]
Merdler, Ilan [1 ,2 ]
Laufer, Michal [1 ,2 ]
Lubman, Natan [3 ]
Ghantous, Eihab [1 ,2 ]
Viskin, Sami [1 ,2 ,3 ]
Rosso, Raphael [1 ,2 ]
机构
[1] Tel Aviv Univ, Tel Aviv Sourasky Med Ctr, Dept Cardiol, Tel Aviv, Israel
[2] Tel Aviv Univ, Sackler Sch Med, Tel Aviv, Israel
[3] CardiacSense LTD, Caesarea, Israel
关键词
Photoplethysmography; Continuous monitoring; Atrial fibrillation; Wearable diagnostics; ATRIAL-FIBRILLATION; CIRCADIAN VARIATION;
D O I
10.1016/j.jelectrocard.2020.04.017
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Wearable devices using photo-plethysmography (PPG) can accurately detect heart beats and may be useful for heart rate measurement and diagnosis of arrhythmias such as atrial fibrillation (AF). A previous study of a new portable PPG sensor (CardiacSense) showed high accuracy in heart rate measurement and AF detection in resting patients. We report a trial done to test the same device in active ambulatory patients with diverse characteristics. Methods: A cohort of 24 ambulatory volunteers, underwent simultaneous PPG recording and continuous electrocardiogram (ECG) recording under different environmental conditions and situations. Per study protocol, the subjects were diverse in age, BMI, hair density and skin tone. Four subjects had AF. Heart rate measurement using the PPG device was compared to measurements by ECG. Results: Of 163,527 recorded ECG-detected beats in the trial, 86,929 (53.2%) were also recorded by the PPG device. Most undetected heart beats were due to motion induced noise. Correlation between ECG and PPG was high (R = 0.94, p < 0.0001), yet in subjects with AF correlation was lower (R = 0.80, p < 0.0001). A Bland-Altman analysis showed the mean difference between measurements was -0.7 ms (95% limit of agreement -93.8 to 92.2). A total of 86,217 (99.9%) of all RR measurements were reliably measured (RR difference within 100 ms). Reliability was sustained (>99.8%) in subjects of all groups including subjects with AF. Conclusions: This study showed that, in the absence of movement-related noise, the CardiacSense PPG device can reliably detect HR in a variety of situations and subjects' characteristics. (C) 2020 Elsevier Inc. All rights reserved.
引用
收藏
页码:138 / 141
页数:4
相关论文
共 19 条
[1]   Heart rate variability: a review [J].
Acharya, U. Rajendra ;
Joseph, K. Paul ;
Kannathal, N. ;
Lim, Choo Min ;
Suri, Jasjit S. .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2006, 44 (12) :1031-1051
[2]   Asymptomatic Atrial Fibrillation The Case for Screening to Prevent Stroke [J].
Ben Freedman, S. ;
Lowres, Nicole .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2015, 314 (18) :1911-1912
[3]   Atrial Fibrillation Detection Using a Novel Cardiac Ambulatory Monitor Based on Photo-Plethysmography at the Wrist [J].
Bonomi, Alberto G. ;
Schipper, Fons ;
Eerikainen, Linda M. ;
Margarito, Jenny ;
van Dinther, Ralph ;
Muesch, Guido ;
de Morree, Helma M. ;
Aarts, Ronald M. ;
Babaeizadeh, Saeed ;
McManus, David D. ;
Dekker, Lukas R. C. .
JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2018, 7 (15)
[4]  
Bonomi AG, 2016, COMPUT CARDIOL CONF, V43, P277
[5]   The WATCH AF Trial: SmartWATCHes for Detection of Atrial Fibrillation [J].
Doerr, Marcus ;
Nohturfft, Vivien ;
Brasier, Noe ;
Bosshard, Emil ;
Djurdjevic, Aleksandar ;
Gross, Stefan ;
Raichle, Christina J. ;
Rhinisperger, Mattias ;
Stoeckli, Raphael ;
Eckstein, Jens .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2019, 5 (02) :199-208
[6]  
Floyer John., 1707, The Physicians Pulse-watch, Or an Essay to Explain the Old Art of Feeling the Pulse, and to Improve it by the Help of a Pulse Watch
[7]   Stroke prevention in atrial fibrillation [J].
Freedman, Ben ;
Potpara, Tatjana S. ;
Lip, Gregory Y. H. .
LANCET, 2016, 388 (10046) :806-817
[8]   High Prevalence of Atrial Fibrillation Among Patients With Ischemic Stroke [J].
Friberg, Leif ;
Rosenqvist, Marten ;
Lindgren, Arne ;
Terent, Andreas ;
Norrving, Bo ;
Asplund, Kjell .
STROKE, 2014, 45 (09) :2599-+
[9]   Continuous heart rate monitoring for automatic detection of atrial fibrillation with novel bio-sensing technology [J].
Hochstadt, Aviram ;
Chorin, Ehud ;
Viskin, Sami ;
Schwartz, Arie Lorin ;
Lubman, Natan ;
Rosso, Raphael .
JOURNAL OF ELECTROCARDIOLOGY, 2019, 52 :23-27
[10]   Heart rate monitors: State of the art [J].
Laukkanen, RMT ;
Virtanen, PK .
JOURNAL OF SPORTS SCIENCES, 1998, 16 :S3-S7