The construction of the scoliosis 3D finite element model and the biomechanical analysis of PVCR orthopaedy

被引:10
作者
Chen, Xuanhuang [1 ]
Cai, Hanhua [1 ]
Zhang, Guodong [1 ]
Zheng, Feng [1 ]
Wu, Changfu [1 ]
Lin, Haibin [1 ]
机构
[1] Putian Univ, Dept Orthoped, Affiliated Hosp, Putian 351100, Peoples R China
关键词
Scoliosis; Stress; Model; Vertebral body; PVCR;
D O I
10.1016/j.sjbs.2019.12.005
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Objective: The objective is to investigate the biomechanical conditions of the Posterior Vertebral Column Resection (PVCR) of the constructed scoliosis 3D finite element model. Methods: A patient with scoliosis was selected; before the PVCR orthopaedy, the patient was submitted to the radiography of normal and lateral full-length vertebral column scans and the total magnetic resonance imaging (MRI) scans; then, the idiopathic scoliosis model was constructed by the 3D finite element method, and the 3D finite element software utilized in the process of model construction included Mimics software, Geomagic Studio 12 software, and Unigraphic 8.0 (UG 8.0) software; in addition, PVCR orthopaedy was utilized to correct the scoliosis of the patient, and the biomechanical parameters, such as orthodontic force, vertebral body displacement, orthopedic rod stress, stress on the pinbone interface of the vertebral body surface, and the stress on the intervertebral disc, were studied. Results: The 3D effective finite element model of scoliosis was successfully constructed by the Mimics software, the Geomagic Studio 12 software, and the UG 8.0 software, and the effectiveness was tested. PVCR orthopaedy could effectively solve the problem of scoliosis. The magnitude of the orthodontic force that a patient needed depended on the physical conditions and the personal orthodontic requirements of the patient. The maximum vertebral body displacement on the X-axis was the vertebral body Ll, the maximum displacement on the Y-axis was the vertebral body T3, the maximum displacement on the Z-axis was the vertebral body T1, and the rang of orthopedic rod stress was 0.0050214e(7) MPa to 0.045217e(7) MPa, in which the maximum stress of 2 vertebral bodies in, above, and below the osteotomy area reached 0.045217e(7) MPa, the stress on the pin-bone interface of the T10 vertebral body surface reached 11.83 MPa, and the stress of T8/T9 intervertebral disc reached 13.84 MPa. Conclusion: The 3D finite element model based on 3D finite element software was highly efficient, and its numerical simulation was accurate, which was important for the subsequent biomechanical analysis of PVCR orthopaedy. In addition, the vertebral stress of PVCR orthopaedy was different in each body part, which was mainly affected by the applied orthodontic force and the sites of the orthodontic area. (C) 2019 The Author(s). Published by Elsevier B.V. on behalf of King Saud University.
引用
收藏
页码:695 / 700
页数:6
相关论文
共 14 条
[1]  
[Anonymous], [No title captured]
[2]  
[Anonymous], [No title captured]
[3]  
[Anonymous], [No title captured]
[4]  
[Anonymous], [No title captured]
[5]  
Cannestra A., 2018, GLOB SPINE J, V8, P2
[6]  
Clark M., 2019, Brain Neurosci. Adv, V3, P229
[7]   Effectiveness of braces designed using computer-aided design and manufacturing (CAD/CAM) and finite element simulation compared to CAD/CAM only for the conservative treatment of adolescent idiopathic scoliosis: a prospective randomized controlled trial [J].
Cobetto, N. ;
Aubin, C. E. ;
Parent, S. ;
Clin, J. ;
Barchi, S. ;
Turgeon, I. ;
Labelle, Hubert .
EUROPEAN SPINE JOURNAL, 2016, 25 (10) :3056-3064
[8]   Patient-specific finite element model of the spine and spinal cord to assess the neurological impact of scoliosis correction: preliminary application on two cases with and without intraoperative neurological complications [J].
Henao, Juan ;
Aubin, Carl-Eric ;
Labelle, Hubert ;
Arnoux, Pierre-Jean .
COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2016, 19 (08) :901-910
[9]   Biomechanical comparison of multilevel lateral interbody fusion with and without supplementary instrumentation: a three-dimensional finite element study [J].
Liu, Xilin ;
Ma, Jun ;
Park, Paul ;
Huang, Xiaodong ;
Xie, Ning ;
Ye, Xiaojian .
BMC MUSCULOSKELETAL DISORDERS, 2017, 18
[10]   Biomechanical Evaluation of Osteoporosis Treating with Mechanical Vibrations: A Finite Element Analysis [J].
Liu, Yun ;
Zhou, Guangquan ;
Xu, Nenggui .
NANOSCIENCE AND NANOTECHNOLOGY LETTERS, 2019, 11 (04) :506-513