Evaluation of a flash glucose monitoring system in nondiabetic dogs with rapidly changing blood glucose concentrations

被引:14
|
作者
Howard, Leigh A. [1 ]
Lidbury, Jonathan A. [1 ]
Jeffe, Nicholas [1 ]
Washburn, Shannon E. [2 ]
Patterson, Carly A. [2 ]
机构
[1] Texas A&M Univ, Coll Vet Med & Biomed Sci, Dept Small Anim Clin Sci, College Stn, TX 77843 USA
[2] Texas A&M Univ, Coll Vet Med & Biomed Sci, Dept Vet Physiol & Pharmacol, College Stn, TX 77843 USA
关键词
dog; FreeStyle Libre; hypoglycemia; interstitial glucose; OF-CARE GLUCOMETER; ACCURACY; METERS; PLASMA; CANINE; CATS;
D O I
10.1111/jvim.16273
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background A flash glucose monitoring system (FGMS; FreeStyle Libre) is useful for monitoring hypoglycemic dogs with diabetes. Objective To assess the utility of this FGMS in dogs with induced hypoglycemia and rapid fluctuations in blood glucose (BG) concentrations. Animals Twenty-four apparently healthy research (n = 10) and teaching (n = 14) dogs. Methods Prospective, observational study performed in tandem with a teaching laboratory. Regular insulin was administered to dogs and resulting hypoglycemia was corrected. Before insulin administration and every 10 minutes over a 90-minute period, serial measurements of interstitial glucose (IG) with FGMS and BG with a portable blood glucose meter (PBGM) and clinical chemistry analyzer concentrations were made. Portable blood glucose meter and FGMS readings were compared to that of the clinical chemistry analyzer. Analytical and clinical accuracy were assessed using ISO 15197:2013 criteria, including Parkes error grid analysis. Results The proportions of readings in the low BG range (BG <100 mg/dL) for which the test method measurement was within +/- 15 mg/dL of the reference BG for the PBGM and FGMS were 81.7% (161/197) and 39.1% (72/184), respectively. The proportions of readings for the PBGM and FGMS, which were not likely to affect clinical outcome according to Parkes error grid analysis, were 97.9% (233/238) and 80.1% (177/221), respectively. Conclusions and Clinical Importance In this model, there was limited agreement between the FGMS and reference standard BG measurements. The FGMS (measuring IG concentrations) was compared to peripheral BG concentrations, not brain-tissue glucose concentrations, and failed to reliably detect hypoglycemia.
引用
收藏
页码:2628 / 2635
页数:8
相关论文
共 50 条
  • [41] Comparison of a human portable blood glucose meter and automated chemistry analyser for measurement of blood glucose concentrations in healthy dogs
    Ismail-Hamdi, Sihem
    Romdane, Mohamed Nejib
    Ben Romdhane, Samir
    VETERINARY MEDICINE AND SCIENCE, 2021, 7 (06) : 2185 - 2190
  • [42] A pilot study to evaluate a novel subcutaneous continuous glucose monitoring system in healthy Beagle dogs
    Affenzeller, Nadja
    Benesch, Thomas
    Thalhammer, Johann G.
    Willmann, Michael
    VETERINARY JOURNAL, 2010, 184 (01) : 105 - 110
  • [43] Accuracy and precision assessment of a new blood glucose monitoring system
    Ji, Li-Nong
    Guo, Li-Xin
    Liu, Li-Bin
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2016, 54 (01) : 181 - 188
  • [44] Performance of the FreeStyle Libre Flash Glucose Monitoring System during an Oral Glucose Tolerance Test and Exercise in Healthy Adolescents
    Afeef, Sahar
    Tolfrey, Keith
    Zakrzewski-Fruer, Julia K.
    Barrett, Laura A.
    SENSORS, 2023, 23 (09)
  • [45] Evaluation of blood glucose fluctuation in Japanese patients with type 1 diabetes mellitus by self-monitoring of blood glucose and continuous glucose monitoring
    Kusunoki, Yoshiki
    Katsuno, Tomoyuki
    Nakae, Rie
    Watanabe, Kahori
    Akagami, Takafumi
    Ochi, Fumihiro
    Tokuda, Masaru
    Murai, Kazuki
    Miuchi, Masayuki
    Miyagawa, Jun-ichiro
    Namba, Mitsuyoshi
    DIABETES RESEARCH AND CLINICAL PRACTICE, 2015, 108 (02) : 342 - 349
  • [46] Flash Glucose Monitoring: The Future Is Here
    Garg, Satish K.
    Akturk, Halis Kaan
    DIABETES TECHNOLOGY & THERAPEUTICS, 2017, 19 : S1 - S3
  • [47] Analysis of "Multicenter Evaluation Study Comparing a New Factory-Calibrated Real-Time Continuous Glucose Monitoring System to Existing Flash Glucose Monitoring System"
    Peyser, Thomas
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2023, 17 (01): : 214 - 216
  • [48] Clinical evaluation of the eBsensor hand-held blood glucose monitoring system
    Chen, Chien-Chih
    Tai, Der-Yan
    Ho, Ping
    Lai, Yiu-Kay
    Lin, Jui-Jane
    Mao, Simon J. T.
    Lin, Chia-Chi
    Hsieh, Yu-Chen
    Chiang, Hsiang-Cheng
    Dai, Ken-Shwo
    CLINICA CHIMICA ACTA, 2007, 377 (1-2) : 170 - 173
  • [49] Evaluation of Three Human-Use Glucometers for Blood Glucose Measurement in Dogs
    dos Santos, Matheus Albuquerque Basilio
    Vargas, Alesssandra Martins
    Rosato, Paula Nunes
    Andrade, Carolina Gomes
    Martins, Camila Marinelli
    Petri, Giuliana
    VETERINARY MEDICINE INTERNATIONAL, 2022, 2022
  • [50] Hip sensor location for flash interstitial glucose monitoring system provides adequate clinical accuracy in non-diabetic hypoglycemic dogs
    Evans, Jeremy B.
    Lidbury, Jonathan A.
    Jeffery, Nicholas
    Washburn, Shannon E.
    Patterson, Carly A.
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2023, 84 (06)